VICTOR 24Pro Multi functional Process Calibration InstrumentIt is a multifunctional, high-precision, and high stability on-site process calibrator that provides * measurement and output functions, almost meeting all calibration work in process control sites.
| measurement function | range | measurement range | resolution | accuracy | Remarks | ||
| DC voltage DCV |
50mV | -55.000~55.000mV | 1μV | 0.01+0.01 | Input resistance: approximately 100M Ω (50mV/500mV) About 1M Ω (5V/50V) Input current less than 1nA |
||
| 500mV | -550.00~550.00mV | 10μV | 0.01+0.01 | ||||
| 5V | -5.5000~5.5000V | 0.1mV | 0.01+0.01 | ||||
| 30V | -35.000~35.000V | 1mV | 0.01+0.01 | ||||
| direct current DCI |
50mA | -55.000m~55.000mA | 1μA | 0.01+0.01 | Shunt resistance: approximately 10 Ω Input resistance: approximately 20 Ω |
||
| Resistance OHM | 500Ω | 0.00~550.00Ω | 0.01Ω | 0.01+0.01 | 2W/3W/4W measurement 500 Ω approximately 1mA excitation 5K Ω approximately 0.1mA excitation Open circuit voltage: approximately 2.5V The measurement accuracy does not include lead resistance |
||
| 5KΩ | 0.0000~5.5000KΩ | 0.1Ω | 0.01+0.01 | ||||
| thermocouple TC |
R | 0°C~1767°C | 1°C | 0~500℃ 500~1767℃ |
1.8℃ 1.5℃ |
Accuracy does not include errors in cold end compensation Accuracy does not include the inaccuracy of sensors Accuracy does not include the influence of thermoelectric potential |
|
| S | 0°C~1767°C | ||||||
| K | -100.0°~1372.0°C | 0.1°C | -100.0~0.0℃ 0.0~1372.0℃ |
1.2℃ 0.8℃ |
|||
| E | -50.0°C~1000.0°C | -50.0℃~0.0℃ 0.0~1000.0℃ |
0.9℃ 1.5℃ |
||||
| J | -60.0°C~1200.0°C | -60.0~0.0℃ 0.0~1200.0℃ |
1.0℃ 0.7℃ |
||||
| T | -100.0°C~400.0°C | -100.0~0.0℃ 0.0~400.0℃ |
1.0℃ 0.7℃ |
||||
| N | -200.0°~1300.0°C | -200.0~0.0℃ 0.0~1300.0℃ |
1.5℃ 0.9℃ |
||||
| B | 600°C~1820°C | 1°C | 600~800℃ 800~1000℃ 1000~1820℃ |
2.2℃ 1.8℃ 1.4℃ |
|||
| L | -60.0°C~900.0°C | 0.1°C | -60.0~0.0℃ 0.0~900.0℃ |
0.7℃ 0.5℃ |
|||
| U | -100.0°C~600.0°C | -100.0~0.0℃ 0.0~600.0℃ |
0.7℃ 0.5℃ |
||||
| A | 0°C~2500°C | 1°C | 0~1600℃ 1600~2000℃ 2000~2500℃ |
2.0℃ 2.2℃ 2.4℃ |
|||
| C | 0°C~2310°C | 0~1600℃ 1600~2000℃ 2000~2310℃ |
2.0℃ 2.2℃ 2.4℃ |
||||
| D | 0°C~2310°C | 0~100℃ 100~270℃ 270~1200℃ 270~2310℃ |
2.6℃ 2.4℃ 2.2℃ 2.4℃ |
||||
| G | 100°C~2315°C | 0~300℃ 300~1500℃ 1500~2315℃ |
1.6℃ 1.0℃ 2.0℃ |
||||
| P | 0°C~1395°C | 0~1000℃ 1000~1395℃ |
1.6℃ 2.0℃ |
||||
| M | -50.0°C~1410.0°C | 0.1°C | -50.0°C~100.0°C 100.0°C~1000.0°C 1000.0°C~1410.0°C |
1.0℃ 1.6℃ 2.0℃ |
|||
| thermal resistor RTD |
PT100 385 |
-200.0°C~800.0°C | 0.1°C | -200.0~0.0℃ 0.0~400.0℃ 400.0~800.0℃ |
0.5℃ 0.7℃ 0.8℃ |
2W/3W/4W measurement 500 Ω approximately 1mA excitation 5K Ω approximately 0.1mA excitation Open circuit voltage: approximately 2.5V The accuracy does not include the error caused by the mismatch of 2 W/3 W measurement lead resistance Accuracy does not include the inaccuracy of sensors Accuracy does not include the influence of thermoelectric potential |
|
| PT200 385 |
-200.0°C~630.0°C | -200.0~100.0℃ 100.0~300.0℃ 300.0~630.0℃ |
0.8℃ 0.9℃ 1.0℃ |
||||
| PT500 385 |
-200.0°C~630.0°C | -200.0~100.0℃ 100.0~300.0℃ 300.0~630.0℃ |
