SZD-12 reciprocating horizontal porous oscillator
1、 Product Introduction:
SZD-12 reciprocating horizontal porous oscillatorIt is a biochemical instrument for cultivating and preparing biological samples, widely used in laboratories of scientific research, universities and other enterprises and institutions for shaking cultivation of plants, organisms, microorganisms, genetics, viruses, various fungi, etc. It can also be used forOscillatory leaching of environmental water samples, soil samples, and solid waste, in compliance with the toxicity leaching equipment for solid waste leaching-Chinese Environmental Protection Standards(HJ577-2009)Equipment standards are applied in industries related to environmental protection, solid waste disposal, and toxicity identification, research, treatment, and disposal of solid waste.
2、 Technical parameters:
Specification Model |
SZD-9 |
SZD-12 |
oscillation frequency |
0~180time/part(Stable in)110time/Divide ±10) |
|
Oscillation amplitude |
40mm (reciprocating) |
|
timing range |
0~999minute |
|
Bottled quantity |
9only |
12only |
Overall dimensions |
485×485×300mm |
640×485×300mm |
power supply voltage |
220V 50HZ |
|
3、 Usage:
Before use, place the instrument on a sturdy workbench or ground and keep it stable. The usage environment should be clean, dry, and well ventilated. When installing test bottles on the oscillator tray, avoid placing heavy test bottles on one end of the tray. The containers should be evenly placed, and the culture medium in each bottle should be roughly equal. Connect the external power supply and place the power switch in“ON”Position, indicator light on, set the desired speed and time, detailed operation with instrument instructions attached!
4、 Instrument Description:
Example: Set the oscillator to every minute200Next job120Stop working in minutes.
1Turn on the power switch, and the upper row will display the original speed value.(Last run speed)
The bottom row displays the original timing values.
2According to the ︾(Speed fast/slow)Set the speed to200.
3Press ▲ (Up key) Set the timing to120.
4Press the ▼ (down arrow) to adjust the timing downwards (setting 0 points indicates continuous operation with an indefinite opening)
5Press the RUN button to start the machine, and it will slowly start until it reaches 200 revolutions per minute. At the same time, the timer will start (countdown), and then run steadily for 120 minutes before stopping. The END alarm will be displayed for 10 seconds. Timed operation completed.
6Speed and time adjustments can be made during both standby and timed operation.
7: PressSTOP(stop button)It can be stopped at any time (at which point the scheduled time also ends).
8:The green light is the running light, and the red light is the stop light. The corresponding indicator light is on, indicating that the controller is in the corresponding state.
5、 Attention:
1Before use, please test the machine with no load to see if it is working properly, and then proceed with container testing.
2Before each use, check whether all screws are tightened.
3If the number of reagent bottles is insufficient, they can be placed symmetrically.
4When oscillating, the solution inside the bottle should not exceed that of the reagent bottle1/3For good.
5Medium speed use is beneficial for extending the service life.
6Local use220V 50HzSo please use a three hole socket with safety grounding.
7、Please keep the instructions properly for future reference. In case of any discrepancy between the text description or image display and the actual product, the actual product shall prevail.