The YHKC-2A particle strength tester uses pressure sensors with different ranges. At present, there are four digit displays of 100N, 300N, 500N, and 1000N on the instrument, which directly display the particle strength value of the tested sample. Pressure sensors with various ranges can be configured to measure the intensity values of various samples. This instrument has a small volume, direct reading of strength values, high accuracy, and is more convenient to use. It is an ideal new generation particle strength tester for measuring the strength of samples such as fertilizers, drugs, catalysts, alumina, and coal rods.

Product Details

YHKC-2AType particle strength testerUse pressure sensors with different ranges.Currently, there are 100N, 300N, 500N, and 1000N availableThe instrument displays four digits, directly showing the particle strength value of the tested sample. Pressure sensors with various ranges can be configured to measure the intensity values of various samples. This instrument has a small volume, direct reading of intensity values, high accuracy, and is suitable for usemoreconvenient,It's fertilizers, drugs, catalystsAluminum oxide, coal rodWaiting for sample strength determinationThe ideal next-generation particle strength tester.Instrument increase3Highlight: Firstly, the instrument has a self checking and zero calibration function. 2、 The instrument has a one click calibration function. 3、 The instrument has an over range protection function.

YHKC-2AtypeParticle strength testertechnical parameters

1 Measurement range 099.99N

2 Particle diameter:diameter< 25mm(Large diameter can be machined)

3 Relative humidity: 85%

4 Environmental temperature050

5 Power supply voltage: 220Vcommunication

6 Resolution:0.01N

7 Measurement accuracy:≤±0.5%F.S

8 External dimensions:18Omm×250mm×280mm

(The sensor accuracy is ±0.02%F.S,The instrument display design is0.00N)

Instrument usage method :

1 Turn on the power switch, the instrument indicator light is on, and the instrument displays1234After entering the self checking zero calibration function, it displays0.00N. (preheat2minute)

2 Place the test sample at the center of the sample trayOn a stainless steel surface,If the sample is too heavy,The displayed numbers have changed,thenpressa bitClear key,Make the instrument display as0.00.

Rotate the handwheel clockwise,Through spiral rotation, the center is pressedLower the force rod,When the force rod approaches the sample, Press the peak button once,At the same time, the indicator light flashes, and thenContinue slowlyclockwiseRotating handwheel,Slowly press the sample downwards, and as the pressure increases, the intensity value displayed by the instrument gradually increases.When the sample particles are broken,Under forceLarge values are lockedAt this point, the numbers are displayed directlyThe tested sampleIntensity value.

Press the peak button again, the indicator light will not flash, and the display will return to0Status, enter standby mode

3 When continuing the experiment,Repeat the above steps.

4 When the sample strength is too high,The displayed value exceeds95Ntime,Instrument peak beep indication, while instrument displayFFFFWhen, Should exit in a timely manner, To prevent damage to the sensor.