Laser headspace sealing integrity/leakage aperture detector

NegotiableUpdate on 01/30
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The HGA-1400v laser headspace sealing integrity/leakage aperture detector is designed for non-destructive testing of micro pore diameters in samples such as penicillin bottles and ampoules. It conducts in-depth research on micro pore drug packaging, microbial invasion, and other issues to ensure the quality and safety of drugs throughout their entire lifecycle.
Product Details

HGA-1400vLaser headspace sealing integrity/leakage aperture detectorConduct non-destructive testing on the micropore diameter of samples such as penicillin bottles and ampoules, and conduct in-depth research on micropore drug packaging, microbial invasion, and other issues to ensure the quality and safety of drugs throughout their entire lifecycle.

Technical Principle

The HGA-1400v laser headspace micro hole leakage detection system is based on laser absorption spectroscopy non-destructive testing technology to perform high-precision detection of micro hole leakage in sample bottles. The system has the characteristics of low detection limit, non-destructive measurement, and easy operation.

measured data

The following figure shows the test data. The red color represents the complete test data of the standard perforated sample penicillin bottle (8R-8.5 μ m) in the headspace microporous leakage system. The leakage aperture can be accurately calculated within the specified time, and the measurement results are consistent with those of the standard perforated sample bottle.

激光顶空密封完整性/泄漏孔径检测仪


System Application

The HGA-1400v laser headspace microporous leakage detection system does not require nitrogen purging. The system adopts a split design and can be matched with different specifications of penicillin/ampoule bottle fixtures according to testing needs. Non standard sample bottles can be customized with matching fixtures to meet users' precise measurement and research on the sealing integrity/leakage aperture of drug packaging materials throughout their entire life cycle.


Product advantages

Non destructive testing, full lifecycle testing

No need for nitrogen purging

Directly measure the internal pressure changes of samples such as penicillin bottles

Research on the integrity of sealing throughout the entire lifecycle

Ultra high precision micro hole leakage detection

Visual and intuitive analysis of measurement data

Split design, the cavity can be replaced as needed


HGA-1400vLaser headspace sealing integrity/leakage aperture detectorSpecifications

Detection function: sealing integrity/leakage aperture

Detection technology: laser absorption spectroscopy

Detection range: 0-10 μ m

Testing specifications: 2R~30R penicillin bottle, 1-25ml ampoule bottle, non-standard product consulting engineer

Detection time: 30s (related to the detection aperture)

Working environment: temperature 15 ℃~30 ℃, humidity 0~90%

Detection accuracy: 0.1 μ m (slight deviation due to different environments and packaging materials)

Communication interface: USB

Calibration cycle: 12 months

Laser grade: Class1 complies with IEC 60825-1

Shell material: Metal shell

Weight: 20kg

Chassis size: 240mm × 210mm × 300mm

Chamber size: 300mm × 270mm × 205mm

Power supply voltage: 220V AV, 200W, 50/60Hz