Standard:
GB 9706.1-2020 (IEC60601) Medical Electrical Equipment Part 1: General Requirements for Basic Safety and Basic Performance -5.1 Type Tests/7.2.6 Connection to Power Supply Network/7.2.10 Application Part/8.3 Classification of Application Parts B/BF/CF
GB 4793.1 (IEC61010) Safety requirements for electrical equipment for measurement, control and laboratory use Part 1
YY/T 0841-2011 (IEC62353-2007) Periodic testing and post repair testing of medical electrical equipment
VDE 751、ANSI/AAMI ES1:1993、NFPA-99、AN/NZS 3551
JJF (Zhejiang) 1147-2018 Calibration Specification for Enteral Nutrition Pump (Patient Leakage Current Test)
This is a fully automatic warning electrical safety testing instrument suitable for medical facilities and testing laboratories to conduct electrical safety testing on medical equipment. With the advantages of its special built-in electrical circuit prompt diagram and 50 automatic testing sequences in the board storage, engineers only need to memorize simple symbols to operate the equipment safely and correctly, greatly reducing the technical difficulty of electrical testing for non professionals.
Under the condition of combining its special asset management function, directly call the one-to-one characteristic test sequence for fully automated electrical testing, and automatically record the results after completion. You can browse test records in real time, achieving paperless office for convenience and speed.

One of the features of this asset management function is the ability to perform programmatic and hierarchical management of equipment users. By presetting the operating program and locking the operation keys, it can effectively prevent quality control personnel who are unfamiliar with the tested equipment from illegally testing and causing asset losses.
It is a small-sized and ultra portable one handed handheld electrical safety analyzer on the market. Its special panel integrated keyboard and board type memory make it free from the annoyance of carrying a laptop, allowing senior managers to edit the required test sequences anytime and anywhere.
Built in solutions for electrical safety testing standards such as GB 9706.1 (IEC60601-1) and GB 4793.1 (IEC61010). At the same time, it has grounding impedance testing, leakage current testing for various types of tests, and insulation testing for four different voltage levels. It can provide 50V/100V/250V/500V DC insulation testing voltage, which is more in line with the testing requirements of 24Vdc and 48Vdc operating equipment executed by manufacturers such as operating tables, surgical lamps, and portable X-ray machines. It provides great convenience for routine inspection/maintenance performance testing/asset management of medical electrical equipment, compliance with IEC standards, testing of medical and non-medical devices, and fixed installation testing of large equipment in different situations.

The special standard AA battery powered testing mode is no longer limited by the location without power supply, and can measure wherever you go. Can perform ground impedance testing, touch or point-to-point leakage current testing, and insulation testing.
Special power cable testing and evaluation function, built-in wire gauge calculation table can automatically calculate the limit value of conduction impedance. About 90% of medical failures are caused by unqualified electrical cables of medical equipment, so quickly determining the conductivity and insulation of power lines is of utmost importance in the work.

Its various levels of testing programs, including fully automatic, semi-automatic, and manual, as well as the ability to store and print real-time data reports at any time, make it an extremely rigorous safety analyzer among similar products. After combining with the asset management software Med eBase, its expertise in creating test sequences and its ability to manage data and assets can be further expanded.
To further improve the flexibility and efficiency of on-site operations, a range of optional accessories are supported, including barcode scanners and Bluetooth printers. Combining asset management barcodes to achieve one-on-one testing of one device and one condition.

