Application scope:
Compliant with the safety requirements of GB9706.4, capable of verifying and calibrating electric power output devices. Unlike the traditional method of measuring current through resistance voltage division and calculation, Uni Therm uses high-resolution current transformers for direct current measurement. Therefore, Covidien FT10 energy workstation chose Uni Therm for its calibration work. At the same time, through the user customized and edited automatic testing program, it can safely guide users to complete the entire testing process automatically, saving more time. Continuous measurement and single measurement mode can also be performed.

The main advantages and functions include an integrated onboard memory manager (with built-in electronic documents such as test circuits, the device can directly input the required automated test program through the keyboard, and can store test results on site), graphical color LCD screen prompts for operation steps, built-in user edited automated test programs, direct connection to energy output ports without the need to prepare consumables such as test pens and negative plates, comprehensive test result judgment, data storage and export functions.

The load resistor with high power and low inductance characteristics can be freely set from 0 to 1 with a current of up to 8 amperes and a duty cycle, ensuring that the high-power energy output device can continue to operate during calibration and performance testing, avoiding overheating protection that may cause it to malfunction and stop working. Suitable for calibrating high current vascular closure working mode.
The device provides all necessary resistors within a single chassis, and the non inductive load box has over 1000 levels of resolution within the testing range of 0-5115 Ω, providing high detail power curve analysis.
The testing circuit of the negative electrode contact monitor can be executed automatically or manually. High precision potentiometers can provide a 1 Ω resolution variation, simulate changes in the impedance of the negative electrode plate in contact with the human body, and test whether the alarm for poor contact of the electrode plate has failed.
Uni Therm has excellent operability, and inserting a USB keyboard can easily work with the internal memory of the machine to quickly edit test programs. Bluetooth technology allows wireless connection between PCs and direct reading printing devices, enabling quick downloading of test data and uploading of specific testing plans from manufacturers.
By using the built-in oscilloscope output, the output waveform can be checked, making it easy to confirm the desired waveform shape.
☆ Three independent resistive loads (CQM for electrode contact, high-frequency leakage current, energy output)
☆ Compatible with all types of high-frequency electric knives, including electrosurgical equipment using high current vascular closure technology
☆ Meet all current requirements for electrode contact testing without the need for additional equipment purchase
☆ Built in electronic imaging, electrical connection user manual
☆ By executing automated testing sequences, you can quickly complete your personalized testing
☆ Automatically save results, avoid manual recording errors, and provide work efficiency
☆ Accelerate the testing process, save your valuable time, and improve testing security
Electrical testing function |
application |
High frequency leakage current |
Routine testing and calibration of high-frequency electric knife oscillation output device, production line testing |
Output load test |
Development tool for high-frequency electric knife research and development |
Power distribution curve |
Formal testing and evaluation tools for electrosurgical equipment |
Patient Polar Circuit Monitoring Alarm Test (CQM) |
Teaching tools, providing training on quality control procedures for electrosurgical equipment |
Rigel Uni Therm Electric Knife Analyzer Product Introduction:
Performance testing tool for output power, high-frequency leakage current, and electrode contact monitor of electrosurgical instruments
power measurement |
The true effective value of the applied waveform |
active power |
0-500W (true effective value) |
precision |
When>50w, ± (5% of the reading) When<50w, ± (5% of reading+1w) |
duty cycle |
0-100% duty cycle, opening time Z up to 60 seconds |
Variable resistance box |
0 - 5115Ω |
Step value |
5Ω |
precision |
± (1% set value+0.5 Ω) |
Load resistance type |
Non inductive ceramic resistor |
peak voltage |
0-10KV (peak to peak) - Closed loop load only |
precision |
± (10% of reading+15V) - Measurements are only taken between the excitation electrode and the dispersion electrode under closed load |
Effective voltage |
0-700V (true RMS) |
precision |
± (10% of reading+5V) |
High frequency current (true effective value) |
0-6000mA built-in variable load 0-8000mA external non inductive load |
precision |
± (2% of reading+1mA) |
quality factor |
1.4-20 (peak voltage/true RMS voltage) Use the higher of the two peak voltage measurements. |
Effective bandwidth | |
True RMS bandwidth (instrument only) |
30 Hz to 10MHz (-3 dB) |
True effective bandwidth value (built-in load) |
30 Hz to 5.0 MHz (-3 dB) |
Variable load |
0 - 5115Ω |
resolution |
5Ω |
precision |
± (1% set value+0.5 Ω) |
Load resistance type |
Non inductive ceramic resistor |
Measurement delay |
The foot switch delay can be selected between 200-5000ms (10ms resolution) |
High frequency leakage current | |
High frequency current (true effective value) |
0-6000mA built-in variable load 0-8000mA external non inductive load |
precision |
± (2% of reading+1mA) |
Motivational Pole |
From surgical electrode to ground |
Diffuse pole |
From negative plate to ground |
load |
Variable load, please refer to load/power curve measurement 2 x 200 Ω fixed load |
precision |
±1% + 0.5Ω |
Neutral electrode contact safety monitoring (CQM-REM) | |
range |
0-475 Ω, step @ 1 Ω, variable potentiometer driven by stepper motor |
precision |
±5% |
Alarm register |
High and low, manually confirm |
adjustment method |
manual or automatic |
Output connector | |
Remote foot switch control (CUT) for electric cutting mode |
2 x 4 mm - yellow, single relay contact |
Remote foot switch control (COAG) for electrocoagulation mode |
2 x 4 mm - Blue, single relay contact |
Electrode contact safety impedance (REM) 0-475 ohm |
2 x 4 mm - Black |
Variable load 0-5115 Ω |
2 x 4mm red=excitation electrode, black=dispersion electrode |
Fixed load 2 x 200 Ω |
1 x 4 mm black (common end), 2 x 4 mm red |
Device under test (MD) |
2 x 4mm white wire |
USB A 端口 |
Keyboard interface |
USB B port |
PC interface |
Oscilloscope output port |
0.5V/A, 100mA RF current Z small input, uncalibrated, only used for display purposes |
isolation | |
Isolation (grounding to measuring equipment) |
10kV |
Low frequency filter |
|
Interference frequency filter |
100 Hz high pass filter |
General Specifications | |
Data storage and memory calling |
4Mb, Approximately 5000 pieces of data |
Data output format |
CSV format and SSS format Test sequence, results, reference power diagram |
Overall dimensions |
370 X 300 X 204mm |
weight |
11 Kg |
Operating Temperature |
10 ˚C - 40 ˚C |
Storage temperature |
0˚C - 50 ˚C |
main power supply |
115/230 ±10% VAC; 48 - 66 Hz, 35 VA |
fuse |
2 x 1.6 A (T) ceramic slow melting fuse |