0.4℃ 0.5℃ 0.7℃ |
||||
| PT1000 385 |
-200.0°C~630.0°C | -200.0~100.0℃ 100.0~300.0℃ 300.0~630.0℃ |
0.3℃ 0.5℃ 0.7℃ |
||||
| PT100 3916 |
-200.0°C~630.0°C | -200.0~0.0℃ 0.0~300.0℃ 300.0~630.0℃ |
0.5℃ 0.7℃ 0.8℃ |
||||
| PT100 3926 |
-200.0°C~630.0°C | -200.0~0.0℃ 0.0~300.0℃ 300.0~630.0℃ |
0.5℃ 0.7℃ 0.8℃ |
||||
| Ni120 | -80.0°C~260.0°C | -80.0°C~0.0°C 0.0°C~260.0°C |
0.3℃ 0.4℃ |
||||
| Cu10 | -100.0°C~260.0°C | -100.0°C~260.0°C | 1.8℃ | ||||
| Cu50 | -50.0°C~150.0°C | -50.0~150.0℃ | 0.7℃ | ||||
| frequency FREQ |
500Hz | 1.00Hz~500.00Hz | 0.01Hz | ± 2 words | Input coupling: communication Input impedance: greater than 100 K Ω <100 pF Frequency sensitivity: Square wave at least 3V Vp-p 50% duty cycle |
||
| 5KHz | 0.5001KHz~5.0000KHz | 0.1Hz | |||||
| 50KHz | 5.001KHz~50.000KHz | 1Hz | |||||
| CPM | 1~100000 CPM | 1CPM | ±1CPM | ||||
| Circuit power supply LOOP |
24 V | 10% | Short circuit protection Maximum current: 22 mA Maximum input voltage: 30 V DC Output resistance: 250 Ω nominal value |
||||
| output function | range | output range | resolution | accuracy | Remarks | |
| DC voltage DCV |
100mV | -10.000~110.000mV | 1μV | 0.01+0.01 | Maximum output current: 0.5mA Output resistance: XXXm Ω |
|
| 1V | -0.10000~1.10000V | 10μV | 0.01+0.01 | Maximum output current 2mA Output resistance: XXXm Ω |
||
| 10V | -1.0000~11.0000V | 0.1mV | 0.01+0.01 | Maximum output current 5mA Output resistance: XXXm Ω |
||
| direct current DCI |
33mA | 0.000~33.000mA | 1μA | 0.01+0.01 | At 20 mA, the maximum load is 670 Ω resistor When simulating a transmitter, the external circuit provides a power supply of 5-28V |
|
| Resistance OHM | 400Ω | 0.00~400.00Ω | 0.01Ω | 0.01+0.01 | The excitation current is ± 0.1 to ± 5mA Maximum output voltage 2V When the excitation current is ± 0.1-0.5mA, an additional error of 0.1 Ω is added, and the accuracy does not include lead resistance |
|
| 4KΩ | 0.0000~4.0000 KΩ | 0.1Ω | 0.01+0.01 | |||
| thermocouple TC |
R | 0°C~1767°C | 1°C | 0~100℃ 10~1767℃ 0~100℃ 100~1767℃ |
1.5℃ 1.2℃ 1.5℃ 1.2℃ |
Accuracy does not include errors in cold end compensation Accuracy does not include the inaccuracy of sensors Accuracy does not include the influence of thermoelectric potential |
| S | 0°C~1767°C | |||||
| K | -200.0°C~1372.0°C | 0.1°C | -200.0~-100.0℃ -100.0~400.0℃ 400.0~1200.0℃ 1200.0~1372.0℃ |
0.6℃ 0.5℃ 0.7℃ 0.9℃ |
||
| E | -200.0°C~1000.0°C | -200.0~-100.0℃ -100.0~600.0℃ 600.0~1000.0℃ |
0.6℃ 0.5℃ 0.4℃ |
|||
| J | -200.0°C~1200.0°C | -200.0~-100.0℃ -100.0~800.0℃ 800.0~1200.0℃ |
0.6℃ 0.5℃ 0.7℃ |
|||
| T | -250.0°C~400.0°C | -250.0~400.0℃ | 0.6℃ | |||
| N | -200.0°C~1300.0°C | -200.0~-100.0℃ -100.0~900.0℃ 900.0~1300.0℃ |
1.0℃ 0.7℃ 0.8℃ |
|||
| B | 600°C~1820°C | 1°C | 600~800℃ 800~1820℃ |
1.5℃ 1.1℃ |
||
| L | -200.0°C~900.0°C | 0.1°C | -200.0~0.0℃ 0.0~900.0℃ |
0.7℃ 0.5℃ |
||
| U | -200.0°C~600.0°C | -200.0~0.0℃ 0.0~600.0℃ |
0.7℃ 0.5℃ |
|||
| A | 0°C~2500°C | 1°C | 0~1600℃ 1600~2000℃ 2000~2500℃ |
2.0℃ 2.2℃ 2.4℃ |
||
| C | 0°C~2310°C | 0~1600℃ 1600~2000℃ 2000~2310℃ |
2.0℃ 2.2℃ 2.4℃ |
|||
| D | 0°C~2310°C | 0~100℃ 100~270℃ 270~1200℃ 270~2310℃ |
