Welcome Customer !

Membership

Help

Hangzhou Dawei Science and Education Instrument Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Hangzhou Dawei Science and Education Instrument Co., Ltd

  • E-mail

    416183088@qq.com

  • Phone

    13461341555

  • Address

    2nd Floor, Building 3, No. 1-1 Huayi Road, Yuhangxing Street, Yuhang District, Hangzhou City, Zhejiang Province

Contact Now

Stainless steel molecular distillation apparatus

NegotiableUpdate on 02/08
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
Material pretreatment of stainless steel molecular distillation apparatus: High viscosity materials (such as resins and oils) need to be vacuum degassed (≤ 100Pa, 2h) and added with 0.1% -0.5% antioxidant; Materials containing solid particles need to be filtered to a particle size of less than 50 μ m; Sealing verification: Helium mass spectrometer leak detector is used to detect system leakage rate ≤ 1; 10⁻⁹Pa m /s, Focus on inspecting the feed valve, cold trap interface, and scraper sealing ring;
Product Details

Stainless steel molecular distillation apparatusAnalysis of the entire process of operation details

1. Preliminary preparation and equipment inspection

Material pretreatment: High viscosity materials (such as resins and oils) require vacuum degassing (≤ 100Pa, 2h) and the addition of 0.1% -0.5% antioxidants; Materials containing solid particles need to be filtered to a particle size of less than 50 μ m;

Sealing verification: Helium mass spectrometer leak detector is used to detect system leakage rate ≤ 1 × 10 ⁻⁹ Pa · m ³/s, with a focus on checking the feed valve, cold trap interface, and scraper seal ring;

Cold trap pre cooling: Inject liquid nitrogen to 2/3 of the volume, run for 30 minutes to confirm that the temperature is stable at -150 ℃ to -120 ℃, and replenish liquid nitrogen every 2 hours;

Electrical safety: Grounding resistance ≤ 4 Ω, independent power supply wiring, avoid sharing circuits with high-power equipment.

2. Core operating procedures

Vacuum establishment: After the mechanical pump is pre pumped to 10Pa, switch to the diffusion pump and control the vacuum gradient (pre pumping section 10-100Pa, distillation section 0.1-10Pa, collection section 10-100Pa);

Temperature control: three-stage heating strategy - evaporation surface ≤ 200 ℃ (to avoid coking), condensation surface -80 ℃ to -50 ℃, scraper preheating to 80 ℃;

Feeding and film formation: Start the film scraping motor (speed 200-400rpm), with an initial feeding speed of ≤ 10mL/min, gradually increase to 50mL/min after the system stabilizes;

Distillate collection: Light components (such as terpenes) are first evaporated and then liquefied by a condenser before entering the primary collection bottle; Residual components (such as wax) remain in the evaporator and are discharged through a residual valve;

Process monitoring: Record vacuum degree, temperature, and feed rate data every 15 minutes. When the temperature fluctuation on the evaporation surface is greater than ± 5 ℃, an alarm is triggered, and the heater power or scraper operation status needs to be checked.

3. Ending and Cleaning

Shutdown sequence: Close the feed valve → Stop heating → Wait for the temperature to drop below 80 ℃ → Close the vacuum pump → Slowly release pressure to atmospheric pressure;

Cleaning process: Ethanol/isopropanol is circulated to rinse the distillation chamber three times, and high viscosity materials need to be rinsed with hot water; Clean the condenser quarterly with a 5% citric acid solution;

Residue treatment: Waste materials are disposed of according to hazardous waste standards, with a solvent recovery rate of ≥ 95%, meeting environmental emission requirements.

Stainless steel molecular distillation apparatusFull cycle management of maintenance and upkeep

1. Daily maintenance

Cleaning and disinfection: Wipe the surface of the equipment with 75% ethanol, and disinfect the internal pipelines with 0.5% sodium hypochlorite solution every month;

Seal inspection: Check the wear of O-rings and mechanical seals after each batch of operation, with a replacement cycle of ≤ 100 batches;

Lubrication management: Food grade lubricating grease is added to the scraper bearings every month, and the lubricating oil for the gearbox is replaced every quarter;

Vacuum pump maintenance: Mechanical pump oil is replaced every 3 months, diffusion pump oil is replaced every month, and filter element is cleaned every quarter.

2. Regular maintenance

Calibration and testing: Temperature sensors are calibrated every six months (compared to standard platinum resistors), and vacuum gauges are sent for inspection every year;

Performance testing: Conduct a maximum vacuum degree test (≤ 0.01Pa) and a condensation efficiency test (condensation recovery rate ≥ 98%) every six months;

Structural inspection: Conduct an annual endoscopic inspection of the evaporator and condenser to assess corrosion and scaling conditions;

Safety valve calibration: The pressure safety valve is calibrated once a year to ensure that the opening and closing pressure meets the design requirements.

3. Long term disuse and storage

Shutdown preparation: Drain all media, blow the system with nitrogen, and apply rust proof oil to the surface of the equipment;

Storage environment: temperature 5-35 ℃, humidity ≤ 60%, avoid direct sunlight, covered with dust cover;

Regular startup: Start up and run for 1 hour every month to check if all system functions are normal.

Stainless steel molecular distillation apparatusSafety regulations and emergency response

1. Operational safety regulations

Personal protection: Wear anti freezing gloves (nitrile rubber), goggles, and lab clothes during operation to avoid contact with low-temperature surfaces (below -80 ℃);

Electrical safety: Do not operate with wet hands, ensure good grounding of equipment, and equip with leakage protection devices;

Material safety: Familiar with MSDS of materials, thermal sensitive substances need to be tested for thermal stability in advance;

Ventilation requirements: The operating room should maintain good ventilation and be equipped with combustible gas alarms.

2. Emergency response plan

Vacuum leakage: Immediately close the feed valve, start nitrogen purging, and investigate the leakage point;

Temperature out of control: Cut off the heating power and activate the emergency cooling system;

Material leakage: Absorb with oil absorbing cotton, cover with sand, and dilute with ventilation;

Fire accident: Use dry powder fire extinguishers and strictly prohibit using water to extinguish fires.