VICTOR 02S temperature calibrator

NegotiableUpdate on 05/14
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

VICTOR 02S temperature calibrator (hereinafter referred to as calibrator) is a handheld portable instrument powered by batteries that can measure and output electrical and physical parameters. Can be used to measure DC voltage, resistance, thermocouple, and thermal resistance; Simulate output DC voltage, resistance, thermocouple, and thermistor.

Product Details

VICTOR 02S temperature calibrator(hereinafter referred to as the calibrator) is a handheld portable instrument powered by batteries that can measure and output electrical and physical parameters. Can be used to measure DC voltage, resistance, thermocouple, and thermal resistance; Simulate output DC voltage, resistance, thermocouple, and thermistor.

Basic characteristics of the product
1. Input measurement characteristics

*Basic functions: DC voltage DCV, resistance OHM, thermocouple TC, thermal resistance RTD, on/off
*The highest measurement accuracy is 0.05%, and the measurement display is 5 digits.
*Measurement rate: 2-3 times per second.
*OHM and RTD can perform 2W/3W/4W measurements.
*The measured temperature value and corresponding temperature signal are displayed on the same screen.

2. Analog output characteristics

*Basic functions: DC voltage DCV, resistance OHM, thermocouple TC, thermal resistance RTD
*The highest output accuracy is 0.05%, and the output is displayed as 5 bits.
*The output setting can be manually adjusted by bit.
*There are two settings for OHM excitation current: H/L.
*The output temperature value and corresponding temperature signal are displayed on the same screen.

3. Built in high-precision temperature sensor with high-precision cold end compensation.
The compensation methods for the cold end of the thermocouple can be set, including manual compensation, automatic compensation, compensation connection disconnection, etc.
4. Maintain the settings function, including:
Automatic shutdown, backlight, flashlight, buzzer, etc
5. Other

*Segment code LCD screen display, white backlight.
*LED flashlight illumination.
*Powered by 3 No. 5 alkaline batteries, the battery compartment door allows for easy battery replacement and low battery detection.
*Compact, sturdy, reliable, suitable for on-site use, and affordable.
*User selectable high suction magnetic strap.

1VICTOR 02S temperature calibratorBasic functions and performance (technical indicators) of the product

Input measurement function[Applicable toWithin one year after calibration23±52070%RHAccuracy%set value+Character)]

measurement function

range

measurement range

resolution

accuracy

remark

DC voltage

DCV

100mV

-110.000mV~110.000mV

0.01mV

0.05%+0.03mV

Input resistance: approximately1MΩ

1V

-1.1000V~1.1000V

0.0001V

0.05%+0.3mV

30V

-30.000V30.000V

0.001V

0.02%+2mV

resistance

OHM

500Ω

0~550.00Ω

0.01Ω

0.05%+0.2Ω

2W/3W/4Wmeasurement

500Ωabout1mAmotivation

5KΩabout0.1mAmotivation

Open circuit voltage: approximately2.5V

The measurement accuracy does not include lead resistance

5KΩ

0~5.5000KΩ

0.0001KΩ

0.05%+2Ω

thermocouple

TC

R

0°C1767°C

1°C

0.05%+3(Less than or equal to)100°C

0.05%+2(Greater than)100°C

adoptITS-90Temperature scale

Accuracy does not include errors in cold end compensation

Accuracy does not include the inaccuracy of sensors

Accuracy does not include the influence of thermoelectric potential

The material corresponding to the left division mark is as follows:

R: platinum-rhodium13-platinum

S: platinum-rhodium10-platinum

K: nickel-chromium-Nickel silicon

E: nickel-chromium-Copper nickel (constantan)

J: iron-Copper nickel (constantan)

T: copper-Copper nickel (constantan)

N: Nickel chromium silicon-Nickel silicon

B: platinum-rhodium30-platinum

A: Tungsten rhenium 5 -Tungsten rhenium 20

C: Tungsten rhenium 5 -Tungsten rhenium 26

D: Tungsten rhenium 3 -Tungsten rhenium 25

S

0°C1767°C

K

-100.0°1372.0°C

0.1°C

0.05%+2(Less than or equal to)-100°C

0.05%+10(Greater than)-100°C

E

-50.0°C1000.0°C

J

-60.0°C1200.0°C

T

-100.0°C400.0°C

N

-200.0°1300.0°C

B

600°C1820°C

1°C

0.05%+3

A

0°C2500°C

1°C

0.05%+3

C

0°C2310°C

1°C

D

0°C2310°C

1°C

thermal resistor

RTD

3W

Pt100

385

-200.0°C800.0°C

0.1°C

0.05%+0.6

adoptPt385)Temperature scale

2W/3W/4Wmeasurement

500Ωabout1mAmotivation

5KΩabout0.1mAmotivation

Open circuit voltage: approximately2.5V

Accuracy does not include2 W/3 W/4WError caused by mismatched lead resistance measurement