2.6℃ 2.4℃ 2.2℃ 2.4℃ |
|||
| G | 100°C~2315°C | 0~300℃ 300~1500℃ 1500~2315℃ |
1.6℃ 1.0℃ 2.0℃ |
|||
| P | 0°C~1395°C | 0~1000℃ 1000~1395℃ |
1.6℃ 2.0℃ |
|||
| M | -50.0°C~1410.0°C | 0.1°C | -50.0°C~100.0°C 100.0°C~1000.0°C 1000.0°C~1410.0°C |
1.0℃ 1.6℃ 2.0℃ |
||
| thermal resistor RTD |
PT100 385 |
-200.0°C~800.0°C | 0.1°C | -200.0~0.0℃ 0.0~400.0℃ 400.0~800.0℃ |
0.3℃ 0.5℃ 0.8℃ |
The excitation current is ± 0.1 to ± 5mA Maximum output voltage 2V When the excitation current is ± 0.1-0.5mA, add an additional error of 0.1 Ω Accuracy does not include lead resistance Accuracy does not include the influence of thermoelectric potential Supports pulse transmitters and PLCs with pulse times as short as 10ms |
| PT200 385 |
-200.0°C~630.0°C | -200.0~100.0℃ 100.0~300.0℃ 300.0~630.0℃ |
0.8℃ 0.9℃ 1.0℃ |
|||
| PT500 385 |
-200.0°C~630.0°C | -200.0~100.0℃ 100.0~300.0℃ 300.0~630.0℃ |
0.4℃ 0.5℃ 0.7℃ |
|||
| PT1000 385 |
-200.0°C~630.0°C | -200.0~100.0℃ 100.0~300.0℃ 300.0~630.0℃ |
0.2℃ 0.5℃ 0.7℃ |
|||
| PT100 3916 |
-200.0°C~630.0°C | -200.0~0.0℃ 0.0~300.0℃ 300.0~630.0℃ |
0.5℃ 0.7℃ 0.8℃ |
|||
| PT100 3926 |
-200.0°C~630.0°C | -200.0~0.0℃ 0.0~300.0℃ 300.0~630.0℃ |
0.5℃ 0.7℃ 0.8℃ |
|||
| Ni120 | -80.0°C~260.0°C | -80.0°C~0.0°C 0.0°C~260.0°C |
0.3℃ 0.4℃ |
|||
| Cu10 | -10.0°C~250.0°C | -100.0°C~260.0°C | 1.8℃ | |||
| Cu50 | -50.0°C~150.0°C | -50.0~150.0℃ | 0.7℃ | |||
| frequency FREQ |
100Hz | 1.00 Hz ~110.00Hz | 0.01Hz | ± 2 words | The output waveform is a square wave with an adjustable amplitude of 0-11V The above amplitude errors are all ± 5% reading+0.5V Duty cycle: 50% Load:>100K Ω |
|
| 1KHz | 0.100 KHz ~1.100KHz | 1Hz | ± 2 words | |||
| 10KHz | 1.0KHz~11.0KHz | 0. 1KHz | ± 2 words | |||
| 50KHz | 10KHz~50KHz | 2KHz | ± 5 words | |||
| CPM | 1~1200 CPM | 1CPM | ±1CPM | |||
3、 General characteristics of the product
*Working temperature and humidity: 0 to 50 º C ≤ 80% RH, no condensation; 40 to 50 º C ≤ 70% RH
*Storage temperature and humidity: -25 to 60 º C ≤ 90% RH, no condensation.
*Electrical safety: EN61010-1:2001
Voltage endurance: Measure AC350V/1 minute between the terminal and output terminal.
Insulation impedance: Measure the DC500V/50M Ω or greater between the terminal and the output terminal.
*Electromagnetic compatibility (EMC): EN61326-1:2006
Meets performance criterion 2, which means that the functionality and performance are temporarily reduced or lost, but can be restored on their own.
*Protection level IP65: dustproof and water-resistant
*Vibration and drop: IEC 60068-2-64:2008 IEC 60068-2-32:2008
Random, 2g, 5-500Hz; 1-meter drop test.
*CE certification and national CMC certification.
*Quality standards: Developed, designed, and manufactured in accordance with DIN ISO 9001.
*Calibration cycle and preheating time:
To ensure the accuracy of this table, its calibration cycle is one year
The startup preheating time is greater than 10 minutes
*Display:
3.2-inch TFT color screen display
*Power supply:
Powered by 4 × 1.5V AAA alkaline batteries
Power adapter for power supply
*Size and weight
206 × 97 × 60 mm
About 600g