- Phone
-
Address
6th Floor, Building 8, Lane 4855, Guangfulin Road, Songjiang District, Shanghai
Shanghai Mowei Biotechnology Co., Ltd
6th Floor, Building 8, Lane 4855, Guangfulin Road, Songjiang District, Shanghai
HU-200SSOn site installation type MLSS instrument (four wire transmission)Connect the immersion MLSS sensor (SS-90) and measure MLSS in activated sludge. The use of near-infrared radiation transmission method can measure a wide range of SS content (0-2000mg/L) with good repeatability. MLSS、 Any calibration curve of clay (inorganic mud) and kaolin can be used as a standard substance for SS. If "MLSS" is used for the activated sludge treatment tank, "clay" is used for the construction sludge sedimentation treatment, and "kaolin" is used for the white sludge treatment, a concentration closer to the analytical value can be obtained.
range
Activated sludge (MLSS): 0~20000 mg/L
White inorganic powder: 0~10000 mg/L
Grey inorganic mud: 0~20000 mg/L
sturdy
Dealing with harsh conditions during on-site measurement.
Operating temperature range (transmitter): -20-55 ℃
Rated power supply voltage: AC 100~240 V ± 10% 50/60 Hz
Aluminum alloy watch box: protection level IP65
Compliant with CE marking and FCC rules, suitable for use on overseas sites.
intelligence
The self diagnostic function can immediately detect sensor errors.
The lead-free electrode design complies with WEEE and RoHS standards.
Built in three calibration curves, which can be selected according to the sample.
Long life LED light source that eliminates ambient light through flickering.
Near infrared radiation transmission method with stable sensitivity.
easy to maintain
Easy to view large display screen.
Easy to operate.
Can store calibration records.
Use threaded terminals in the wiring board that will not come off for wiring.
Stain resistant PFA sensor head.
HU-200SSOn site installation type MLSS instrument (four wire transmission)specification
transmitter
measurement method |
Light transmission method |
range |
Activated sludge (MLSS): 0~20000mg/L Resolution: 1mg/L (0~10000mg/L) |
repetitiveness |
Within ± 3% of the reading or ± 10mg/L, whichever is greater. (Sensor connected to sludge measurement value) |
Transfer output |
1. 4-20mA DC input/output insulation type maximum load impedance 900 Ω |
Contact output |
3-point no voltage contact output relay contact, SPDT |
Cleaning output |
1-point active voltage contact output (connected to power supply voltage output) |
Contact input |
1-point contact form: collector open circuit no voltage contact |
communication function |
RS-485 2-wire type, input/output insulation type (transmission output is non insulated) |
ambient temperature |
-20~55 ℃ (no freezing) |
calibration method |
Zero point calibration: use clean water |
Self-diagnostic function |
Sensor diagnostic error, transmitter malfunction |
power supply |
AC 100~240 V ± 10% 50/60 Hz Power consumption: 20W or lower |
construction |
Outdoor installation type: IP65 protection level |
weight |
About 4.5 kg |
Electrical certification |
CE marking,FCC rules |
detector
| Product Name | MLSS detector for industrial use |
|---|---|
| model | SS-90 |
| Measuring principle | Light transmission method |
| light source | Infrared LED, 880 nm, approximately 10 Hz, AC flashing |
| detector | (transmitted light/reference light) |
| Measuring pool length | About 7 mm |
| data transmission | RS-485 (communication with transmitter) |
| Self-diagnostic function | Water immersion testing |
| Operating temperature | 5 ℃~50 ℃ (no icing) |
| Storage temperature | −25℃ ~ 65℃ |
| Measure liquid pressure | 0 MPa ~ 0.2 MPa |
| Moisture resistant material | PPO SUS316 PFA FKM EPDM PVC |
| cable length | 10 meters (standard), length: 50 meters (total length: 60 meters) |
| power supply | 12 V DC |
| Overall dimensions | Diameter 48 millimeters, length 250 millimeters (excluding cables) |
| weight | About 0.8 Kg (excluding cables) |