Electrical Performance Analysis
Electrical safety testing:
ground impedance
Qualification of power supply wires
insulation resistance
(Power cord, power PE, power accessible non grounded conductive component NE, equipment casing, patient application end AP-PE, AP-NE, application terminal pressurization conditions)
GB9706.1 / IEC 60601-1:
leakage current
Shell leakage
Patient leakage
Patient assisted leakage
F-type patient leakage
IEC 61010 / IEC 62353:
Contact leakage
Equipment leakage (direct, differential, and substitute)
Application of three-phase power supply equipment:
Ground conduction resistance
Insulation properties of materials
Leakage current (equipment leakage, shell leakage, patient leakage)
You can use the above variables or combinations to create custom tests.
Grounding continuity | |
test method |
2-wire pre pulse technology, using wire zero elimination impedance correction technology |
Pre pulse technology |
25A peak current, eliminating thin film resistance |
test current |
>200mA |
Current loading time |
6 optional values: 2 seconds/5 seconds/10 seconds/20 seconds/30 seconds/duration |
test voltage |
Effective value of 4-24V (IEC60601 test specifies 6V) |
Measurement range (small range) |
0.001 - 0.999Ω / 分辨率 0.001Ω |
Measurement range (medium range) |
1.00 - 9.99Ω / 分辨率 0.01Ω |
Measurement range (large range) |
10.0 - 19.9Ω / 分辨率 0.1Ω |
precision |
Reading ± 2%+10 m Ω |
insulation resistance | |
measurement items |
5 optional values: The power supply end of the test object - the shell part that fails to protect the grounding (Class I), the power supply end of the test object - the shell part (Class II), the equipment application part - the shell part that fails to protect the grounding (Class I, with grounding), the equipment application part - the shell part that fails to protect the grounding (Class II), the power supply end of the test object - the equipment application part |
Loading voltage |
4 optional values: 50 / 100 / 250 / 500 VDC @1mA |
Voltage loading time |
6 optional values: 2 seconds/5 seconds/10 seconds/20 seconds/30 seconds/duration |
Test limit conditions |
13 optional values: 0.01/0.1/0.3/1.0/2.0/4.0/5.0/7.0/10.0/20.0/50.0/70.0/100.0MΩ |
|
Measurement range (small range) 50V DC Measurement range (small range) above 50V DC Accuracy (small range) |
0.01MΩ-10MΩ 0.01MΩ-20MΩ Reading ± 1%+2 counts |
|
Measurement range (large range) 250V DC Measurement range (large range) 500V DC Accuracy (large range) |
20MΩ-50MΩ 20MΩ-100MΩ Reading ± 3%+2 counts |
resolution |
0.01MΩ |
Direct method for measuring equipment leakage current |
(IEC60601/GB9706.1&IEC62353/YYT 0841 Medical Electrical Equipment Periodic Testing and Post Repair Testing) |
measurement range |
4 ~ 9999μA |
precision |
Reading ± 3%+2 counts |
Loading power supply voltage on the application end |
Only F-type devices can be loaded with 110% grid power |
Measurement device |
According to the requirements of IEC 60601-1 |
measurement type |
Compliant with IEC 60601 measurement regulations for patient (or auxiliary) leakage current, providing true RMS values for both AC and DC measurements |
loading time |
6 optional values: 2 seconds/5 seconds/10 seconds/20 seconds/30 seconds/duration |
Limit conditions for evaluation results |
10 optional values: 10/ 25/ 50/ 100/ 300/ 500/1000/2500/5000/10000uA |
Application Part AP |
1-10 lead independent editing |
Anti electric shock level |
Class I / Class II |
Differential method for measuring equipment leakage current (IEC62353&YYT 0841-2011 Medical electrical equipment Periodic testing and post repair testing of medical electrical equipment) | |
measurement range |
75μA to 9999μA |
precision |
Reading ± 3%+5 counts |
Measurement/Display |
Resolution 1 μ A |
measurement type |
True effective value |
Measurement device |
Frequency response complies with IEC 60601-1 regulations |
loading time |
6 optional values: 2 seconds/5 seconds/10 seconds/20 seconds/30 seconds/duration |
Limit conditions for evaluation results |
10 optional values: 10/ 25/ 50/ 100/ 300/ 500/1000/2500/5000/10000uA |
Alternative method for measuring equipment leakage current (IEC62353&YYT 0841-2011 Medical electrical equipment Periodic testing and post repair testing of medical electrical equipment) | |
test voltage |
250V power frequency |
test current |
3.5mA current limiting |
measurement range |
4μA - 9999μA |
measurement resolution |
1μA |
measurement accuracy |
Reading ± 3%+2 counts |
measurement type |
True effective value |
Measurement device |
According to IEC 60601-1 |
power measurement | |
method |
apparent power |
scope |
0.1kVA-4kVA |
precision |
± 5%+2 counts |
loading time |
6 optional values: 2 seconds/5 seconds/10 seconds/20 seconds/30 seconds/duration |
Network power output test | |
Input voltage range |
0-300VAC |
Can input current |
16A |
precision |
Reading ± 5%+2 counts |
Pluggable IEC network power cord test | |
Test time |
2 seconds |
test |
Conductivity, insulation, and polarity changes |
Conductivity limit |
Automatic calculation of conductivity limit using wire gauge calculation table 15 optional values: 0.01/0.02/0.03/0.04/0.05/0.06/0.07/0.08/0.10/0.20/0.30/0.50/1.0/2.0/5.0Ω |
Insulation limit |
13 optional values: 0.01/0.1/0.3/1.0/2.0/4.0/5.0/7.0/10.0/20.0/50.0/70.0/100.0MΩ |
Loading voltage conditions |
4 optional values: 50 / 100 / 250 / 500 VDC @1mA |
Test polarity |
auto reverse |
General Specifications | |
main power supply |
230V AC ±10%, 50-60Hz +/- 1Hz 120V AC ± 10%, 60Hz+/-1Hz (American model) |
battery |
6 x 1.5V AA |
weight |
1.6 kilograms (including battery) |
size |
270 x 110 x 75 mm |
Operating conditions |
0℃ - 40℃, 0-90% RH - NC |
storage environment |
-15℃ - +60℃ / 5℃ - 140 ℉ |
Environmental protection |
IP 40 |
Recommended configuration:1 set of 288 type electrical safety tester host, 1 set of 4mm buckle type pin type dual-use terminal posts for ECG leads (10 pieces), 1 detachable 2-meter IEC power cord, 1 application terminal test connection module, 1 standard test probe, fixture and cable, 1 grounding test wide fixture, verification test kit (including 1 pair of thin head red and black probe, 1 thin head crocodile clip, 1 pair of 2-meter test cable, 1 0.2 Ω low resistance verification box, 1 10M Ω high resistance verification box, 1 surface insulation defect detection brush (optional accessory), 1 tool package), 1 Chinese quick use guide, 1 Chinese user manual, factory test data report 1 set, 1 one handed portable bag, 1 portable bag