The DRPL-3B thermal conductivity meter (protective flat plate heat flow meter method) adopts the internationally popular heat flow meter method for detecting thermal conductivity, and is equipped with a computer for fully automatic detection. It has automatic pressurization and automatic thickness measurement functions, and can be used for testing under different pressures, reducing manual measurement errors and adapting to the testing requirements of thermal conductivity and thermal resistance of various common materials.
Product Details

DRPL-3BThermal conductivity meter (protective plate heat flow meter method)

Product Introduction

DRPL-3BThermal conductivity meter (protective plate heat flow meter method)Adopting the internationally popular method of using a heat flux meter to detect thermal conductivity, coupled with a computer to achieve fully automatic detection, it has automatic pressure and thickness measurement functions, can be tested under different pressures, and reduces manual measurement errors, adapting to the testing requirements of thermal conductivity and thermal resistance of various common materials.

This instrument is based on the principle of steady-state flat plate method testing. A stable hot surface temperature is added to the hot surface, and heat is transferred to the cold surface through the sample. The transferred heat flux is measured, and the thermal conductivity and thermal resistance can be calculated based on the thickness and heat transfer area of the sample. This testing method is simple, fast, and has good repeatability. Very suitable for research and development in material heat transfer. Can be used for testing nylon, plastic, silicone rubber, graphite, insulation materials, etc., widely used in colleges, research institutes, quality inspection, factories and mines.

Reference standard for this instrument:ASTM C518-04Test method for determining steady-state heat flux and heat transfer characteristics using a heat flux meter;GB10295-2008Determination of steady-state thermal resistance and related characteristics of insulation materials (heat flow meter method) andGB/T25261-2018Reflective thermal insulation coating for buildings AppendixARequirements for determination of thermal conductivity.

Applicable Materials

This instrumentCan be used for testing the thermal conductivity of materials such as nylon, plastic, silicone rubber, graphite, insulation materials, and reflective insulation coatings for construction.

Applicable standard

The product is applicable to the following standards:

ASTM C518-04Test method for determining steady-state heat flux and heat transfer characteristics using a heat flux meter

GB10295-2008Determination of steady-state thermal resistance and related characteristics of insulation materials (heat flux meter method)

GB/T25261-2018Reflective thermal insulation coating for buildings AppendixA

Main technical parameters

1Range of thermal conductivity: low thermal conductivity materials0.00105W/(mk)High thermal conductivity material3500W/(mk)Accurate to four decimal places.

2Accuracy: superior to±4%Repeatability:±2%.

3Hot surface temperature: room temperature+10℃~100.00℃Using a high-precision display meter for temperature measurement with high resolution0.01℃.

Or room temperature~300.00℃; Resolution0.1℃Choose between two specifications.

4Cold face temperature:070℃Equipped with a constant temperature water bath for cooling; Using a high degree display meter for temperature measurement,0.2Grade accuracy, resolution0.01℃.

5Hot surface temperature control: room temperature~100.00℃, Resolution0.01℃Flat panel heater, bidirectional thyristor control.

Or room temperature~300.00℃; Resolution0.1℃Choose between two specifications.

6Using computer automated testing, data processing, and the ability to generate detection reports and print them out.

7Range of heat flow:80~2000W/m2.

8Using an organic glass windproof cover, the experimental process can be directly observed.

9Test surface size:100×100mmTest sample size for low thermal conductivity materials100×100×(1~5)mm.

10Equipped with a thermal protection device to ensure testing accuracy.

11Equipped with automatic pressurization and thickness measurement functions, it can detect the thermal conductivity and resistance under different pressure states, making it very suitable for materials with compressibility

Automatic pressure measurement:0.01000.0N, Resolution0.1N

Automatic displacement measurement:0.00100.00mm, Resolution1Mm.

12Loading and unloading method: stepper motor.

13Voltage:220V50HZ.

14Optional powder sample frame can be added for powder detection.

Configuration List

1Test host 1Taiwan

2Software 1set

3Organic glass windproof cover 1Taiwan

4High precision constant temperature water bath 1Taiwan

5Lenovo original computer 1Taiwan