Defibrillator pacemaker analyzer

NegotiableUpdate on 12/15
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Rigel Uni Pulse 400 defibrillator pacemaker analyzer application scope: The Uni Pulse 400 defibrillator analyzer is a super compact and portable energy tester on the market, which can accurately verify all defibrillation and pacing instruments currently on the market. Defibrillators are compatible with traditional single-phase wave modes, standard biphasic wave, biphasic square wave, modulated biphasic wave, pulse mode defibrillators, as well as all automatic external defibrillators (AEDs).

Product Details

Rigel Uni-Pulse 400Defibrillator pacemaker analyzerProduct Introduction

Application scope of Rigel Uni Pulse 400 defibrillator analyzer:

The Uni Pulse 400 pacemaker analyzer is an ultra compact and portable energy testing device on the market, capable of accurately verifying all defibrillation and pacing instruments currently on the market. Defibrillators are compatible with traditional single-phase wave modes, standard biphasic wave, biphasic square wave, modulated biphasic wave, pulse mode defibrillators, as well as all automatic external defibrillators (AEDs).

除颤器起搏器分析仪

Rigel Uni-Pulse 400Defibrillator pacemaker analyzerKey Features:

☆ Test various types of energy release, compatible with single-phase, two-phase, two-phase square wave, pulse and various discharge forms

☆ Test the synchronous electrical cardioversion performance of the heart

☆ Test the performance of automatic external defibrillation

☆ Provide patient sign testing loads with different chest impedances ranging from 10-300 Ω

☆ Support energy release (voltage, current, load) process data analysis, waveform drawing, and comparison of morphological details

☆ Built in standard analog signal generator, capable of outputting electrical performance test signals and standard electrocardiogram signals required for electrocardiogram testing

☆ Simulate over 40 selectable arrhythmias

☆ Can detect the function of percutaneous pacemaker

☆ Simulation of 5 types of pacemakers

☆ Charge and discharge time test

☆ High sampling rate supports fine data analysis, USB can be directly downloaded to PC

☆ Rechargeable battery, can run without external power supply

☆ Color LCD display screen, large characters, clearer reading

☆ Use full character buttons and navigation keys to quickly and intuitively enter customer information and store all test results

☆ Multi language firmware to meet different user needs

☆ Portable, compact, and sturdy


Product Introduction:


Used for defibrillators (monitors)& Automatic External Defibrillator (AED)& Pacemaker performance testing


Rigel Uni Pulse 400 Defibrillator Analyzer Technical Parameters:

Defibrillation energy measurement

load resistor

50 Ω± 1% insensitivity (with built-in standard load)

External variable load box 10-300 Ω; ±1%,

10,15,20,25,50,75,100,125,150,175,200,250,300  Ω

(External load  Optional)

Low energy range

0-199.9 joules

accuracy

± 1% ± 0.1 joules of reading

High energy range

200-600 joules

accuracy

± 1% ± 1.0 joules of reading

display resolution

0.01 joules

peak voltage

0 - 5000Vdc

peak current

0 – 120A

sampling rate

250 kHz sampling frequency

pulse width

5us - 120ms

counter

Charging time

0-999.9s, Accuracy ± 0.05s

Pulse waveform of external automatic defibrillator (AED)

Normal sinus rhythm (NSR)

20 - 300 BPM, Amplitude 1.00mV (± 2%)

Cardiac arrest (ASYS)


Ventricular fibrillation rough (VFBC)

240 BPM, Amplitude 1.00mV (± 2%)

Ventricular fibrillation fine (VFBF)

240 BPM, Amplitude 1.00mV (± 2%)

Monomorphic ventricular tachycardia (MVT)

210 BPM, Amplitude 1.00mV (± 2%)

Atrial fibrillation (AFB)

20 - 300 BPM, Amplitude 1.00mV (± 2%)

ECG simulator with 12 lead ECG simulation, including high-level output

Electrocardiogram waveform - sinus type

20 - 300 BPM, Accuracy ± 1%

Amplitude 0.50-5.00mV, accuracy ± 2%

Normal sinus rhythm (NSR)

