Rigel 288+Fully Automatic Electrical Safety Tester Application Range: $r $n This is a fully automatic warning electrical safety testing instrument suitable for medical facilities and testing laboratories to conduct electrical safety testing on medical equipment. With its special built-in electrical circuit prompt diagram and 50 automatic test sequences stored in the board memory, engineers only need to memorize simple symbols to safely and correctly operate the equipment, greatly reducing the technical difficulty of electrical testing for non professionals. $r $n $r $n, combined with its unique asset management capabilities, directly calls one-on-one feature test sequences for fully automated testing

Product Details

Fully automatic electrical safety detectorRigel 288+Product Introduction:

Standard:

GB 9706.1-2020 (IEC60601) Medical Electrical Equipment Part 1: General Requirements for Basic Safety and Basic Performance -5.1 Type Tests/7.2.6 Connection to Power Supply Network/7.2.10 Application Part/8.3 Classification of Application Parts B/BF/CF

GB 4793.1 (IEC61010) Safety requirements for electrical equipment for measurement, control and laboratory use Part 1

YY/T 0841-2011 (IEC62353-2007) Periodic testing and post repair testing of medical electrical equipment

VDE 751、ANSI/AAMI ES1:1993、NFPA-99、AN/NZS 3551

JJF (Zhejiang) 1147-2018 Calibration Specification for Enteral Nutrition Pump (Patient Leakage Current Test)


Fully automatic electrical safety detectorRigel 288+application scope:


This is a fully automatic warning electrical safety testing instrument suitable for medical facilities and testing laboratories to conduct electrical safety testing on medical equipment. With the advantages of its special built-in electrical circuit prompt diagram and 50 automatic testing sequences in the board storage, engineers only need to memorize simple symbols to operate the equipment safely and correctly, greatly reducing the technical difficulty of electrical testing for non professionals.

Under the condition of combining its special asset management function, directly call the one-to-one characteristic test sequence for fully automated electrical testing, and automatically record the results after completion. You can browse test records in real time, achieving paperless office for convenience and speed.

全自动电气安全检测仪

One of the features of this asset management function is the ability to perform programmatic and hierarchical management of equipment users. By presetting the operating program and locking the operation keys, it can effectively prevent quality control personnel who are unfamiliar with the tested equipment from illegally testing and causing asset losses.

It is a small-sized and ultra portable one handed handheld electrical safety analyzer on the market. Its special panel integrated keyboard and board type memory make it free from the annoyance of carrying a laptop, allowing senior managers to edit the required test sequences anytime and anywhere.

Built in solutions for electrical safety testing standards such as GB 9706.1 (IEC60601-1) and GB 4793.1 (IEC61010). At the same time, it has grounding impedance testing, leakage current testing for various types of tests, and insulation testing for four different voltage levels. It can provide 50V/100V/250V/500V DC insulation testing voltage, which is more in line with the testing requirements of 24Vdc and 48Vdc operating equipment executed by manufacturers such as operating tables, surgical lamps, and portable X-ray machines. It provides great convenience for routine inspection/maintenance performance testing/asset management of medical electrical equipment, compliance with IEC standards, testing of medical and non-medical devices, and fixed installation testing of large equipment in different situations.

全自动电气安全检测仪

The special standard AA battery powered testing mode is no longer limited by the location without power supply, and can measure wherever you go. Can perform ground impedance testing, touch or point-to-point leakage current testing, and insulation testing.

Special power cable testing and evaluation function, built-in wire gauge calculation table can automatically calculate the limit value of conduction impedance. About 90% of medical failures are caused by unqualified electrical cables of medical equipment, so quickly determining the conductivity and insulation of power lines is of utmost importance in the work.

全自动电气安全检测仪


Its various levels of testing programs, including fully automatic, semi-automatic, and manual, as well as the ability to store and print real-time data reports at any time, make it an extremely rigorous safety analyzer among similar products. After combining with the asset management software Med eBase, its expertise in creating test sequences and its ability to manage data and assets can be further expanded.

To further improve the flexibility and efficiency of on-site operations, a range of optional accessories are supported, including barcode scanners and Bluetooth printers. Combining asset management barcodes to achieve one-on-one testing of one device and one condition.

全自动电气安全检测仪





Electrical Performance Analysis

Electrical safety testing:

ground impedance

Qualification of power supply wires

insulation resistance

(Power cord, power PE, power accessible non grounded conductive component NE, equipment casing, patient application end AP-PE, AP-NE, application terminal pressurization conditions)

GB9706.1 / IEC 60601-1:

leakage current

Shell leakage

Patient leakage

Patient assisted leakage

F-type patient leakage

IEC 61010 /  IEC 62353:

Contact leakage

Equipment leakage (direct, differential, and substitute)

Application of three-phase power supply equipment:

Ground conduction resistance

Insulation properties of materials

Leakage current (equipment leakage, shell leakage, patient leakage)

You can use the above variables or combinations to create custom tests.

