High chlorine wastewater standard COD digestion device

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

KHCOD-8G high chlorine wastewater standard COD digestion device, * in accordance with HJ/T 70-2001 high chlorine wastewater chemical oxygen demand determination CL calibration method. The $r $n instrument adopts microcomputer technology for timed control of the heating plate, which can simultaneously heat and digest 8 250ml conical flask reflux devices. Suitable for the determination of COD in wastewater from oil fields, coastal oil refineries, oil depots, chlor alkali plants, and deep-sea discharge of wastewater

Product Details

KHCOD-8GtypeHigh chlorine wastewater standard COD digestion deviceThe instrument adopts microcomputer technology for timed control of the heating plate, which can simultaneously heat and digest 8 250ml conical flask reflux devices.

The instrument adopts a glass burr reflux tube instead of a spherical reflux tube, and uses air cooling technology instead of tap water cooling, which can save water and standardize the instrument. Preparation, operation, and COD calculation of chemical solutions for instruments.

Suitable for the determination of COD in wastewater from oil fields, coastal oil refineries, oil depots, chlor alkali plants, and deep-sea wastewater discharge.

KHCOD-8G typeHigh chlorine wastewater standard COD digestion deviceTechnical Specifications

1Anti chlorine interference: 1000-20000mg/L

  1. Nitrogen flow rate: 6-60mL/min

3Measurement range: ≤ 1000mg/L (greater than dilution)

4Dissolving time: 2 hours and 30 minutes (time adjustable)

5Dissolving volume: 250mL

6Measurement error: ≤ 500mg/L, relative standard deviation not greater than 5.0;

≤1000mg/L, The relative standard deviation is not greater than 8.0%

7Heating power: 1600W (220V, 50HZ)

8Cooling method: dual fan cooling with water cooling

9Microcrystalline plate heating table, make 8 samples

Instructions for using the buttons on the front panel of the instrument

  1. blow-dry

Control the opening and closing of the fan.

  1. timing

Dissolve time control setting key. The instrument is set to power on for 2 hours and 30 minutes. Each press of the setting key automatically deducts 10 minutes,When using the standard reflux method to determine COD, it should be set to 2 hours.

  1. start

Start control button for digestion. After pressing the "Start" button, the instrument will perform heating reflux digestion and automatically count down from the set value (such as 2 hours). The heating indicator light will turn on. When the set time is reached, the instrument automatically stops heating and the heating indicator light goes out.

4There are 8 flow meters on the front panel of the instrument.

Instrument usage instructions

Install all connections according to the diagram first, familiarize yourself with them before starting the experiment.

Before powering on the instrument, as much distilled water as possible should be added from the water injection port of the reflux pipe to ensure cooling effect.

  1. Turn on the power switch on the right side panel of the instrument.

At this time, the instrument displays "000", "111", "222"... "999", and then shows that the digestion time is 2 hours and 30 minutes higher than the set value. The fan is turned on at the same time, and the fan indicator light is on.

  1. Change the digestion time setting value

Press the "timer" button to decrease the digestion time until the set value is 2 hours. Of course, the digestion time value can also be set arbitrarily as needed.

  1. Start reflux digestion

Press the "Start" button, and the instrument will start heating, digestion, and reflux. The "Start" indicator light will turn on, and the heating indicator light will flash.

  1. Reflux digestion completed

After digestion is complete, the instrument displays "000" and flashes, the "start" indicator light goes out, and the heating indicator light goes out.

Explanation: 1. To remove all the connecting pipes of the instrument, simply press down on the gasket at the quick connector and pull the pipeline outward.

2The nitrogen adjustment knob of the flowmeter on the instrument is precision adjustable to control the nitrogen flow rate.

Note: Sample processing and analysis methods refer to the environmental protection industry standard of the People's Republic of China - Determination of Chemical Oxygen Demand for High Chlorine Wastewater - Chlorine Calibration Method.

Packing list for high chlorine COD digester

serial number

Name

Unit

Quantity

1

High chlorine COD digestion machine host

platform

1

2

Burr reflux pipe