SLIA-5000Silicon dioxide analyzer
Quickly and automatically measure trace amounts of silica in boiler water and purified water
The management of silicon dioxide (SiO2) is crucial for preventing the formation of scale. This is because the presence of scale in turbines can reduce power generation efficiency, especially for thermal and nuclear power plants. Based on HORIBA's years of technical experience in the field of analysis, we have developed the SLIA-5000 silica water quality analyzer with stability and maintainability for measuring trace amounts of silica in boiler water and pure water.
Measurement time 5 minutes, repeatability ± 2% F.S
Has extremely fast response speed and high repeatability in measurement results.
● Wide range
Measurement can be achieved from high sensitivity of 0-10 μ g/L to high concentration of 0-5.0 mg/L.
● Self diagnostic function
Self diagnosis of various abnormalities, such as sampling or reagent supply abnormalities, calibration abnormalities, light source abnormalities, and internal temperature abnormalities.
● Automatic calibration function
Automatic zero point calibration and range calibration can be preset to be performed at specific times for each measurement.
● Able to automatically switch measurement cycles
Automatically shorten the measurement cycle when the sampling concentration exceeds the set concentration range, and automatically extend the measurement cycle when the sampling concentration is below the set concentration range, which can save power and reagents.
● Improve operability
Equipped with a color LCD touch screen, it has strong operability.
SLIA-5000Silicon dioxide analyzerspecification
Product Name |
silicon dioxide |
model |
SLIA-5000 |
Overall dimensions |
600(W)x750(D)x1600(H)mm |
weight |
130kg |
power supply |
100V AC±10% 50/60Hz |
Consumption voltage |
500VA |
measurement range |
0~10/0~100μg/L |
repetitiveness |
±2%F.S. |
Measuring principle |
Molybdenum Blue |
measurement point |
1-4 o'clock (please specify when purchasing) |
measure time |
5min |
measurement period |
5~999min (can be selected in the control panel) |
correction |
Zero point calibration: automatic calibration for each measurement |
compensation |
Zero drift -300.0~300.0 μ g/L (can be selected in the control panel) |
reagent |
A: Sulfuric acid and ammonium molybdate |
Reagent exchange cycle |
One week (reagent storage temperature): 30 ℃ (86 ° F) or lower |
ambient temperature |
0~45℃(32~113°F) |
ambient humidity |
Relative humidity of 85% or lower (no condensation) |
Installation environment |
Indoor installation, with no dust or mist in the air, no corrosive gases, no direct sunlight, and minimal vibration |
Sampling conditions |
Temperature 2~40 ℃ (36~104 ° F), flow rate 0.5~1.0L/min, pressure 0.02~0.20MPa, pH value ≤ 11.0 |
| Concentration output | Standard: DC 4~20mA or DC 0~16mA (operation setting) Optional: DC 0~1V or DC 1~5V Output format: Maintain output or mark output (operation setting) Maximum load below 500 Ω, input/output insulation type, non insulated between channels |
| Contact output | Standard: A contact output (please contact us if you want to set it to B contact) Contact capacity AC250V 3A, DC 30V 3A Universal measurement circuit: in measurement, standby, maintenance, external control status, synchronization signal, instrument abnormality, power interruption, alarm, etc Each measurement line: measurement line, measurement range, sampling interruption, abnormal concentration |
| Contact input | Effective for remote control without voltage a contact input (optional contact) Insulated input (-) side universal ON impedance: max100 Ω Open voltage: maxDC 26V Short circuit power supply: maxDC 13mA Universal measurement circuit: measurement starts, calibration starts, cycle switching Each measurement line: measurement line setting, measurement range setting High, measurement range setting Low |
display |
Color LCD touch screen |
data storage |
Measurement value: Over one year |