- Phone
-
Address
No. 14, 8th Floor, Building 3, Dahankou International Digital City, east of Huanhu Road and west of Beijing Guangzhou Railway, Fukou Street, Huangpi District, Wuhan City, Hubei Province
Wuhan Tonghui Technology Co., Ltd
No. 14, 8th Floor, Building 3, Dahankou International Digital City, east of Huanhu Road and west of Beijing Guangzhou Railway, Fukou Street, Huangpi District, Wuhan City, Hubei Province
THYM3000Insulator salt density testerUsed for anti pollution flashover detection in power systems, it is a specialized measuring instrument for measuring the equivalent salt density (hereinafter referred to as "salt density") on the surface of insulators, and can also measure the conductivity and temperature of solutions. The whole machine is widely used in power, teaching, scientific research, and other related industries due to its high measurement accuracy, large measurement range, and convenient use.
The classification of pollution levels and the drawing of pollution level distribution maps are the basis for anti pollution flashover work. Accurate pollution level distribution maps are the basis for selecting the external insulation creepage distance of transmission and transformation equipment. The equivalent salt density value on the surface of insulators is a quantitative data for determining the severity of external insulation pollution in electric ceramics, and is an important basis for dividing pollution levels and drawing pollution zone maps. Therefore, salt density measurement work is of great significance for the safe operation of the power system.
reference standard:
GB/T16434 – 1996Classification of Environmental Pollution and Selection Standards for External Insulation of High Voltage Overhead Lines, Power Plants, and Substations
GB/T16434-200XSelection and size determination of high-voltage insulators under polluted conditions1Part: Definitions, Information, and General Principles
Q/GDW152-2006Classification of Environmental Pollution and Selection Standards for External Insulation of High Voltage Overhead Lines, Power Plants, and Substations
THYM3000Insulator salt density testerFeatures:
1)Equipped with automatic range switching function and fast measurement speed(2.5s/Next time).
2)Large measurement range, salt density range0.0001mg/cm2~9.9999mg/cm2.
3)The Chinese and English interfaces can be switched independently.
4)adopt480*272Color touch LCD screen.
5)Can directly display and print salt density, conductivity, temperature, pollution level, and uniform creepage distance.
6)Automatically perform temperature compensation and display directly20℃Standard conductivity and equivalent attached salt at time(ESDD).
7)It has the function of automatically removing the salt content of the original solution, reducing the requirement for cleaning solution.
8)Automatically measure the salt density without charging(ESDD)Convert salt density to charged measurement(ESDD).
9)storable10Ten thousand sets of records, and the records can be exported toUDisk or print through a printer.
10)You can view, delete, and export individual records, as well as delete all records.
11)Built in large capacity rechargeable lithium battery(2600mAh)Suitable for field use.
THYM3000Insulator salt density testerTechnical Specifications:
3.1 measurement range:
Salt density: 0.0001mg/cm2 to 9.9999mg/cm2 (based on X-4.5 type insulator)
Measurement temperature: 0 ℃~100 ℃
Measurement of conductivity: 0-200000 μ s/cm
3.2 Basic Error:
Measurement of salt density: resolution 0.0001 mg/cm2
Full scale accuracy: ± 2%
Temperature measurement: resolution 0.1 ℃, accuracy ± 0.5 ℃
Measurement of conductivity: resolution 0.01 μ s/cm
3.3 Environmental temperature: 0 ℃ to 60 ℃.
3.4 Environmental humidity:<90%.
3.5 Volume and Weight
The dimensions of the entire machine chassis are 356mm in length, 260mm in width, and 133mm in height.
Whole machine weight: approximately 2.5Kg.