productDetails
1Overview
thatinstrumentUsing a new high-performance catalyst, the hydrocarbon removal effect is better, the speed is faster, and the operating temperature is only350 ℃ makes instrument use safer and more energy-efficient.There are pressure regulating devices and gases in the zero gas generatorpurifyEquipment, reaction chamber, and temperature controller.It uses oil-free compressors or air cylinders as the gas source, and obtains clean air through catalytic conversion, adsorption and filtrationThe zero gas generator generates a gas that does not contain the measured pollutant gas (such as:SO2,CO,O3,NO,NO2AndHCWaiting for zero gas.
PR-2LThe zero gas generator can provide sufficient and stable zero gas for various automatic monitoring systems.
IISchematic diagram of appearance structure

IIISchematic diagram
After being compressed and dried, the air enters the hydrocarbon removal device, and then enters the activated carbon for preliminary adsorption of organic pollutants, and then enters the system filled with precious metal catalystsreaction chamberPerforming oxidation, the main reaction chambermake, do, create, compose, write, etc.It is used to remove substances from the airCOOxidation intoCO2,HCAnd it oxidizes into water andCO2After removal, the pure air enters the three-stage purification system equipped with molecular sieves99.99%The particle size0.5umAfter the formation of gas particles, dry and pure zero level air is eventually formed.
4Online installation
1. Remove the instrument from the packaging box, first check the appearance and interior for any serious damage caused by transportation, and then verify whether the instruction manual and accessories are complete.
2. Use an outer diameter of Φ 3pipelineConnect the instrument with the gas sourceconnectAnd check for any air leaks. Turn on the gas source switch, and the pressure on the instrument gauge should be consistent with the pressure on the gas source gauge.When the pressure is greater than0.2MPaFirst, close the input port, then quickly close the output port,2No significant change in hourly pressure indicates no air leakage.And carefully inspect all pipelines to confirm that there are no leaks
5Operation steps
1.Turn on the gas source switch and confirm that there is no leakage.
2.Turn on the power of the hydrocarbon remover and set the control temperature to350℃,Start warming up(Please refer to the instructions for using the temperature control table in the appendix for details).After one hour of stable startup, the gas can be used for testing normally.
3. After use, turn off the power and cool down for 30 minutes, then turn off the gas source switch.
6、 Technical parameters
aInput concentration of hydrocarbons (e.g.)<100ppm
b.Hydrocarbon output concentration (e.g.)<0.1ppm
cInput gas pressure0~0.5MPa
dInput gas flow rate0~1000ml/min
eTemperature control range of high temperature furnaceroom temperature~600℃
7、 Application Description
aInstrument stability time<1h
bUsing ambient temperature0-40℃
c.Environmental humidity for use<85%
dWorking power supply220V/50Hz
e.Exterior dimensions long*wide*High370mm*485mm*225mm
8、 Precautions
aRegularly check the gas source silica gel, molecular sieve, and activated carbon. If there is any failure (based on silica gel discoloration), replace the packing or activate it in a timely manner. When replacing, please cut off the power first and wait for the pressure to reach zero before proceeding with the operation.
bWhen replacing the gas source or packing, a leak test is required. (As long as there are changes in the pneumatic system, a leak test must be conducted when starting up)
c.Regularly check the hydrocarbon removal effect of the instrument. If the concentration of purified air exceeds the standard, replace the packing or reactivate it.
dWhen the instrument is idle, the gas input and output ports need to be usedBlind hole pressure cap seal.
9random spare parts
1. Connecting pipelineΦ 3 5 meters
2. fuse10A2a
3. power cord1article
4. Instruction Manual1copy
5. Certificate of Conformity1Zhang
10、 Quality Assurance
This product comes with a one-year warranty from the date of sale,Free maintenance during the warranty period (excluding consumables),Lifetime access to maintenance services.
Customers can communicate through phoneE-mailPlease contact the sales or technical department at any time by fax or other means. If the matter is urgent, both parties will arrive on time as agreed. After receiving the information, our company willRespond within 2 hours, if necessary, within 24 hours within the cityArrive at the site for repair within hours (depending on the actual situation in other regions) until the fault is resolved. (The costs incurred due to instrument quality issues during the warranty period shall be borne by us, and other costs shall be borne by the user.)