Instrument features:
The data results can also be used to determine the safety critical parameters of the tested substance. The current standards in the industry focus on determining whether energetic materials can be safely operated and processed in experimentation, development, and production. The amplification ratio program for materials that are potentially sensitive to electric shock generally requires testing to evaluate the spark energy required for material detonation. The sensitivity measured is compared with the energy voltage during the production process to determine if it exceeds the safety limit. The electrostatic spark sensitivity test is very effective in providing such information.
Technical parameters:
1. Compliant with standards: MIL-STD-1751A, STANAG4490, etc;
2. High voltage power supply: output voltage 0.50-10.0KV, drift not exceeding 5% within 30 minutes, accuracy: ± 0.01KV;
3. Current output: 10mA;
4. Maximum voltage not exceeding 10KV, electrode gap adjustment range: 0-10mm;
5. High voltage capacitors: 5, 15, 30, 100, 200nF (built-in, withstand voltage 15KV);
6. High voltage capacitors: 20, 40, 100, 250, 400pF; 1, 2, 3 nF (external, withstand voltage 15KV);
7. Series discharge resistance: 10k Ω;
8. Energy recording device, equipped with high-voltage and current probes, to measure and record the true ignition energy caused by sample absorption leading to explosion, as well as the total energy of electrostatic sparks, with an accuracy of ± 1 μ J;
9. Spark energy range: 25 μ J-25 J, accuracy: ± 1 μ J;
10. Specialized software package: computer-controlled operation, software collects data from high-voltage probes and current probes during the discharge process, accurately analyzes and evaluates voltage and current attenuation characteristics, integrates and calculates the actual spark energy value between electrodes;
11. Automatically record the number of electrode ignitions, prompt timely replacement of electrodes, and prevent electrode loss from affecting testing;
12. Support remote control device ignition, instrument end delayed ignition, and remote control host selection of test conditions to ensure the safety of test personnel;
13. Double interlock protection, when the door is not closed, the high-voltage system power supply fails; After the door is closed, the high-voltage system is activated, allowing spark gap discharge;
14. Compressed air drive device: The air source pressure is 3 bar, which drives the opening and disconnection of the high-pressure system. The pneumatic valve controls the electrode to move and reset quickly, which can reduce the energy loss of the discharge circuit and ensure the safety of the operator's test.