Residual voltage tester/Residual voltage tester/residual voltage testerH18453
one Equipment usage:
Residual voltage testerAccording to Chapter 22.5 of the national standard GB4706.1-2005 "Safety Part of Household and Similar Electrical Appliances: Safety Requirements" and GB9706.1-2007 "Medical Electrical Equipment Part: General Requirements for Safety", equipment connected to the power grid with a plug must be designed for voltage between each power pin and between each power pin and the equipment casing 1 second after unplugging the plug. ”Design. This instrument has a sufficiently high test impedance and uses the method of simulating the disconnection of the plug to disconnect the device from the grid voltage at peak state, for residual voltage testing of electrical equipment such as medical devices and household appliances.
2、 Main technical parameters:
1. Output voltage: AC 0-245V;
2. Test voltage: 0-150V;
3. Test time: 1 second;
4. Test impedance: 1000 M Ω;
5. Test alarm range: 60/34V;
6. Output power: 2000W;
7. Test discrimination: Sound and light alarm, while cutting off voltage;
8. Display mode: touch screen
9. Used for residual voltage testing between power plugs and between plugs and enclosures of equipment specified in GB9706.1-2007 "Medical Electrical Equipment Part: General Requirements for Safety" and GB4706.1-2005 "Safety Part: Safety Requirements for Household and Similar Electrical Appliances", as well as residual voltage testing of other parts of the equipment.
10. Equipment size approximately: length 450X width 450X height 200mm
11. Working power supply: AC220V/50Hz
2.Particle abrasion tester/Particle abrasion tester/Fertilizer catalysts, molecular sieves, adsorbents, etcH18452
Overview
Welcome to use our company's particle abrasion tester.
This instrument is an upgraded product of the DP-4 wear tester, with improvements made to the operating system and driver program. Suitable for measuring the wear rate of particulate materials such as catalysts, molecular sieves, alumina, adsorbents, etc. in industries such as chemical, pharmaceutical, and petroleum.
This instrument has complete functions, including preset grinding drum speed, grinding drum rotation time, vibrating screen vibration frequency, and vibrating screen vibration time, which can be selected according to different experimental conditions.
Reliable performance, simple and intuitive operation.
Technical Specifications
(1). Working environment: -10C-60C, humidity ≤ 85%
(2). Power supply: AC 220V ± H10% 50 Hz, power ≤ 100W
(3) Wear test:
Motor speed: 25/30/50/60/80/100/125 r/min, 7 options in total
Counting range: 1~9999 r
Grinding cylinder diameter: 36x300mm/50x300mm (inner diameter).
Two types of grinding cylinders are available for selection, which can test two parallel samples simultaneously.
Screening test:
Screening frequency: 1-9HZ
Screening time: 1-9999sec
3.Filter material integrity tester/integrity tester/bubble point tester Model:H18451
1、 Main technical indicators
Input voltage: 100-240VAC. 50/60 Hz
Rated power: 30W Insurance fuse: less than 2A
Basic dimensions: length, height, and depth 255 * 160 * 360 mm
Working temperature: -10~55 ℃
Storage temperature: -20~65 ℃
Environmental humidity:<95% RH (no icing, no condensation) ≠
Time setting: 1~9999 minutes
Detection accuracy: 0.2 level (higher than the national standard of 0.5 level)
Detection range: 0-600 kPa Resolution 0.1 kPa:
Gas source requirements: 1. Use dry and clean compressed air or nitrogen;
2. The intake pressure is 0.4-0.8 MPa.
2、 Instrument usage instructions
1. The pressure of the gas source should be 100-200 kPa higher than the required pressure for detection, not exceeding 800 kPa; four
2. After connecting the gas source, there is a slight leakage sound inside the instrument, which belongs to normal pressure regulation and exhaust, not a fault;.
3. Try to use it in situations with small temperature gradient changes, while avoiding strong vibrations and impacts; four
4. Do not use in situations with strong electromagnetic interference, otherwise it may cause uncontrollable deviation of the instrument;
5. Please be careful to avoid liquid entering the interior of the instrument during operation
4.Handheld surface copper tester/surface copper tester/surface copper detectorH18450
1. Product Overview
Handheld copper thickness tester, accurately grasping customer needs in the circuit board industry
2. Instrument principle:
The handheld surface copper tester adopts micro resistance testing technology, using four contact probes to generate electrical signals on the surface copper foil for measurement. When the probe contacts the copper foil sample, a constant current passes through the outer two probes, and the inner two probes measure the change in voltage.
According to Ohm's law, the voltage value is converted into a resistance value, and the thickness value is calculated using a certain function, which is not affected by the insulation layer and the copper layer on the back of the circuit board.
2. Product Usage:
Conduct relevant copper foil thickness testing before and after PCB and FPC cutting, drilling, electroplating, etching, and other processes.
3. Product features:
1) Two point calibration and multi-point calibration can be selected to achieve accurate measurement of copper thickness over a large range without the need to switch files.
2) Handheld copper thickness tester.
3) Handheld design, easy to carry.
4) The curvature of the test head and screen is easy to observe.
5) Easy to operate and user-friendly.
6) Replaceable probe, economical and practical.
7) Probe with illumination for easy positioning.
4. Technical indicators and parameters:
Ø Test range: 5-254 μ m (0.197-10mil)
Ø ruler Dimensions: 125mm * 52mm * 35mm (flat: length * width * thickness)
Ø Heavy Quantity: 100g
Ø Single Position: Data display unit can be selected mil、μm、oz
Ø Operating language: The instrument operation interface has two languages: English and Chinese
Calibration method: two-point calibration or multi-point calibration, no need to switch test files.
Ø Test mode: Single or continuous measurement mode can be selected
Ø Electricity Source: Powered by a regular AA (size 5) battery
Ø Data interface: The USB interface is used to download test data to the computer.
Ø Display: Color with backlight TFT2.0 LCD display screen, 240 × 320 pixels
Ø Data statistics: Automatic data recording, reminders for upper and lower limits of large and small values, mean, and standard deviation.
Ø Testing accuracy: ± 3% (based on standard film), testing product is ± 5%, with temperature sensing.
Ø Testing speed: 0.5S/time
Ø Minimum resolution: 0.01um
5.Immersion pulling coating machine model:H18449
It is a precision instrument suitable for preparing thin films by dip coating method, which is one of the common methods for preparing thin films by sol-gel method. Using high-quality accessories, precise control of lifting speed, and repeated coating without positional deviation. The pulling speed, lifting height, number of coating cycles, immersion time, descent speed, and dwell time can all be precisely controlled through the program. Suitable for glass SI(100)、SI(111)、 Commonly used substrates such as sapphire (AL203), porcelain tiles, and resin substrates.
Technical Specifications:
Lifting stroke: 0~300mm, subject to actual conditions, optional 600mm
Range of lifting speed 1~9999um/s
Position accuracy <0.1mm
speed and accuracy ±0.01%
Immersion time 1~3200s
Coating frequency 1-1000 times
Coating interval time 1~3200s
Power supply 220V 50Hz
The above parameter information corresponds to the image