Technical Specifications:

Grounding continuity

test method

2-wire pre pulse technology, using wire zero elimination impedance correction technology

Pre pulse technology

25A peak current, eliminating thin film resistance

test current

>200mA

Current loading time

6 optional values: 2 seconds/5 seconds/10 seconds/20 seconds/30 seconds/duration

test voltage

Effective value of 4-24V (IEC60601 test specifies 6V)

Measurement range (small range)

0.001 - 0.999Ω / 分辨率 0.001Ω

Measurement range (medium range)

1.00 - 9.99Ω / 分辨率 0.01Ω

Measurement range (large range)

10.0 - 19.9Ω / 分辨率 0.1Ω

precision

Reading ± 2%+10 m Ω

insulation resistance

measurement items

5 optional values:

The power supply end of the test object - the shell part that fails to protect the grounding (Class I), the power supply end of the test object - the shell part (Class II), the equipment application part - the shell part that fails to protect the grounding (Class I, with grounding), the equipment application part - the shell part that fails to protect the grounding (Class II), the power supply end of the test object - the equipment application part

Loading voltage

4 optional values:

50 / 100 / 250 / 500 VDC @1mA

Voltage loading time

6 optional values:

2 seconds/5 seconds/10 seconds/20 seconds/30 seconds/duration

Test limit conditions


13 optional values:

0.01/0.1/0.3/1.0/2.0/4.0/5.0/7.0/10.0/20.0/50.0/70.0/100.0MΩ

Measurement range (small range) 50V DC

Measurement range (small range) above 50V DC

Accuracy (small range)


0.01MΩ-10MΩ

0.01MΩ-20MΩ

Reading ± 1%+2 counts

Measurement range (large range) 250V DC

Measurement range (large range) 500V DC

Accuracy (large range)


20MΩ-50MΩ

20MΩ-100MΩ

Reading ± 3%+2 counts


resolution

0.01MΩ

Direct method for measuring equipment leakage current

(IEC60601/GB9706.1&IEC62353/YYT 0841 Medical Electrical Equipment Periodic Testing and Post Repair Testing)

measurement range

4 ~ 9999μA

precision

Reading ± 3%+2 counts

Loading power supply voltage on the application end

Only F-type devices can be loaded with 110% grid power

Measurement device

According to the requirements of IEC 60601-1

measurement type

Compliant with IEC 60601 measurement regulations for patient (or auxiliary) leakage current, providing true RMS values for both AC and DC measurements

loading time


6 optional values:

2 seconds/5 seconds/10 seconds/20 seconds/30 seconds/duration

Limit conditions for evaluation results


10 optional values:

10/ 25/ 50/ 100/ 300/ 500/1000/2500/5000/10000uA

Application Part AP

1-10 lead independent editing

Anti electric shock level

Class I / Class II

Differential method for measuring equipment leakage current (IEC62353&YYT 0841-2011 Medical electrical equipment Periodic testing and post repair testing of medical electrical equipment)

measurement range

75μA to 9999μA

precision

Reading ± 3%+5 counts

Measurement/Display

Resolution 1 μ A

measurement type

True effective value

Measurement device

Frequency response complies with IEC 60601-1 regulations

loading time


6 optional values:

2 seconds/5 seconds/10 seconds/20 seconds/30 seconds/duration

Limit conditions for evaluation results


10 optional values:

10/ 25/ 50/ 100/ 300/ 500/1000/2500/5000/10000uA

Alternative method for measuring equipment leakage current (IEC62353&YYT 0841-2011 Medical electrical equipment Periodic testing and post repair testing of medical electrical equipment)

test voltage

250V power frequency

test current

3.5mA current limiting

measurement range

4μA - 9999μA

measurement resolution

1μA

measurement accuracy

Reading ± 3%+2 counts

measurement type

True effective value

Measurement device

According to IEC 60601-1

power measurement

method

apparent power

scope

0.1kVA-4kVA

precision

± 5%+2 counts

loading time

6 optional values:

2 seconds/5 seconds/10 seconds/20 seconds/30 seconds/duration

Network power output test

Input voltage range

0-300VAC

Can input current

16A

precision

Reading ± 5%+2 counts

Pluggable IEC network power cord test

Test time

2 seconds

test

Conductivity, insulation, and polarity changes

Conductivity limit

Automatic calculation of conductivity limit using wire gauge calculation table


15 optional values:

0.01/0.02/0.03/0.04/0.05/0.06/0.07/0.08/0.10/0.20/0.30/0.50/1.0/2.0/5.0Ω

Insulation limit

13 optional values:

0.01/0.1/0.3/1.0/2.0/4.0/5.0/7.0/10.0/20.0/50.0/70.0/100.0MΩ

Loading voltage conditions

4 optional values:

50 / 100 / 250 / 500 VDC @1mA

Test polarity

auto reverse

General Specifications

main power supply

230V AC ±10%, 50-60Hz +/- 1Hz

120V AC ± 10%, 60Hz+/-1Hz (American model)

battery

6 x 1.5V AA

weight

1.6 kilograms (including battery)

size

270 x 110 x 75 mm

Operating conditions

0℃ - 40℃, 0-90% RH - NC

storage environment

-15℃ - +60℃ / 5℃ - 140 ℉

Environmental protection

IP 40


Recommended configuration:1 set of 288 type electrical safety tester host, 1 set of 4mm buckle type pin type dual-use terminal posts for ECG leads (10 pieces), 1 detachable 2-meter IEC power cord, 1 application terminal test connection module, 1 standard test probe, fixture and cable, 1 grounding test wide fixture, verification test kit (including 1 pair of thin head red and black probe, 1 thin head crocodile clip, 1 pair of 2-meter test cable, 1 0.2 Ω low resistance verification box, 1 10M Ω high resistance verification box, 1 surface insulation defect detection brush (optional accessory), 1 tool package), 1 Chinese quick use guide, 1 Chinese user manual, factory test data report 1 set, 1 one handed portable bag, 1 portable bag