Welcome Customer !

Membership

Help

Shanghai Mowei Biotechnology Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Shanghai Mowei Biotechnology Co., Ltd

  • E-mail

  • Phone

  • Address

    6th Floor, Building 8, Lane 4855, Guangfulin Road, Songjiang District, Shanghai

Contact Now

On site installation type MLSS instrument (four wire transmission)

NegotiableUpdate on 05/07
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

HU-200SS on-site installation MLSS instrument (four wire transmission) is connected to an immersion MLSS sensor (SS-90) and measures MLSS in activated sludge. The use of near-infrared radiation transmission method can measure a wide range of SS content (0-2000mg/L) with good repeatability

Product Details

HU-200SSOn site installation type MLSS instrument (four wire transmission)Connect the immersion MLSS sensor (SS-90) and measure MLSS in activated sludge. The use of near-infrared radiation transmission method can measure a wide range of SS content (0-2000mg/L) with good repeatability. MLSS、 Any calibration curve of clay (inorganic mud) and kaolin can be used as a standard substance for SS. If "MLSS" is used for the activated sludge treatment tank, "clay" is used for the construction sludge sedimentation treatment, and "kaolin" is used for the white sludge treatment, a concentration closer to the analytical value can be obtained.

range

Activated sludge (MLSS): 0~20000 mg/L

White inorganic powder: 0~10000 mg/L

Grey inorganic mud: 0~20000 mg/L

sturdy

Dealing with harsh conditions during on-site measurement.

Operating temperature range (transmitter): -20-55 ℃

Rated power supply voltage: AC 100~240 V ± 10% 50/60 Hz

Aluminum alloy watch box: protection level IP65

Compliant with CE marking and FCC rules, suitable for use on overseas sites.

intelligence

The self diagnostic function can immediately detect sensor errors.

The lead-free electrode design complies with WEEE and RoHS standards.

Built in three calibration curves, which can be selected according to the sample.

Long life LED light source that eliminates ambient light through flickering.

Near infrared radiation transmission method with stable sensitivity.

easy to maintain

Easy to view large display screen.

Easy to operate.

Can store calibration records.

Use threaded terminals in the wiring board that will not come off for wiring.

Stain resistant PFA sensor head.

HU-200SSOn site installation type MLSS instrument (four wire transmission)specification

transmitter

measurement method

Light transmission method

range

Activated sludge (MLSS): 0~20000mg/L Resolution: 1mg/L (0~10000mg/L)
高岭土: 0~10000mg/L 10mg/L(10000~22000mg/L )
Clay (inorganic clay): 0-20000mg/L

repetitiveness

Within ± 3% of the reading or ± 10mg/L, whichever is greater. (Sensor connected to sludge measurement value)

Transfer output

1. 4-20mA DC input/output insulation type maximum load impedance 900 Ω

Contact output

3-point no voltage contact output relay contact, SPDT
Contact function: R1, R2: upper limit alarm, lower limit alarm, transmission output hold, cleaning command output (off when alarm action occurs, normally open, off when power is off) optional
FAIL: Fault alarm (normally off, on in fault state, off in power failure)

Cleaning output

1-point active voltage contact output (connected to power supply voltage output)
Contact function: Driven by a cleaning solenoid valve

Contact input

1-point contact form: collector open circuit no voltage contact
Contact function: cleaning command, external input

communication function

RS-485 2-wire type, input/output insulation type (transmission output is non insulated)

ambient temperature

-20~55 ℃ (no freezing)

calibration method

Zero point calibration: use clean water
Compensation correction: concentration conversion method using coefficient input
Selection of measurement line: (Select activated sludge and inorganic sludge) has the function of adjusting the slope by analyzing the displayed value and manually analyzing the sludge value (weighing method)

Self-diagnostic function

Sensor diagnostic error, transmitter malfunction

power supply

AC 100~240 V ± 10% 50/60 Hz Power consumption: 20W or lower

construction

Outdoor installation type: IP65 protection level
Installation method: 50A column or wall installation
Shell material: Aluminum alloy
Installation fixing and protective cover material: SUS304

weight

About 4.5 kg

Electrical certification

CE marking,FCC rules

detector

Product Name MLSS detector for industrial use
model SS-90
Measuring principle Light transmission method
light source Infrared LED, 880 nm, approximately 10 Hz, AC flashing
detector (transmitted light/reference light)
Measuring pool length About 7 mm
data transmission RS-485 (communication with transmitter)
Self-diagnostic function Water immersion testing
Operating temperature 5 ℃~50 ℃ (no icing)
Storage temperature −25℃ ~ 65℃
Measure liquid pressure 0 MPa ~ 0.2 MPa
Moisture resistant material PPO SUS316 PFA FKM EPDM PVC
cable length 10 meters (standard), length: 50 meters (total length: 60 meters)
power supply 12 V DC
Overall dimensions Diameter 48 millimeters, length 250 millimeters (excluding cables)
weight About 0.8 Kg (excluding cables)