Product Name: Linear Thermal Expansion Coefficient Tester
Product Model:HRZ-XPZ492
Product Introduction
Specializing in the determination of the average linear thermal expansion coefficient of various medicinal glass bottles, analyzing the expansion and contraction properties of glass and other materials under high temperature conditions during heating. The product complies with the average linear thermal expansion coefficient method (YBB00202003-2015) of the 2015 version of the National Drug Packaging Material Standard, and meets the requirements of relevant international and domestic standards and regulations.
Product features:
Ø Adopting vertical and single/double push rod options for easy sample installation;
Ø Industrial grade color touch screen, microcomputer control, clear view, sensitive touch, easy to operate;
Ø Intelligent control of the experimental process enables detection and control of displacement and temperature throughout the entire testing process;
Ø The instrument adopts programmed temperature control, and the heating rate can be freely set with high temperature control accuracy;
Ø Strong compatibility with samples, capable of measuring the coefficient of expansion of glass rods, tubes, plates, and finished products;
Ø The software can automatically calculate the coefficient of linear expansion and obtain the expansion curve for data processing;
Ø The testing system displays and records the temperature expansion coefficient curve, all raw data, and calculates the line expansion coefficient;
Ø Support the recording, storage, and viewing of experimental data, and configure a micro printer to directly output and print test results;
Ø Multi level user permission management, password login, convenient for user management.
Reference standard for HRZ-XPZ492 line thermal expansion coefficient tester
YBB00202003-2015 Method for Measuring the Average Linear Thermal Expansion Coefficient
Technical specifications of HRZ-XPZ492 line thermal expansion coefficient tester:
indicator |
parameter |
Furnace temperature range |
Room temperature~1000 ℃ |
heating rate |
0.1-10 ℃/min adjustable, programmed temperature control |
temperature control accuracy |
±1℃ |
Temperature recording value error |
≤±0.5﹪ |
Displacement measurement range |
±2.5mm |
Resolution of measuring expansion value |
0.1um, Imported displacement sensor with automatic range calibration
|
sample size |
Ø (25-50) mm × H (150-200) mm (different size specifications can be customized according to requirements) |
Test accuracy |
Better than ± 0.5% |
Overall dimensions |
1100mm(L)x460mm(B)x460mm(H) |
power supply |
AC 220 V 50 Hz |
Net weight |
About 60kg
|
Product configuration:
Standard configuration: Test host (with touch screen system), professional software, micro printer, quartz bracket, quartz standard sample