20 - 300 BPM, Amplitude 0.50-5.00mV (± 2%)

ST wave (STE)

20 - 300 BPM, Amplitude 0.50-5.00mV (± 2%)

ST Depression (STD)

20 - 300 BPM, Amplitude 0.50-5.00mV (± 2%)

Myocardial infarction (MI)

20 - 300 BPM, Amplitude 0.50-5.00mV (± 2%)

高 T(TT)


Cardiac arrest (ASYS)


Electrocardiogram waveform - ventricular arrhythmia

Ventricular premature beats intermittent (PVCI)

80 BPM, Amplitude 0.50-5.00mV (± 2%)

Bigeminy (BIG)

80 BPM, Amplitude 0.50-5.00mV (± 2%)

Trigeminy (TRIG)

80 BPM, Amplitude 0.50-5.00mV (± 2%)

Ventricular flutter (VFLT)

240 BPM, Amplitude 0.50-5.00mV (± 2%)

Ventricular fibrillation rough (VFBC)

240 BPM, Amplitude 0.50-5.00mV (± 2%)

Ventricular fibrillation fine (VFBF)

240 BPM, Amplitude 0.50-5.00mV (± 2%)

Monomorphic ventricular tachycardia (MVT)

210 BPM, Amplitude 0.50-5.00mV (± 2%)

Right ventricular premature beats (RFPVC)

80 BPM, Amplitude 0.50-5.00mV (± 2%)

Electrocardiogram waveform - atrial conduction arrhythmia

First level AV Bok (FAVB)

80 BPM, Amplitude 0.50-5.00mV (± 2%)

Secondary AV module - Mobitz I (SAVB_MI)

80 BPM, Amplitude 0.50-5.00mV (± 2%)

Secondary AV module - Mobitz II (SAVB_MII)

80 BPM, Amplitude 0.50-5.00mV (± 2%)

Triple AV module (TAVB)

50 BPM, Amplitude 0.50-5.00mV (± 2%)

Electrocardiogram waveform - atrial arrhythmia

Sinus arrhythmia (SAR)

20 - 300 BPM, Amplitude 0.50-5.00mV (± 2%)

Leakage (MB)

20 - 300 BPM, Amplitude 0.50-5.00mV (± 2%)

Atrial flutter (AFLT)

300 BPM, Amplitude 0.50-5.00mV (± 2%)

Atrial fibrillation (AFB)

20 - 300 BPM, Amplitude 0.50-5.00mV (± 2%)

Paroxysmal atrial tachycardia (PAT)

180 BPM, Amplitude 0.50-5.00mV (± 2%)

Premature Contraction of Connections (PJC)

20 - 300 BPM, Amplitude 0.50-5.00mV (± 2%)

Compatible pacing signals with pacemaker types

Atrial inhibitory pacing (AAI) - on-demand type

20 - 300 BPM, Pulse amplitude 0.50-5.00mV, pulse width 0.1-2.0ms

Asynchronous atrial pacing (AOO) - fixed frequency

20 - 300 BPM, Pulse amplitude 0.50-5.00mV, pulse width 0.1-2.0ms

Adjustable pacemaker (PCR) - in vitro remote control

20 - 300 BPM, Pulse amplitude 0.50-5.00mV, pulse width 0.1-2.0ms

Ventricular inhibitory pacing (VVI) - on-demand type

20 - 300 BPM, Pulse amplitude 0.50-5.00mV, pulse width 0.1-2.0ms

Dual chamber sensing and dual chamber pacing sequential pacing (DDD) - multi-mode

20 - 300 BPM, Pulse amplitude 0.50-5.00mV, pulse width 0.1-2.0ms

R-wave inhibitory pacing (RWD) - on-demand type

20 - 300 BPM, Pulse amplitude 0.50-5.00mV

ECG performance waveform

正弦 (SINE)

0.1 - 300Hz,1.00 - 10.00mV

Square wave (SQ)

0.1 - 300Hz,1.00 - 10.00mV

Triangular waves (TRI)

