16 hole constant temperature water bathUsed for distillation, drying, concentration, and temperature preservation of chemical or biological products, it is a commonly used product in various laboratories.
Technical parameters:
1Electricity Source:220V 50HZ
2Heating power:3000W
3Constant temperature range: room temperature-100℃
4Temperature control accuracy: ≤±0.5℃ Resolution:0.1℃
5Product features: The water bath pot is made of stainless steel, which has excellent corrosion resistance.
6Adopting an intelligent temperature controller with high temperature control accuracy, it can set temperature and display temperature,PIDIntelligent temperature control, intuitive and accurate digital display.
7Water tank size:1200×130×120(mm) Including the height of the waterproof board30(mm)
8Configuration: Equipped with water and power outage protection function, silicone cover ring, stainless steel partition
116 hole constant temperature water bath Usage
When using this instrument, please first check if all the accessories accompanying the machine are complete, and then plug in the power plug of the instrument,Turn on the power switch and press according to your work needs“SET”The key or "∧", "∨" can enter the temperature setting state. Display on the lower monitor“SP”Press the "∧" or "∨" key to increase or decrease the temperature setting value until it reaches the set operating temperature, as shown on the monitor. After setting, press“SET”Press the key to approve and exit the setting state.
Operate according to the user manual of the temperature controller. When not working, the power should be cut off to ensure safety, avoid damaging the components, and prevent severe vibration.
5. operation
5.1The order of calling out each function
hController power on operation:
After the controller is powered on, the upper row displaysSYThe lower row displays the division mark, indicating the input type. after passing through2After seconds, the upper row displays the upper range limit and the lower row displays the lower range limit, indicating the measurement range. after passing through2After seconds, the upper row displays the measured value and the lower row displays the set value, entering the standard display mode.
hTemperature setting:
pressSETKey, upper displaySP, press5or6Press the key to display the desired value in the lower row. Press the function key, the upper row will displayST(in minutes), press5or6Press the key to display the desired value in the lower row, then click the function key to return the controller to standard mode.
5.2Detailed description of each function
hPVZone display:
![]()
If displayed“ ”It indicates that the sensor is open or the input signal exceeds the measurement limit; If displayed“ ”If the sensor is short circuited or the input signal is below the lower limit of the range, it means that when the input signal of the controller is not within the range, the buzzer will sound and press any key to silence it.
hTime inquiry:
When the controller is working in standard mode, simply press the shift button, the upper row will display the timer time, the lower row will display the controller set time, and then press5Key, the controller returns to standard mode.
hTimer function:
WhenSTset to0When the timing function is cancelled, the controller runs continuously; WhenSTSet not to0Only when the controller has a timing function, will the lower row display when the controller's running time is upENDThe buzzer beeps, the controller stops working, press any key to silence, and at the same timepress5Keys and6key4secondCan be restarted. When the timer function is activatedRUNIf the indicator light is flashing and there is no timer or if the timer has not been activated yetRUNThe indicator light stays on for a long time,ALMThe constant of the lamp indicates a fault alarm.
hWater level indication and replenishment function:
When the water level is low, display“0000”And it will sound an alarm, the buzzer will beep, the heating output will be automatically turned off, and any key can be pressed to silence the sound
5.3The self-tuning function of the controller (setting the set time to0Cancel the timer and set the timer time after the self-tuning is completed
If the temperature control effect is not ideal, please activate the self-tuning function and press6key4seconds laterATThe indicator light flashes, and the controller starts self-tuning. After self-tuning automatically endsRUNThe indicator light stays on for a long time; At this point, the control effect will be greatly improved.
hDuring the self-tuning process, press6key4The running indicator light will stay on for a long time after seconds, and the self-tuning will stop. The controller will operate as beforePIDControl the parameters.
hpressSETkey4Appears in seconds or moreLK, willLCKtune18Press againSETPrompt displaying control parameters on the top row of keys(Please refer to the Control Parameter List for details), press5or6Press the key to display the desired value in the lower row. long pressSETReturn to standard mode. (No key pressed)1Automatically return to standard mode after minutes)
The functional parameters are shown in the table below
prompt |
name |
set the scope |
explanation |
initial value |
|
Pb |
Zero position adjustment (intercept) |
-100.0~999.9 |
When the zero position error of the controller is large and the full-scale error is small, adjust the value. GenerallyPt100Rarely adjust this value. |
0 |
|
PK |
Full adjustment (Slope) |
-1000~2000
|
When the zero position error of the controller is small and the full-scale error is large, adjust the value.PK=4000×(Mercury thermometer reading value-Current temperature measurement value)/Current temperature measurement value, generallyPt100Adjust the value first. |
0 |
rH |
Upper range setting |
0.0~400.0 |
adjustrHCan make the measuring range of the instrument0~rH(℃). |
|
AL |
alarm setting |
0.0~Full range |
When the temperature exceedsSP+ALWhen the value is reached, the alarm light will light up and an alarm output will be generated. |
6 |
P |
Proportional band |
0.0~Full range |
Proportional adjustment,PThe larger the proportion, the smaller the effect, and the lower the system gain,P=0For positional control, the insensitive area is0.4. |
15 |
|
I |
integration time (Adjust the time again) |
10~3600second |
Integral action time constant,IThe larger the value, the weaker the integration effect |
300 |
|
d |
Differential time (Pre adjustment time) |
0~3600second |
Differential action time constant,dThe larger, the stronger the differential effect, and it can overcome overshoot |
100 |
Ar |
Overshoot suppression |
0~200% |
WhenArThe smaller the value of, the stronger the controller's ability to suppress temperature overshoot |
100 |
T |
Heating cycle |
1~100second |
Relay output ≮20s,SSRAnd thyristor switch ≮2s,Only applicable to the heating side |
3 |
dp |
decimal point |
0;1 |
DP=0The display resolution is1℃;DP=1The display resolution is0.1℃, |
1 |
CE |
time unit |
0-1 |
0It is measured in minutes;1It is measured in hours. |
0 |
three Precautions
1When an alarm is detected after turning on the machine, please check if the water is in place
2If leakage is found after prolonged use, please check if the heating tube is damaged, as the heating wire is220VConnect the AC voltage directly or check if there is any collision causing the circuit and insulation tube to fall off. If you do not understand electrical principles, please absolutely prohibit disassembly and repair. Notify our factory for repair.
Heating temperature is high, resulting in water evaporation. During use, a dedicated person should be present to supervise and prevent direct contact with the instrument casing to avoid burns.