Features:
Tubular design Ultra high light intensity
Balancing multiple factors Free amplification
Module Design Slot replacement
Temperature controlled mixing Integrated as a whole
Featured features:
1. Compatible with both homogeneous and heterogeneous phases, capable of multiple tube series connection
2. Tube channel design, ultra-high light intensity, up to 250W, effectively improves reaction efficiency
3. Adopting modular design, 1-4 external light tubes can be added according to demand to improve reaction efficiency
4. Integrated light source, peristaltic pump, stirring, and temperature control, easy to operate, can effectively control the reaction
5. The module interface adopts a threaded design, which is quick to disassemble and easy to connect
Accessories
Three neck reagent bottle (250mL), compatible with magnet, silicone hose
Specifications:
Operation Manual:
1. Connect the power supply
2. Turn on the switch: The instrument switch is located at the bottom of the side of the instrument, with "one" indicating power on and "O" indicating power off. The startup screen lights up.
3. Set power: Touch the value under "Power" on the display screen, and the value will start flashing. Rotate the knob clockwise (counterclockwise) to control the power increase (decrease).
4. Adjust the mixing speed: Touch the value under "Mixing" on the display screen, and the value will start flashing. Rotate the knob clockwise (counterclockwise) to control the mixing speed to increase (decrease).
5. Adjust the flow rate of the peristaltic pump: Touch the value under "flow rate" on the display screen to enter the flow rate setting interface. Click on the flow rate value, and the value will start flashing. Rotate the knob clockwise (counterclockwise) to control the flow rate to increase (decrease). Click on the arrow icon under "Direction" to switch the direction of the peristaltic pump (in the stopped state). After setting up, click the "OK" button to return to the previous interface.
6. Flow rate calibration function: Touch the value under "Flow rate" on the display screen to enter the flow rate setting interface, and then click the "Calibration" button to enter the calibration interface. Click on the value under "Speed", and the value will start flashing. Rotate the knob clockwise (counterclockwise) to control the increase (decrease) of the peristaltic pump speed. After selecting the speed, click on the value under "Time" and turn the knob clockwise (counterclockwise) to set the calibration time. After selecting the calibration time, click the start button, and the peristaltic pump will start running. It will automatically stop when the set time is reached, and the value under the "volume" on the right will start flashing. Record the volume of liquid flowing through the process and input it under the "Volume" label by rotating the knob. Click the "OK" button to complete the calibration and return to the previous interface.
7. Start experiment: Click the play buttons under "Flow rate", "Power", and "Mixing" on the main interface, and the peristaltic pump, LED light, fan, and mixing will start working respectively to start the experiment.
8. Click the pause button below to stop the reaction.
9. Turn off the instrument: Turn off the power switch and cut off the power.
10. Increase/decrease light strips: Connect the external light strips directly to the card slot at the bottom of the back of the host, plug and play.
11. Cleaning: Good solvents or acid-base solutions can be used to pump into the pipeline for cleaning. Please rinse with solvent before the experiment.
Notes:
1. Suggested working environment temperature range for the instrument: 0~50 ℃; And ensure that the distance between the equipment and other objects is at least 10cm to ensure heat dissipation effect;
2. This device should be placed in a cool, ventilated, and dry place; Stay away from flammable and explosive materials, corrosive materials, dust, etc. It is strictly prohibited to use and store in unprotected acidic or alkaline environments;
If the device is not used for a long time, it should be turned on regularly to check its working condition to avoid circuit damage, usually once a week;
4. The equipment body should be kept clean, and water or other foreign objects are strictly prohibited inside the equipment. Once the above problems occur, immediately shut down and report for repair. After confirming that all hidden dangers inside the instrument have been eliminated, it can be reused;
5. If any abnormalities such as short circuit, smoking, or alarm sound are found, immediately shut down and report for repair. It is strictly prohibited to turn on the machine during the waiting period;
6. Non professionals are prohibited from disassembling and assembling to avoid damaging the circuit.
Product operation: