Performance Characteristics
Injection port flow rate, chromatographic column carrier gas flow rate, detector gas flow rate, precision control
Multiple injection ports can be configured: packed column injection port (with spacer blowing, can be connected to large-diameter capillary column), splitter/Non split capillary injection port, program-controlled cold column head injection port, volatile component series connection injection port (specially equipped with sample pretreatment devices such as headspace/sweep capture/thermal desorption);
Detector Design for Industry *
The amplification plate with special ceramic treatment is stable during humid seasons, automatically ignites, and has automatic protection against hydrogen leakage, making it safe and reliable.
Precise control program temperature rise and furnace temperature, precise furnace temperature tracking design
The internal control accuracy of the furnace is less than0.01℃; The temperature difference between internal points shall not exceed 1 ℃; Rapid heating and cooling, with a programmed heating rate of 120 ℃/min; Balance time of 10 seconds; Cooling time (300 ℃ to 50 ℃) for 6 minutes; The accuracy of temperature control (with a difference of less than 0.3 ℃ from the actual temperature).
Fully imported valve configuration system
*More equipped8 valves (four-way valve, six way valve, ten way valve, and liquid injection valve optional); Automatic control and sequential operation; 0 leakage blowing technology; High speed switching solenoid valve has a long service life (1 million times, 10mS).
Separated columns oven
In a multi column chromatography system, the operating temperatures of different columns can be set separately to achieve the best separation effect for each column;
Convenient implementation of pneumatic cutting technology
It can achieve multidimensional chromatography function to separate and analyze complex samples, and can realize the blowback function of the chromatography column to save analysis time and cycle;
Technical Specifications
| model | GC-9280 | |
| power supply | AC220V±10% 50Hz 2200W | |
| working environment | Environmental temperature: 5-35 ℃, relative humidity: ≤ 85%, no corrosive gases indoors, stable and vibration free workbench, no strong magnetic field around | |
| Related accessories | Detector (matched according to user needs) | Hydrogen flame ionization detector (FID) Thermal conductivity detector (TCD) Electron Capture Detector (ECD) Flame photometric detector (FPD) Nitrogen phosphorus detector (NPD) |
| Column temperature box | temperature range | Room temperature+5~400 ℃ |
| Temperature setting | 1 ° C; Program set heating rate of 0.1 ° C? C | |
| Temperature stability of chromatographic column | When the ambient temperature changes by 1? At C, is it 0.01? C | |
| Cheng Sheng's order | 8th order | |
| Cheng Sheng rate | 0.1 ℃ -40 ℃/min (increments of 0.1 ℃) | |
| Hydrogen flame ionization Detector (FID) | Control temperature | 400℃ |
| detection limit | ≤5×10-12g/s [n-C16] | |
| drift | ≤5×10-13A/30min | |
| noise | ≤2×10-13A | |
| Dynamic linear range | ≥107 | |
| Thermal conductivity pool detector (TCD) | Control temperature | 400℃ |
| sensitivity | ≥ 12000mvml/mg [benzene/toluene] | |
| drift | ≤30uV/30mi | |
| noise | ≤8uV | |
| Dynamic linear range | ≥104 | |
| Electron capture detector (ECD) | Control temperature | 350℃ |
| detection limit | ≤1×10-14g/ml (γ-666) | |
| drift | ≤50uV/30min | |
| noise | ≤20uV | |
| Dynamic linear range | ≥104 | |
| Flame photometric detector (FPD) | Control temperature | 250℃ |
| detection limit | ≤2×10-13g/s(P) ≤5×10-11g/s(S) | |
| drift | ≤4×10-11A/30min | |
| noise | ≤2×10-12A | |
| Dynamic linear range | ≥103(P) ≥102(S) | |
| Nitrogen phosphorus detector (NPD) | Control temperature | 400℃ |
| detection limit | ≤5×10-13g/s(P) ≤1×10-12g/s(N) | |
| drift | ≤25×10-12A/30min | |
| noise | ≤4×10-13A | |
| Dynamic linear range | ≥103(P) ≥103(N) | |
Chromatographic peak diagram









Application field

Service Commitment
Beijing Pru AnalysisinstrumentCo., Ltd.We are a chromatography manufacturer that values commitment and is highly responsible to our customers. Any user who purchases our chromatography products makes the following commitments:
1. Our company provides a one-year product warranty period, during which free on-site repairs are provided (excluding faults or damages caused by human factors or uncontrollable natural phenomena).
2. After receiving the repair notice, respond to the problem within 8 hours and arrive at the scene within 3 working days to solve the problem.
3. Users can consult technical issues through after-sales telephone and obtain clear solutions.
4. When users experience performance failures during normal use, our company promises the above warranty services. In addition, if there are clear provisions in the applicable laws and regulations of the country, our company will comply with the relevant laws and regulations.
5. During the warranty period, paid repair services will be provided for the following situations:
(1) Damage caused by human or uncontrollable natural phenomena
(2) Malfunctions or damages caused by improper operation
(3) Malfunctions or damages that occur due to modifications, disassembly, or assembly of the product.