THA3110Magnetic pressure oxygen analyzer
technical indicators
Typical range:0~2%;0~10%;0~21%;0~100%;98~100%
Linear deviation: ±2%FS
Stability: ±2%FS/7d
Repeatability: 1%
Response time(T90):≤35s
Technical Specifications
Working environment temperature: (5~45)℃
Tested gas flow rate:0.5L/min
Power Supply:220V 50Hz
output: 4~20mA
Communication method:RS485(Modbus RTU)
Reference gas flow rate: approximately10mL/min
Instrument functions
Based on oxygen paramagnetism,THA3110The type magnetic pressure oxygen analyzer uses the magnetic pressure characteristics generated by oxygen molecules in a magnetic field to achieve the analysis process of oxygen concentration. It is used for online analysis of oxygen concentration in industrial processes and scientific laboratories, and has the characteristics of high automation, strong functionality, and digital communication.
working principle
Under the influence of an external magnetic field, matter will be magnetized and exhibit certain magnetic properties. When a substance is magnetized in an external magnetic field, it generates an additional magnetic field. When the direction of the additional magnetic field is the same as that of the external magnetic field, the substance is attracted by the external magnetic field; When the direction is opposite, it is repelled by the external magnetic field. The substance attracted by an external magnetic field is called a paramagnetic substance, or in other words, the substance has paramagnetism; The substance that is repelled by an external magnetic field is called a diamagnetic substance, or in other words, the substance has diamagnetism.
Gas media can also be magnetized in a magnetic field, exhibiting paramagnetism or diamagnetism depending on the type of gas. asO2TheNOTheNO2Waiting for paramagnetic gas,H2TheN2TheCO2TheCH4Wait, it's a diamagnetic gas.
After the measured oxygen enters the magnetic field, the pressure of the gas will change under the action of the magnetic field, resulting in a pressure difference between the gas in the magnetic field and the space without the magnetic field. There is a proportional relationship between oxygen concentration and pressure difference, so magnetic pressure can be used to analyze oxygen concentration.
Technical Features
uresistNH3TheCOTheCO2andH2Waiting for gas cross interference.
uCorrosion resistance, allowing the sample gas to contain an appropriate concentration of corrosive gases (such as sulfides).
uExcellent microfluidic sensors and electrical design ensure high stability.
uNo consumable parts, long service life.
uNot affected by environmental vibrations.
Typical Engineering Application Fields
uAnalysis of Oxygen in Industrial Processes such as Fertilizer and Chemical Industry
uOxygen analysis in the cement and metallurgical industries
uAnalysis of Oxygen Composition in Flue Gas
uOxygen analysis in the biomedical industry
uGas analysis in scientific laboratory