0.1 - 300Hz,1.00 - 10.00mV

Sawtooth (SAW)

0.1 - 300Hz,1.00 - 10.00mV

Anti aliasing (INVSAW)

0.1 - 300Hz,1.00 - 10.00mV

Pulse (PULSE)

0.1 - 300Hz,0.50 - 5.00mV

Noise in electrocardiogram

amplitude

0 - 10.00mV

frequency

50 - 60Hz

ECG output mode

low level

Equipped with 12 lead ECG terminals and defibrillator pad adapters

high level

Jack audio terminal output

Accuracy of electrocardiogram signals

heart rate

±1%

amplitude

± 2% (LA-LL, lead II), ± 10% (defibrillator pad)

Compatible pacing methods

Atrial inhibitory pacing (AAI) - on-demand type

20 - 300 BPM, Pulse amplitude 0.50-5.00mV, pulse width 0.1-2.0ms

Asynchronous atrial pacing (AOO) - fixed frequency

20 - 300 BPM, Pulse amplitude 0.50-5.00mV, pulse width 0.1-2.0ms

Adjustable pacemaker (PCR) - in vitro remote control

20 - 300 BPM, Pulse amplitude 0.50-5.00mV, pulse width 0.1-2.0ms

Ventricular inhibitory pacing (VVI) - on-demand type

20 - 300 BPM, Pulse amplitude 0.50-5.00mV, pulse width 0.1-2.0ms

Dual chamber sensing and dual chamber pacing sequential pacing (DDD) - multi-mode

20 - 300 BPM, Pulse amplitude 0.50-5.00mV, pulse width 0.1-2.0ms

R-wave inhibitory pacing (RWD) - on-demand type

20 - 300 BPM, Pulse amplitude 0.50-5.00mV

Output testing of pacemaker

Load of percutaneous pacing mode

50 Ω (built-in fixed), optional external variable load 25-200 Ω

accuracy

1%, no sensation

Overvoltage protection

5000V

Variable test load for implantable pacemakers

50 to 1600 Ω, 50 Ω step

accuracy

1%, no sensation

Pulse frequency (pulse interval)

5.0 to 800 ppm

accuracy

+/-(0.5% reading+0.1ppm)

heart rate

20 - 300 bpm

Underload and overload

85% (20 bpm) and 115% (300 bpm)

Waveform selection

Sinus heart rate, ventricular fibrillation (coarse), ventricular fibrillation (fine), monophyletic ventricular tachycardia, atrial fibrillation, missed beats, R-wave detection signal

Pulse current amplitude

5.00 - 200mA

accuracy

+/-(1% reading+0.02mA)

Types of pacing current: average value (RMS), leading edge, trailing edge, peak value (peak value in pulse mode)

pulse width

1.00 - 100ms

accuracy

+/-(0.5% reading+0.01ms)

pulse energy

1μJ - 2.00J

accuracy

+/- 4%+10μJ

Compatible with algorithms from pacemaker manufacturers

CU Medical,GE,HP,Laerdal,Mindray, Philips,PhysioControl,Schiller,WelchAllyn,Zoll

Pacemaker refractory testing

Refractory period (for dual pulse generators, pacing refractory period and sensing refractory period)

15 - 500mS

accuracy

+/- 1ms

Pacemaker sensitivity (perception threshold) test

Pacing signal waveform

R-wave

polarity

Positive and negative poles can be set

Pacemaker anti-interference test

heart rate

20-300 bpm

Frequency of power frequency disturbance

50 or 60 Hz

Noise level

0-15.0mV

General Specifications

Language setting

Chinese, English, etc

Overall dimensions

220mm x 150mm x 90mm (length x width x height)

weight

1.5kg

operating power supply

12V/2400mAh nickel metal hydride battery pack

Grid power supply

110/230V AC; 48 ~ 66Hz, 35VA

storage environment

-15°C ~ +60°C

runtime environment

0°C ~ +40°C

Environmental protection level

IP 40

communication

USB

display

High definition LCD color screen

storage

100 test results, including charts