Themys H2A safe hydrogen experimental platform

A safe hydrogen experimental platformThemys H2Used to measure safe temperature, mass changes, heat and heat flux, and escaping gases in a hydrogen environment.
THEMYS H2Dedicated to material characterization research in hydrogen environment, suitable forTMATheTGAorSTAMeasurement.
It is equipped with multiple safety systems, the most important of which are hydrogen and oxygen probes to prevent accidental mixing of these gases inside the instrument.
THEMYS H2Suitable for any material characterization under hydrogen conditions, from hydrogen storage material characterization to oxide reduction, and can also simulate powder treatment conditions, such as metal powder injection molding.
Why are we different from others?
Ultra high temperature capability
The highest single furnace can reach1750℃
Built in multiple security systems
Ensure the safe operation of users and instruments
High precision, high flexibility
Hanging symmetrical upper balance, specially designed forTGAdesign
ultra-high temperatureDTAtechnology
highest1750℃
External integration capability
Used in conjunction with a mass spectrometer, up to1000℃
| Basic Parameters | DTA | DSC | TGA | TMA |
| temperature range(℃) | room temperature~ 1750 | room temperature~ 1750 | room temperature~ 1750 | |
| Programmable temperature scanning rate(℃/min) | 0.01 ~ 100 | |||
| Crucible volume or maximum sample size | 30 ~ 300Ml | 80 ~ 100Ml | 55 ~ 2 500Mlor | height: 20 mm |
| air circuit | 3Road carrier gas (including1Road hydrogen gas), optional among them1Road intake+ 1Road auxiliary gas, 2 MFC, | |||
| vacuum | Secondary vacuum(< 5*10-2mbar),Hydrogen resistant vacuum pump | |||
| MEASUREMENTS | DTA | DSC | TGA | TMA |
| range | ~~ | ±20 mg | ±2mm | |
| Maximum sample size(g) | 35 | ~~ | ||
| resolution | 0.4MW | 1MW | 0.002 Mg | 0.2 nm |
| Calorimetry accuracy(%) | 1.4%b,c | 0.9%b | ±0.06%d | ±0.08 10-6/℃e |
| temperature accuracy(%) | 0.7℃b | ~~ | ||
| a.MV =Micro voltage, tomWThe value in units depends on the type of sensor used; b.Based on the melting of metal standard samples; c.Calibration is required; d.Based on standard sample decomposition; e.Thermal expansion measurement based on sapphire standard sample. | ||||
Place one at each instrument outletHydrogen concentration sensorand aOxygen concentration sensorTo detect the presence of these gases in the exhaust gas.
oneIndependent control unit:
·If hydrogen gas is detected by a sensor,The furnace body and balance will be prohibited from being opened
·If the sensor detects oxygen, it will be prohibitedHydrogen enters the furnace
·passvisual alarmRemind users
·There is a system for stopping hydrogen gas supplyemergency button
Additional security features
·In vacuum or at temperatures higher than70At ℃, the furnace body cannot be opened
·If the full power exceeds one minute or the control thermocouple malfunctions, stop heating