E-mail
sales@gmci-china.cn
-
Phone
13817443004
-
Address
Room 703, Building B, No. 455 Yanzhan Road, Songjiang District, Shanghai
Dianlishi (Shanghai) Electronics Co., Ltd
sales@gmci-china.cn
13817443004
Room 703, Building B, No. 455 Yanzhan Road, Songjiang District, Shanghai
Medical handheld electrical safety analyzer
-Tested in accordance with standards such as IEC/EN 62353, NFPA-99, AS/NZ 3551, etc
-Built in electronic data storage
-Flexible user-defined testing program
-Simple PC communication link
-Insulation and grounding/grounding binding tests for battery powered systems
-50/100/250/500V DC insulation test
-Automatic secondary verification ensures that D obtains the correct result in one attempt
Rigel 62353+Medical handheld electrical safety tester
Rigel 62353+is a small and flexible electrical safety analyzer on the market, with battery powered grounding/grounding binding and insulation testing functions. It provides an accurate and fast testing solution that complies with the IEC 62353 safety standard.
62353+can provide a series of tests without the need for a main power source, and can perform insulation and grounding/grounding binding tests for battery power supply. This provides the ability to move freely in crowded rooms during installation, eliminating the inconvenience of dragging the main power cable.
62353+has low voltage insulation resistance tests (50V DC and 100V DC) to supplement the existing 250V DC and 500V DC insulation tests in its predecessor Rigel 62353. These tests meet the testing requirements of manufacturers for an increasing number of devices operating on 24V DC and 48V DC, such as operating tables, lighting, and mobile X-ray machines.
Rigel 62353+Electrical Safety Analyzer Product Application:
-Routine testing of medical electrical equipment
-Service tool for performance testing
-Asset Management
-Quickly and efficiently test IEC wires
-Binding testing of medical and non-medical devices
Technical Specifications
Grounding continuity
-Method: Use the two-wire technique with the "zero" lead function.
-Test current:>+200mA, -200mA DC enters 2 ohms
-Z-test voltage: 4-24V RMS o/c
-Measurement range (low range): 0.001-0.999 ohms @ 0.001 ohm resolution
-Measurement range (medium range): 1.00-9.99 ohms @ 0.01 ohm resolution
-Measurement range (high range): 10.0-19.9 ohms @ 0.1 ohm resolution
-Accuracy: ± 3% reading+10 micro ohms
Direct leakage measurement
-Measurement range: 4 μ A to 9999 μ A
-Accuracy: ± 5% reading or reading+2 counts
-Measurement equipment: According to the requirements of IEC 60601-1
-Measurement type: True Significant Value
Alternative leakage measurement
-Test voltage: main frequency 250V
-Test current: limited to 3.5 mA
-Measurement range: 4 μ A to 9999 μ A
-Measurement resolution: 1 μ A
-Measurement accuracy: ± 5% reading+2 counts
-Measurement type: True Significant Value
-Measurement equipment: According to the requirements of IEC 60601-1
Main power socket test
-Input voltage range: 0-300V AC
-Z high current: 16A
-Measurement: L-E, N-E
-Accuracy: ± 5% reading+2 counts
Basic specifications
-Main power supply: 230V AC ± 10%, 50-60Hz+/-1Hz, 120V AC ± 10%, 60Hz+/-1Hz (US model)
-Battery: 6 x 1.5V AA (battery not included)
-Weight: 1.6 kg/3.5 lbs (including battery)
-Size: (length x width x height) 270 x 110 x 75 mm/10.5 x 4 x 3 inches
-Operating conditions: 0 ° C-40 ° C, 32 ° F-104 ° F, 0-90% RH - no condensation
-Storage environment: -15 ° C -+60 ° C/5 ° F -140 ° F
-Environmental Protection: IP 40
insulation resistance
-Measurement: EUT to ground/ground, EUT to AP, AP to ground/ground
-Voltage: 50/100/250&500V DC @ 1mA
-Range (low range) @ 50V DC: 0.01Mohms -10 Mohms
-Range (low range) @ Above 50V DC: 0.01Mohms -20Mohms
-Accuracy (low range): ± 5% reading+2 counts
-Range (High Range) @ 250V DC: 20Mohms – 50Mohms
-Range (High Range) @ 500V DC: 20Mohms – 100Mohms
-Accuracy (high range): ± 10%+2 counts
-Resolution: 0.01Mohms
Differential leakage measurement
-Measurement range: 75 μ A to 9999 μ A
-Accuracy: ± 5% reading+5 counts
-Measurement/display resolution: 1 μ A
-Measurement type: True Significant Value
-Measurement equipment: Frequency response characteristics comply with IEC 60601-1 requirements
power measurement
-Method: VA rated value
-Range: 0.1kVA to 4kVA
-Accuracy: ± 10%+2 counts
IEC Main Power Lead Test
-Test duration: IEC main power lead test
-Test: Continuity, insulation, and polarity of all conductors
Standard accessories (provided with 62353+)
-Calibration certificate
-Carrying case
-Grounding/Grounding Binding Test Probe with Clamp
-Grounding/Grounding Binding Clip Lead
-Patient application module
-2 application adapters
-Detachable 2-meter main power cord
-Electronic manual
-Quick Start Guide
-Application software