Accuracy does not include the inaccuracy of sensors

Accuracy does not include the influence of thermoelectric potential

Pt200

385

-200.0°C630.0°C

Pt500

385

-200.0°C630.0°C

Pt1000

385

-200.0°C630.0°C

Cu50

-50.0°C150.0°C

on-off

detection

500Ω

≤50Ωvocalize

0.01Ω


about1mAmotivation

Analog output function[Applicable toWithin one year after calibration23±52070%RHAccuracy%set value+Character)]

output function

range

output range

resolution

accuracy

remark

DC voltage

DCV

100mV

-10.000mV~110.00mV

0.01mV

0.05%+0.03mV

Maximum output current:0.5mA

1V

-0.1000V~1.1000V

0.0001V

0.05%+0.3mV

Maximum output current2mA

ohm

OHM

400Ω

0~400.0Ω

0.1Ω

0.05%+0.2Ω

The excitation current is±0.1±5mA

Maximum output voltage2V

The excitation current is±0.10.5mAWhen, add0.3ΩAdditional error

Accuracy does not include lead resistance

4KΩ

0~4.000KΩ

0.05%+2Ω

thermocouple

TC

R

0°C1767°C

1°C

0.05%+3(Less than or equal to)100°C

0.05%+2(Greater than)100°C

adoptITS-90Temperature scale

Accuracy does not include errors in cold end compensation

Accuracy does not include the inaccuracy of sensors

Accuracy does not include the influence of thermoelectric potential

The material corresponding to the left division mark is as follows:

R: platinum-rhodium13-platinum

S: platinum-rhodium10-platinum

K: nickel-chromium-Nickel silicon

E: nickel-chromium-Copper nickel (constantan)

J: iron-Copper nickel (constantan)

T: copper-Copper nickel (constantan)

N: Nickel chromium silicon-Nickel silicon

B: platinum-rhodium30-platinum

A: Tungsten rhenium 5 -Tungsten rhenium 20

C: Tungsten rhenium 5 -Tungsten rhenium 26

D: Tungsten rhenium 3 -Tungsten rhenium 25

S

0°C1767°C

K

-200.0°C1372.0°C

0.1°C

0.05%+2(Less than or equal to)-100°C

0.05%+1(Greater than)-100°C

E

-200.0°C1000.0°C

J

-200.0°C1200.0°C

T

-250.0°C400.0°C

N

-200.0°C1300.0°C

B

600°C1820°C

1°C

0.05%+3

A

0°C2500°C

1°C

C

0°C2310°C

1°C

D

0°C2310°C

1°C

thermal resistor

RTD

Pt100

385

-200.0°C800.0°C

0.1°C

0.05%+0.6

adoptPt385)Temperature scale

The excitation current is±0.1±5mA

Maximum output voltage2V

The excitation current is±0.10.5mAWhen, add0.3ΩAdditional error

Accuracy does not include lead resistance

Accuracy does not include the influence of thermoelectric potential

Pt200

385

-200.0°C630.0°C

0.1°C

Pt500

385

-200.0°C630.0°C

0.1°C

Pt1000

385

-200.0°C630.0°C

0.1°C

Cu50

-50.0°C150.0°C

0.1°C

IIGeneral characteristics of the product

l Working temperature and humidity:0 to 50ºC ≤80%RHNo condensation;40 to 50ºC ≤70%RH.

l Storage temperature and humidity:-25 to 60ºC ≤90%RHNo condensation.

l Electrical safety:EN61010-12001

Voltage endurance:Between the terminal and the casingAC3540V(50/60Hz)5second.

Insulation impedance:Between the terminal and the casingDC1000V /100MΩOr larger.

l Electromagnetic compatibility(EMC):EN61326-12006

Meets performance criteria2Temporary reduction or loss of functionality and performance, but capable of self recovery.

l Protection levelIP40.

l Vibration and Falling:IEC 60068-2-642008TheIEC 60068-2-322008

Random,2g5500Hz1Rice drop test.

l CEauthentication.

l Quality standards: in accordance withTUV ISO 9001Development, design, and manufacturing.

l Calibration cycle and preheating time:

To ensure the accuracy of this table, its calibration cycle is one year

The startup preheating time is greater than30minute

l Instrument display and buttons

segment codeLCDDisplay, white backlight

Adopting silicone rubber buttons

l Instrument power supply

3section5Powered by alkaline batteries

l Dimensions and Weight