Portable gas analyzer

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Product Name: K850 Portable Gas Analyzer Product Number: K850 Detection Object: Hydrogen and Carbon Dioxide Gas Concentration Instrument Range: 0~100% Resolution: 0.1% Application Field: Hydrogen Purity Analysis Field

Product Details

  Portable gas analyzerCharacteristics of

Capable of detecting multiple gases - including H2, CO2, Ar - with the option to measure O2. It has fast preheating, quick response, small maintenance requirements, and requires fewer calibrations.

Hachi's K850 series analyzer accurately detects changes in a certain component in binary or quasi binary mixed gases (i.e., only one component changes in the mixed gas). The gases that can be detected include H2, CO2, Ar, He, as well as various halogenated hydrocarbons such as Freon. As long as the thermal conductivity of one component in the mixture is different from that of another component, it can be detected in the mixture. Depending on the gas being detected, the range can be from high ppm to 100%.

Use a lossless thermal conductivity sensor to detect the thermal conductivity of gas. This type of sensor has a high sensitivity, low heat capacity, and non attenuating component that does not require maintenance. To achieve accuracy and detection range that conventional thermal conductivity sensors cannot provide, there is a microprocessor inside that provides signal processing. Due to the inherent high stability of the system, instruments require almost or no maintenance

The built-in sampling pump can draw sample gas from a low-pressure source or naturally draw sample gas. Use needle valves and flow indicators to control the sample gas flow through the analyzer.

Can detect two types of gases. It combines the detection function of K850S with the oxygen detection function of electrochemical cells. It can detect oxygen concentrations ranging from 100ppm to 100%, depending on the operating conditions.

The K850 uses sealed rechargeable batteries, so it is truly portable. Rechargeable batteries can be used continuously for 6 hours. If prolonged use is required, the instrument panel can be connected to a battery charger. The entire instrument is placed in a hard ABS box, equipped with a handle that can also be used as a foot to adjust the tilt height. The large and high-definition LCD displays the detected concentration, as well as the information and prompts that appear when setting and calibrating the instrument using the menu.

Applications: hydrogen production process, gas purity testing, food processing, refrigeration systems, power plants, breweries, metallurgical atmosphere control.

  Portable gas analyzerTechnical parameters

Display dot matrix LCD, displaying the range and gas types of thermal conductivity detection for two or four lines of characters from 0 to 10%, up to 0 to 100% (CO2, Ar, CH4) from 0 to 0.5%, up to 0 to 100% (H2, He). For other gases and ranges, please consult Hachi Company for accuracy (typical) within ± 1% of the range

Oxygen detection range (only for KG850)

0~100%

Resolution: 0.1%

Accuracy: ± 0.25%

0~25%

Resolution: 0.1%

Accuracy: ± 0.1% (range of 5% to 100%)

According to the working conditions, the ppm level measurement range can be selected. The typical lifespan of the oxygen battery is 2 years of response speed (typical)

Oxygen detection: 12s (T90)

Thermal conductivity testing: 20s

Sampling interface inlet and outlet: suitable for static sealed compression joints with an outer diameter of 0.25 '(6mm) pipes

sample gas pressure

Pump Off: Large+3barG

Pump on: Small -100mbG

Sample gas flow rate: Operating flow rate is 100-300ml/min

Sample gas temperature: -10 ℃~+60 ℃ (non aggregated) output 0-1V analog output

Environmental temperature: -5~40 ℃ Battery efficiency: 6 hours Battery charging

AC power: 240Vac (only available in the UK)

Car power supply: 12Vdc

The box is equipped with a hard ABS box with a handle, which can also be used as a support foot for adjusting the tilt height.

Size 250mm wide x 255mm deep x 94mm high, with front handle

Weight approximately 2kg