Welcome Customer !

Membership

Help

Changzhou Xunsheng Instrument Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Changzhou Xunsheng Instrument Co., Ltd

  • E-mail

  • Phone

  • Address

    No. 96 Xicheng Road, Jintan District, Changzhou City

Contact Now

Stainless steel temperature controlled electric heating plate

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Stainless steel temperature controlled electric heating plates are widely used for sample baking, drying, and other temperature tests. They are commonly used tools in biology, genetics, medicine and health, environmental protection, biochemistry laboratories, analysis rooms, teaching and research. Its main features are: 1. The working panel is made of stainless steel, which has excellent corrosion resistance. $r$n2、 Fast and uniform heating, easy operation, and safe use.

Product Details

1Stainless steel temperature controlled electric heating plate Product Introduction

Stainless steel electric heating plates are widely used for baking, drying, and other temperature tests of samples, and are commonly used tools in biology, genetics, medicine and health, environmental protection, biochemical laboratories, analysis rooms, teaching and research.

Its main features are:

1The work panel is made of stainless steel, which has excellent corrosion resistance.

2Fast and uniform heating, easy operation, and safe use.

II Stainless steel temperature controlled electric heating plate parameter

1Model:DB-6Atype

2power supply220V 50Hz

3Heating power: 2500W

4Working area:600×400mm

5Digital Display: Room Temperature--300degree Resolution:1degree

III maintenance use

1The power socket should be properly grounded for safety.

2Please wipe the working surface clean before and after use, and there should be no water droplets, dirt, scale, or other foreign objects left on it.

3The sample bottle or other container should be placed on the working surface before heating to prevent bursting.

Connect the power, turn on the power switch, and according to your work needs, press“SET”The key or "∧", "∨" can enter the temperature setting state. Display on the lower monitor“SP”Press the "∧" or "∨" key to increase or decrease the temperature setting value until it reaches the set operating temperature, as shown on the monitor. After setting, press“SET”Press the key to approve and exit the setting state.

For detailed operation, please use according to the temperature control table manual

Setting of instrument working mode and description of various parameters (according toSETkey3Can enter the inner working parameters in seconds)

code name

Parameter meaning

say bright

Set range

default

AL1

alarm1parameter

alarm1Temperature control point for relay operation

0- Full range

5

AL2

alarm2parameter

in703There is no hardware support for this item in the instrument

0- Full range

10

FAL

Alarm point hysteresis

Alarm output action insensitive area

0- Full range

1

P

Proportional band

ifP0For positional control.PThe instrument is a proportional output type when the value is not zeroPIDControl,PFor dynamic working range.

1-200

20

Hyc

Main output hysteresis

Main output action insensitive area(P0(Appears)

050

2

I

Integral time constant

Eliminating static errors and the intensity of instrument control over output energy.IThe smaller the value, the stronger the control ability.

0-999second

200

d

Differential time constant

Pre adjustment capability refers to the speed at which the output energy is regulated when the measured value changes.DWhen the value is high, the response is fast

0-999second

50

T

Proportional cycle

The total heating and termination time is one cycle. The instrument controls the output power by adjusting the heating and termination control

1-200second

20

Sc

Measurement value bias

Sensor error correction

±19.9

0

At

auto-adaptation

The process of automatically selecting appropriate control parameters for instruments

0close/1open

0

EP

Proportional band offset

Used to prevent temperature overshoot of instruments after cold start

10

10

Lck

Parameter lock

0Not locked;1Lock the inner layer parameters;2Lock all parameters

0-2

 0

4、 Operating instructions

1.set temperature: PressSETThe key can set the temperature. pressSETThe data on the digital display below the key flashes(Upper row normal temperature measurement),At this point, the instrument enters the temperature setting state,Press the △ key to increase the set value, press the ▽ key to decrease the set value, and then press againSETThe key instrument panel returns to normal working state and the temperature setting is completed.

2Correction of sensor errors: PressSETkey3Quickly enter the inner menu of the instrument panel and select the parametersSCThe error can be corrected by using the △ or ▽ keys afterwards. The display resolution is1The range of correction for degrees is±50Degree.Press again after completionSETkey3Exit in seconds. The correction value of the instrument when it leaves the factory is0.

3Start self-tuningThis product can adapt to the vast majority of control systems, and can only start self-tuning when the control effect is particularly unsatisfactory. Method 1: After powering on the instrument, press the shift button3Start self-tuning in seconds. Method 2: PressSETkey3Quickly enter the inner menu of the instrument panel and select the parametersAtPress the △ key to adjust the lower row1Press againSETkey3The self-tuning can be activated in seconds. Self tuning in progressActorThe indicator light flashes, and the process is that the instrument control heating system fluctuates twice near the set point. WhenActorWhen the indicator light goes off, the self-tuning ends and the new PID parameters are locked in the chip.

4Operation Flowchart

5Precautions

1The instrument display is normal, but the temperature suddenly loses control. It is necessary to check whether there is a short circuit or open circuit in the external load.

2instrument displayHHHThe measured value exceeds the instrument temperature sensor and the wire is disconnected, indicatingLLLIt's a measured value

Below the lower limit of the instrument range or connected to the wrong sensor.

When the electric heating plate is in working condition, there should be a dedicated person to take care of it. Do not touch the surface of the workbench with your hands to prevent burns.

Work completed, cut off the power supply.