| type | Multi group and multi-mode ultrasonic flaw detector | Acoustic Technical Specifications | |||
| Dimensions (width x height x thickness) | 335 mm × 221 mm × 151 mm | pulse generator | PA channel | UT channel | |
| weight | 5.7 kg (including 1 battery) | voltage | 40 V、80 V、115 V | 85 V、155 V、295 V | |
| Hard drive capacity | 64 GB built-in SSD, can also expand storage capacity with external USB drive | pulse width | Adjustable within the range of 30 ns to 500 ns, with a resolution of 2.5 ns | Adjustable within the range of 30 ns to 1000 ns, with a resolution of 2.5 ns | |
| storage device | SDHC cards and SDXC cards, or most standard USB storage devices | fall time | < 10 ns | < 10 ns | |
| Maximum onboard file capacity | 25 GB | Pulse shape | Negative square wave pulse | Negative square wave pulse | |
| GPS | Yes (unless otherwise specified for certain regions) | output impedance | Pulse echo mode: 28 Ω | < 30 Ω | |
| call the police | 3 of them | One send one receive mode: 24 ohms | |||
| wireless connection | Yes (USB adapter sold separately as an accessory) | receiver | PA channel | UT channel | |
| PA interface | 1 interface | Gain range | 0 dB ~ 80 dB, Maximum input signal; 550 mVp-p (full screen height) | 0 dB ~ 120 dB, Maximum input signal; 34.5 Vp-p (full screen height) | |
| UT interface | 4(2 P/R 通道) | input impedance | Pulse echo mode, at 9 MHz: 57 Ω± 10% | Pulse echo mode: 50 Ω | |
| certification | ISO 18563-1:2015 | One send one receive mode, at 9 MHz: 100 Ω± 10% | Pulse transmission and reception mode: 50 ohms | ||
| EN12668-1:2010 | System bandwidth | 0.5 MHz ~ 18 MHz | 0.25 MHz ~ 28 MHz | ||
| display | Sound beam formation | PA channel | UT channel | ||
| type | TFT LCD (Thin Film Transistor Liquid Crystal Display), resistive touch screen | Scan type | Single, linear, fan-shaped, hybrid, and fully focused modes (TFM) | - | |
| size | 269 millimeters (10.6 inches) | Maximum aperture | OMNIX3-PATFM1664PR=16 chips | - | |
| resolution | 1280 × 768 pixels | OMNIX3-PATFM16128PR=16 chips | |||
| number of colors | 16 million | OMNIX3-PATFM32128PR=32 chips | |||
| viewing angle | Level: -85 ° to 85 ° | Number of chips received | OMNIX3-PATFM1664PR=64 receiver chip | - | |
| Vertical: -85 °~85 ° | OMNIX3-PATFM16128PR=128 receiver chip | ||||
| OMNIX3-PATFM32128PR=128 receiver chip | |||||
| The number of focusing rules | tallReaching 1024 | - | |||
| Input/Output (I/O) ports | Delay range of launch | 0 µ s to 10 µ s, increments of 2.5 ns | - | ||
| USB 2.0 | 2 ports (1 located behind the battery) | Delay range of reception | 0 µ s to 6.4 µ s, increments of 2.5 ns | - | |
| USB 3.0 | 1 port | ||||
| video output | Video output (HDMI) | TFM/FMC | |||
| memory card | SDHC port | Supported modes | Pulse echo: L-L, TT, and TT-TT | ||
| communication | Ethernet | One send, one receive: LL-L, TT-T, TT-L, TL-T, LT-T, TTT-TT, and TL-L | |||
| Input/Output (I/O) Line | Number of sound beam groups | Simultaneously display up to 4 TFM groups | |||
| encoder | Dual axis encoder line (orthogonal or clock/direction), can be connected to a third encoder | Maximum aperture | OMNIX3-PATFM1664PR=32 chip extended aperture | ||
| digital input | 6 numerical inputs, TTL | OMNIX3-PATFM166128PR=32 chip extended aperture | |||
| digital output | 5 digital outputs, TTL | OMNIX3-PATFM32128PR=64 chip extended aperture | |||
| collection switch | Configure 1 digit input as the acquisition switch | image resolution | Up to 1024 × 1024 (1 M point), for each fully focused mode (TFM) sound wave group | ||
| Power output line | 5 V rated, 1 A (short circuit protection), 12 V output at 1 A | Real time full focus mode (TFM) envelope | have | ||
| Operating environment | |||||
| Phased array flaw detectorExternal DC power supply | Intrusion protection rating | Compliant with IP65 rating standards (* dustproof and resistant to water jets from all directions, 6.3mm nozzle) | |||
| DC input (DC-IN) voltage | 15 VDC to 18 VDC (minimum 50 W) | Anti collision rating | Through the drop test of MIL-STD-810G | ||
| interface | Circular, 2.5mm pin diameter, center positive pole | intended use | Indoor and outdoor use | ||
| altitude | Up to 2000 meters | ||||
| battery | Operating temperature | 0 °C ~ 45 °C | |||
| type | lithium-ion battery | Storage temperature | -20 ° C to 60 ° C (including battery) | ||
| capacity | 93 Wh | -20 ° C to 70 ° C (excluding battery) | |||
