Digital online ozone detection system

NegotiableUpdate on 03/18
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
Brief introduction: The Clark battery technology of the digital online ozone detection system is pre calibrated (no on-site calibration required), plug and play sensor probe, strong anti fouling membrane electrode protection, simple membrane replacement, direct data report, plug and play industrial communication adapter. This ozone sensor can be directly connected to PLC to achieve seamless integration with industrial control systems. Any computer can be used to display data, calibrate and customize measurement systems without the need for intermediate analyzer electronic modules. The sensor probe has been pre calibrated and can replace any other type of sensor without shutting down the system; Save space, time, and cost. Measurement system performance (25 degrees Celsius)
Product Details
Details introduction:

Digital online ozone detection system
Work environment:
PEEK or Ryton CPVC
Sensor Probe Sensor Probe
Temperature range: -5 to 50 degrees Celsius -5 to 50 degrees Celsius
*High pressure: 150 psig@50 85 degrees Celsius psig@50 Celsius
*High flow rate: 10 ft/S 10 ft/S

Power requirements (requiring Class II DC power supply):
Voltage range: DC 10-30V
*High power: 200mW
Typical power: 120mW

Structure:
Processing electrodes: gold cathode, silver anode, silver reference grade (3-electrode polarographic Clark cell)
Protection: Platinum
Membrane: PFA Teflon
Sealing ring: Viton (other materials can be used)
Sensor probe material: PEEK, Ryton or CPVC
DataStick material: 316 stainless steel, PEEK, Ryton or CPVC
Weight:
1.2 pounds=0.54kg (PEEK, Ryton or CPVC)
2.6 pounds=1.18kg (316 stainless steel)

Measurement unit:
Measurement unit: ppm, mg/l, %
Temperature units: Celsius, Fahrenheit

Calibration:
Air: automatically adjusted to the set pressure
Sample: 1 point
Zero: 1 point
Temperature: 1 and 2 o'clock
Ozone and temperature have been pre calibrated before leaving the factory

Compensation options:
Salinity: 0mS/cm to 999.9 mS/cm
Pressure: 539.2 mmHg to 792.4 mmHg
Temperature: Automatically ranging from -5 degrees Celsius to 50 degrees Celsius

Other settings options:
Sensor filter: 0 to 100 seconds
Temperature filter: 0 to 100 seconds

authentication:
**And divergence: CE (EN61326 and EN61326:1998)
Safety: CE (EN 61010-1),UL 508
Dangerous places: UL Class I, Division 2