Phone
13701307156
-
Address
225, 2nd Floor, Qingshang, Qinghe Xiaoying, Haidian District, Beijing
Beijing Purui Analytical Instrument Co., Ltd
13701307156
225, 2nd Floor, Qingshang, Qinghe Xiaoying, Haidian District, Beijing

instrumentfeature
1. adopt2.4Industrial grade color screen with high-definition display, meeting the requirements of users for clear viewing of instrument data during night work;
2. Simultaneously supporting multiple signal outputs:4~20mATheRS485Real time communication with the computer, two sets of alarm signals output, which can facilitate communication withDCSThePLCIntegrate real-time data analysis and security protection systems with various control devices.
3. Coordinate curve drawing detection mode;
4. Real time analysis and display of data * small value, average value, * value;
5. Plug and play original imported international brand sensors, easy to maintain on site;
6. High precision, extremely strong stability, fast response speed, and alarm without delay;
7. Built in temperature detection, full range support for temperature compensation function;
8. The Chinese menu supports remote control operation without opening the cover;
9. Design of intrinsically safe circuits and explosion-proof enclosures, rainproof, corrosion-resistant, and wear-resistant;
10. Adopting circuit protection design to effectively prevent instrument damage caused by various human factors and uncontrollable factors, on-site maintenance is safe, convenient, and fast;
11. Supports various installation methods such as wall mounted, pipeline mounted, flow-through, and pump suction to meet the environmental requirements of users in different occasions
Technical Specifications
| gas detection | Volatile Organic Compounds |
| quantity Cheng | 0-500,1000,2000,5000,6000ppm |
| divide distinguish rate | 1——0.001Depending on the specific sensor |
| Detection Principle | PIDPrinciple of photoionization detection |
| essence degree | <±2%F.S |
| Detection Time | <10S(T90) |
| recovery time | <20S |
| linear error | ≤±1% |
| zero drift | ≤±1%(F.S/Year) |
| heavy again sex | ≤±1% |
| Detection method | Fixed, online detection, wall mounted, pipeline, pump suction, circulation and other installation methods are optional |
| Installation specifications | pipeline-styleM40*1.5MMinternal thread ; Outer diameter of flow type air nozzle7mminner diameter5mmlength10mm |
| Detection Mode | Two detection modes: regular and drawing |
| concentration unit | ppmThe%VOLThe%LELThemg/m3optional |
| signal output | 4-20mASignal: Standard16Positional accuracy4-20mAOutput chip, transmission distance1Km RS485Signal: Original importedRS485Communication chip, using standardMODBUS RTUProtocol, transmission distance2Km Voltage signal:0-5VThe0-10VOutput, customizable Pulse signal: also known as frequency signal, with adjustable frequency range (optional) Switching signal: standard configuration2Group relay, passive contact of relay, capacity220VAC 3A/24VDC 3A |
| connected device | Computer host, alarm control hostDCSThePLCExternal control and controlled equipment such as secondary instruments |
| electrical interface | 1/2″NPTinternal thread |
| working environment | 0 ~ 95% RHnon-condensing temperature-40~70℃ |
| Explosion-proof rating | Exia II CT6 |
| Protection level | IP65 |
| Overall dimensions | 240×152×95mm(L×W×H) |
| Working Voltage | 24VDC(10~30VDC) |
| Total weight | 1.8KG |
| match document | Complete machine, manual, remote control, certificate of conformity, (air nozzle, hose, air pump, sound and light alarm) optional |
Updated on September 2, 2020 at 11:59:14