| Number of batteries | 2 of them | ||||
| runtime | 2 batteries run for 5 hours (with hot swappable performance) | ||||
| Programmable TCG | |||||
| PA/UT configuration | number of points | 32, each focusing rule has a TCG (Time Correction Gain) curve | |||
| Depth of Position | 16 bits | scope | Phased array (standard): 40 dB, step pitch of 0.1 dB | ||
| Maximum Pulse Repetition Frequency (PRF) | 20 kHz | Phased array (extended): 65 dB, step size of 0.1 dB | |||
| Conventional ultrasound: 100 dB, step size of 0.1 dB | |||||
| frequency | Maximum slope | Phased array (standard): 40 dB/10 ns | |||
| Effective digital frequency | Maximum 100 MHz | Phased array (extended): 0.1 dB/10 ns | |||
| Conventional ultrasound: 40 dB/10 ns | |||||
| display | |||||
| 刷新率 | A-scan: 60 Hz; S-scan: 20 Hz to 30 Hz | process | |||
| Envelope (echo dynamic mode) | Yes: Volume corrected S-scan (30 Hz) | The maximum number of scanned data points for A | tallReaching 16384 | ||
| A-scan height | Up to 800% | Real time average | PA:2、4、8、16 | ||
| UT:2、4、8、16、32、64 | |||||
| synchronize | detection | RF, full wave, positive half wave, negative half wave | |||
| According to the internal clock | 1 Hz ~ 10 kHz | filtering | PA channel: 3 low-pass, 6 band-pass, and 4 high pass filters | ||
| External walking speed | have | UT channel: 8 low-pass, 6 band-pass, and 4 high pass filters (3 low-pass filters when configured as TOFD) | |||
| According to the encoder | Dual axis: 1 step to 65536 steps | Video filtering | Smooth (adjusted according to the frequency range of the probe) | ||
| Data Technical Specifications | |||||
| The maximum number of scanned data points for A | tallReaching 16384 | ||||
| Real time average | PA:2、4、8、16 | ||||
| UT:2、4、8、16、32、64 | |||||
| detection | RF, full wave, positive half wave, negative half wave | ||||
| filtering | PA channel: 3 low-pass, 6 band-pass, and 4 high pass filters | ||||
| UT channel: 8 low-pass, 6 band-pass, and 4 high pass filters (3 low-pass filters when configured as TOFD) | |||||
| Video filtering | Smooth (adjusted according to the frequency range of the probe) | ||||
| Programmable TCG | |||||
| number of points | 32, each focusing rule has a TCG (Time Correction Gain) curve | ||||
| scope | Phased array (standard): 40 dB, step pitch of 0.1 dB | ||||
| Phased array (extended): 65 dB, step size of 0.1 dB | |||||
| Conventional ultrasound: 100 dB, step size of 0.1 dB | |||||
| Maximum slope | Phased array (standard): 40 dB/10 ns | ||||
| Phased array (extended): 0.1 dB/10 ns | |||||
| Conventional ultrasound: 40 dB/10 ns | |||||
Shanghai Jiuheng Instrument Co., Ltd. is a professional enterprise engaged in the research and development, production, sales, and service of instruments and meters. Ultrasonic flaw detectors: EPOCH 600 ultrasonic flaw detector, EPOCH 6LT ultrasonic flaw detector, EPOCH 650 ultrasonic flaw detector, EPOCH 1000 ultrasonic flaw detector, USM 36 ultrasonic flaw detector, USN 60 ultrasonic flaw detector, USM go+ultrasonic flaw detector Ominscan SXPhased array flaw detector、 OmniScan MX2 phased array TOFD flaw detector, new OmniScan MX3 fully focused phased array ultrasonic flaw detector); Ultrasonic thickness gauge: (27MG ultrasonic thickness gauge, 45MG ultrasonic thickness gauge, 38DL PLUS ultrasonic thickness gauge, DM5E ultrasonic thickness gauge, CL5 ultrasonic thickness gauge, MAGNA-MIKE 8600 Hall effect thickness gauge, ETG-100 electromagnetic high-temperature thickness gauge); Ultrasonic probe; Customized phased array probe; Eddy current flaw detector: (NORTEC 600 eddy current flaw detector); Hardness tester; DR flat panel detector: 1313 DX ,1515DXT-I ,2520DX ,2530HE,4343HE, Dexela 1207 ,Dexela 1313 ,Dexela 2121S ,Dexela CMOS Flat panel detector, XRD 0822 ,XRD 1611 ,XRD 1621 ,XRD 4343RF ,XRD 3025 ,XRpad 4336 ,XRpad 4343 F ,XRpad2 3025 ,XRpad2 4336。 Endoscopes: (IPLEX NX Olympus Endoscope, IPLEX GX/GT Olympus Endoscope, IPLEX G Lite Olympus Endoscope, IPLEX TX Olympus Endoscope, IPLEX YS Olympus Endoscope, XLGo+Industrial Video Endoscope GE, XL Detect+Industrial Video Endoscope GE New, UVin Professional UV Video Endoscope); Pulse generator: (DRP300 replaces Olympus' 5072PR, 5073PR); Alloy analyzer: (Vanta Olympus handheld alloy analyzer, DELTA Professional handheld alloy analyzer Olympus, DELTA Element handheld alloy analyzer Olympus) and other instruments and meters