1、 Product Introduction
Air cabintest process---Pollution discharge stage
1. After the experiment is completed, use a UV lamp for sterilization and disinfection,
2. After the testing sampling is completed, in order to proceed to the next experiment, it is necessary to completely replace the air in the inner box. Open the device door and use the high air volume ventilation fan at the bottom of the device to extract air for air exchange.
3. The discharge time is determined by the testing period during which the concentration of pollutants in the air environment inside the test chamber approaches the background concentration or meets the testing requirements.
The inner wall adopts 10mmThick floating glass made of stainless steel 304 # as a framework, sealing materialSpecially designed sealant with low adsorption capacity. The indoor air purification system is equipped with a multifunctional air filter, which can remove particulate matter(including bacteria))It can also remove chemical pollutants such as formaldehyde and toluene. The room is equipped with a temperature and humidity control system to regulate the temperature and humidity of the air,After being filtered by the system, it is sent to the laboratory and equipped with an indoor temperature and humidity monitoring system, which can constantly observe changes in indoor temperature and humidity.
Air cabinThe overall structure mainly consists of the chamber body, temperature and humidity control system, air mixing fan system, lighting systemexhaust system(including air valve))UV disinfection and sterilization, transfer window, air intake system(including air valve))Sampling and refuelingComposition of mouth and pollution source generation device.
2、 Technical parameters
project |
parameter |
|
1 |
Test chamber volume |
1m3 ±0.1m3 |
2 |
studio dimensions |
850 × 1080 × 1100mm (width × depth × height) |
3 |
Overall Dimensions |
1460 × 1225 × 1820mm (width × depth × height) |
4 |
temperature range |
18~30 ℃ |
5 |
temperature control |
Loop temperature control |
6 |
humidity range |
≤ ± 3% RH |
7 |
humidity control |
Freezing dehumidification, steam humidification (external) |
|
8 |
Background concentration |
Compliant with GBT18883-2002_ Indoor Air Quality Standards |
11 |
sealing performance |
Gas leakage less than5% |
|
12
|
Environmental cabin materials |
SUS304 mirror stainless steel 1.5mm, rounded corners, fully welded and polished |
Q235A cold-rolled steel plate 1.5mm, surface electrostatic spraying | ||
High density polyurethane foam molding thickness80 mm | ||
13 |
Sealing material |
Polytetrafluoroethylene, fluororubber, silicone |
14 |
air duct |
SUS304 stainless steel 1.0mm |
15 |
safety protection |
Power overload, leakage, phase loss, undervoltage, overvoltage |
16 |
power supply |
AC380V 3KW |
17 |
glove box |
a set |