E-mail
xiaoming@herunyq.com
-
Phone
18312225276
-
Address
Building A, Pepsi Smart Park, No. 179 Spectrum East Road, Huangpu District
Guangzhou Herun Instrument Technology Co., Ltd
xiaoming@herunyq.com
18312225276
Building A, Pepsi Smart Park, No. 179 Spectrum East Road, Huangpu District
HGA-1400sLaser Headspace H2O/Leakage Online DetectorBy adopting a non-destructive and non-contact detection method, customers can perform headspace gas online detection on packaging products such as penicillin bottles, ampoules, and plastic bottles.
Technical Principle
The HGA-1400s laser headspace leak online detection system, based on laser absorption spectroscopy detection technology, can be used for headspace analysis (HSA) and leak detection (LD), and is integrated into linkage lines, lamp inspection machines, etc. for 100% detection and quality control.
measured data
The following figure shows the simulated production line test data. The red and blue colors represent the test data of two types of freeze-dried products, which were tested 1000 times each. The data is stable and the repeatability deviation is within ± 0.2kPa.

System Application
The HGA-1400s laser headspace detection system adopts laser absorption spectroscopy detection technology, with a wide range of product applications. The system can detect product leaks, vacuum degree, and gas components without waiting for leakage time. After freeze-drying products are out of the cabinet, leakage can be detected within 2-3 minutes, fully meeting customer needs.
HGA-1400sLaser Headspace H2O/Leakage Online DetectorProduct advantages
Non contact non-destructive testing
• Detection speed up to 600 bottles/minute
• Detection accuracy ± 0.2kPa (test data for controlled penicillin bottles)
• High system stability
No need for nitrogen purging
• Easy to integrate light inspection machines, linkage lines, etc
Specifications
Detection function: pressure/H2O concentration/leakage
Detection range: 0~120kPa (absolute pressure)
Testing specifications: 2R~100R penicillin bottle non-standard product consulting engineer
Detection speed: up to 600 bottles/minute
Working environment: temperature 15 ℃~30 ℃, humidity 0~90%
Detection accuracy: ± 0.2kPa (slight deviation due to different environments and packaging materials)
Communication interface: Numerical and additional information are communicated through RS-422 serial communication
Laser grade: Class1 complies with IEC 60825-1
Shell material: Metal shell
Probe size: 169mm × 84mm × 101mm
Chassis size: 240mm × 210mm × 300mm
Power supply voltage: 220V AV, 200W, 50/60Hz