Welcome Customer !

Membership

Help

Xi'an Shengli Instrument Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Xi'an Shengli Instrument Co., Ltd

  • E-mail

  • Phone

  • Address

    25th Floor, Building G, Wangzuo Modern City, No. 35 Tangyan Road, High tech Zone, Xi'an City, Shaanxi Province

Contact Now

VICTOR 72B Signal Generator

NegotiableUpdate on 05/14
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

VICTOR 72B signal generator is a portable handheld signal generator powered by lithium batteries, which can be used to measure DC voltage, DC current, on-off, and simulate output DC voltage, DC current, thermocouple, and frequency. Suitable for debugging and calibration of on-site transmitters, DCS, PLC and other systems. VICTOR 72B has been widely used in fields such as petrochemicals, biopharmaceuticals, industrial automation, universities, and research institutions. It is a "pocket level" practical tool for engineers, technical workers, and researchers.

Product Details

VICTOR 72B signal generator is a portable handheld signal generator powered by lithium batteries, which can be used to measure DC voltage, DC current, on-off, and simulate output DC voltage, DC current, thermocouple, and frequency. Suitable for debugging and calibration of on-site transmitters, DCS, PLC and other systems. VICTOR 72B has been widely used in fields such as petrochemicals, biopharmaceuticals, industrial automation, universities, and research institutions. It is a "pocket level" practical tool for engineers, technical workers, and researchers.

VICTOR 72B Signal GeneratorKey Features:
1. Measurement function:
*DC voltage, DC current, on-off, thermocouple (only available in 72B)
*Basic accuracy: 0.1%
*Loop detection function, internal 24V circuit power supply, and simultaneous measurement of current.
2. Output function:
*DC voltage, DC current, frequency, and thermocouple (only available in 72B)
*Basic accuracy: 0.1%.
*Preset value setting and quantity conversion mode
*DCI output includes active output (internal 24V power supply) and passive output (external 24V power supply).
*When outputting DCI, manual stepping and automatic waveform output functions can be provided.
*The manual step output includes: 25% step and 100% step.
The output current can be selected from 0~20mA or 4~20mA.
*The automatic wave output includes: automatic stepping, sawtooth wave, triangular wave, and the working parameters of the automatic waveform can be configured.
3. Other functions:
*Display: 2.4-inch TFT color screen
*Battery specification: 3.7V/3000mAh
*TYPE-C interface: rechargeable, communicable (communication protocol available)
*Support language switching between Chinese and English

Technical Indicators:
Input measurement function [applicable within one year after calibration, 23 ℃± 5 ℃, 20-70% RH, accuracy=± (% set value+word)]

measurement function range measurement range resolution accuracy remark
DC voltage 100mV -10.0mV~110.0mV 100uV 0.1%+4
DC voltage 30V -5.00V~30.00V 10mV 0.1%+4
direct current 20mA 0.00mA~22.00mA 0.01mA 0.1%+4
Circuit power supply
LOOP
24V
±10%
on-off
OPEN/CLOSE

thermocouple R、S、B、K、E、J、T、N 1℃ 0.1%+4 Only 72B
DCV characteristics: Input impedance: 30V: 300K Ω (nominal value) mV:>1M Ω (nominal value)
DCI feature: Overload protection: 50mA/30V; Load voltage: approximately 18mV/mA
TC Explanation: Thermocouple measurement uses the ITS-90 temperature scale, and the accuracy does not include errors in cold junction compensation or the influence of thermoelectric potential.

Analog output function [applicable within one year after calibration, 23 ℃± 5 ℃, 20-70% RH, accuracy=± (% set value+word)]

output function range Output setting range resolution accuracy remark
DC voltage 100mV 0.0mV~110.0mV 100uV 0.1%+4
20V 0.00V~24.00V 10mV 0.1%+4
direct current
(Active)
20mA 0.00~22.00mA 0.01mA 0.1%+4
direct current
(Passive)
-20mA 0.00~-22.00mA 0.01mA 0.1%+4
frequency 100Hz、1kHz、5kHz、10kHz、100kHz 1.0Hz~100kHz 0.1Hz 0.1%+2 Only 72B
thermocouple R、S、B、K、E、J、T、N 1℃ 0.1%+4
DCI feature: 20mA maximum load 1K Ω
FREQ characteristics: 60% duty cycle of rectangular wave, 1-20Vp-p
TC Explanation: Thermocouple measurement uses the ITS-90 temperature scale, and the accuracy does not include errors in cold junction compensation or the influence of thermoelectric potential.