1、 Product Introduction
20m ³ air chamberPreparation of bacteriaSuspension Take fresh culture of Staphylococcus aureus (8032) (3rd to 8th generation) on nutrient agar medium slant (18 h~24 h), and use a 5.0ml pipette to aspirate 3.0mL~5.0mL. Add nutrient broth to the slant tube, repeatedly blow and aspirate, wash off the bacterial coating, filter with sterile defatted cotton, and dilute with nutrient broth into a test bacterial suspension with a concentration of 5X10 * CFU/mL~3X10'CFU/mL.
7 adjacent pairs of aerosol cabinets or chambers, one for disinfectionToxicity test - one for experimental control. -Regarding aerosols
The environment in which the cabinet or aerosol chamber is located(including temperature, humidity, lighting, airtightness, and ventilation conditions) should be consistent. Cabinets (or rooms) should be constructed with aluminum alloy and glass. Temperature and humidity regulating devices, ventilator filtration and sterilization or other disinfection devices and corresponding pipelines shall be installed. In addition, windows such as sleeve operation and sample transmission for spray bacteria, disinfectant and sampling shall be set up.
20m ³ air chamberThis standard specifies the raw material requirements, technical requirements, inspection methods, usage methods, label instructions, and precautions for disinfectants used for indoor air disinfection. This standard is applicable for the main purpose of killing microorganisms in the airAn indoor air disinfectant that meets disinfection requirements.
2、 Technical parameters
serial number |
project |
parameter |
1 |
Test chamber volume |
20m3 ±0.5m3 |
2 |
studio dimensions |
2350 × 3400 × 2500mm (width × depth × height) |
3 |
Overall Dimensions |
3050 × 4350 × 3050mm (width × depth × height) |
4 |
temperature range |
18~30 ℃ |
5 |
Temperature control method |
Cold Heat Balance Method |
6 |
Temperature fluctuation range |
≤ ±0.5 ℃ |
7 |
humidity range |
40~70 % RH |
8 |
Humidity fluctuation range |
≤ ± 3% RH |
9 |
Humidity control method |
Freezing dehumidification, steam humidification (external) |
|
10 |
Background concentration |
Compliant with GBT18883-2002_ Indoor Air Quality Standards |
11 |
sealing performance |
Gas leakage less than5% |
12 |
Environmental cabin materials |
Floor: SUS304 mirror stainless steel 2.0mm |
Top plate: mirror stainless steel 1mm | ||
Side panel: 10mm tempered glass | ||
Frame: 50 × 50 × 1.5mm stainless steel frame | ||
13 |
Sealing material |
Polytetrafluoroethylene, fluororubber, silicone |
14 |
Mixing fan |
3 leaves |
15 |
air duct |
SUS304 stainless steel 1.0mm |
16 |
safety protection |
Power overload, leakage, phase loss, undervoltage, overvoltage, phase sequence protection |
17 |
power supply |
AC380V protective grounding 9KW |
18 |
Sampling interface |
6 of them(Φ50mm), can be increased according to actual requirements |
19 |
cleanliness level |
Open the high-efficiency air filter of the aerosol cabinet, purify the air in the test room, and achieve a cleanliness level of 10000 for particles with a diameter of ≥ 0.5 μ m within 30 minutes |