Phased array flaw detector UT channel: 8 low-pass, 6 band-pass, and 4 high pass filters (3 low-pass filters when configured as TOFD)
Product Details
type Multi group and multi-mode ultrasonic flaw detector Acoustic Technical Specifications
Dimensions (width x height x thickness) 335 mm × 221 mm × 151 mm pulse generator PA channel UT channel
weight 5.7 kg (including 1 battery) voltage 40 V、80 V、115 V 85 V、155 V、295 V
Hard drive capacity 64 GB built-in SSD, can also expand storage capacity with external USB drive pulse width Adjustable within the range of 30 ns to 500 ns, with a resolution of 2.5 ns Adjustable within the range of 30 ns to 1000 ns, with a resolution of 2.5 ns
storage device SDHC cards and SDXC cards, or most standard USB storage devices fall time < 10 ns < 10 ns
Maximum onboard file capacity 25 GB Pulse shape Negative square wave pulse Negative square wave pulse
GPS Yes (unless otherwise specified for certain regions) output impedance Pulse echo mode: 28 Ω < 30 Ω
call the police 3 of them One send one receive mode: 24 ohms
wireless connection Yes (USB adapter sold separately as an accessory) receiver PA channel UT channel
PA interface 1 interface Gain range 0 dB ~ 80 dB, Maximum input signal; 550 mVp-p (full screen height) 0 dB ~ 120 dB, Maximum input signal; 34.5 Vp-p (full screen height)
UT interface 4(2 P/R 通道) input impedance Pulse echo mode, at 9 MHz: 57 Ω± 10% Pulse echo mode: 50 Ω
certification ISO 18563-1:2015 One send one receive mode, at 9 MHz: 100 Ω± 10% Pulse transmission and reception mode: 50 ohms
EN12668-1:2010 System bandwidth 0.5 MHz ~ 18 MHz 0.25 MHz ~ 28 MHz
display Sound beam formation PA channel UT channel
type TFT LCD (Thin Film Transistor Liquid Crystal Display), resistive touch screen Scan type Single, linear, fan-shaped, hybrid, and fully focused modes (TFM) -
size 269 millimeters (10.6 inches) Maximum aperture OMNIX3-PATFM1664PR=16 chips -
resolution 1280 × 768 pixels OMNIX3-PATFM16128PR=16 chips
number of colors 16 million OMNIX3-PATFM32128PR=32 chips
viewing angle Level: -85 ° to 85 ° Number of chips received OMNIX3-PATFM1664PR=64 receiver chip -
Vertical: -85 °~85 ° OMNIX3-PATFM16128PR=128 receiver chip
OMNIX3-PATFM32128PR=128 receiver chip
The number of focusing rules tallReaching 1024 -
Input/Output (I/O) ports Delay range of launch 0 µ s to 10 µ s, increments of 2.5 ns -
USB 2.0 2 ports (1 located behind the battery) Delay range of reception 0 µ s to 6.4 µ s, increments of 2.5 ns -
USB 3.0 1 port
video output Video output (HDMI) TFM/FMC
memory card SDHC port Supported modes Pulse echo: L-L, TT, and TT-TT
communication Ethernet One send, one receive: LL-L, TT-T, TT-L, TL-T, LT-T, TTT-TT, and TL-L
Input/Output (I/O) Line Number of sound beam groups Simultaneously display up to 4 TFM groups
encoder Dual axis encoder line (orthogonal or clock/direction), can be connected to a third encoder Maximum aperture OMNIX3-PATFM1664PR=32 chip extended aperture
digital input 6 numerical inputs, TTL OMNIX3-PATFM166128PR=32 chip extended aperture
digital output 5 digital outputs, TTL OMNIX3-PATFM32128PR=64 chip extended aperture
collection switch Configure 1 digit input as the acquisition switch image resolution Up to 1024 × 1024 (1 M point), for each fully focused mode (TFM) sound wave group
Power output line 5 V rated, 1 A (short circuit protection), 12 V output at 1 A Real time full focus mode (TFM) envelope have
Operating environment
Phased array flaw detectorExternal DC power supply Intrusion protection rating Compliant with IP65 rating standards (* dustproof and resistant to water jets from all directions, 6.3mm nozzle)
DC input (DC-IN) voltage 15 VDC to 18 VDC (minimum 50 W) Anti collision rating Through the drop test of MIL-STD-810G
interface Circular, 2.5mm pin diameter, center positive pole intended use Indoor and outdoor use
altitude Up to 2000 meters
battery Operating temperature 0 °C ~ 45 °C
type lithium-ion battery Storage temperature -20 ° C to 60 ° C (including battery)
capacity 93 Wh -20 ° C to 70 ° C (excluding battery)
Number of batteries 2 of them
runtime 2 batteries run for 5 hours (with hot swappable performance)
Programmable TCG
PA/UT configuration number of points 32, each focusing rule has a TCG (Time Correction Gain) curve
Depth of Position 16 bits scope Phased array (standard): 40 dB, step pitch of 0.1 dB
Maximum Pulse Repetition Frequency (PRF) 20 kHz Phased array (extended): 65 dB, step size of 0.1 dB
Conventional ultrasound: 100 dB, step size of 0.1 dB
frequency Maximum slope Phased array (standard): 40 dB/10 ns
Effective digital frequency Maximum 100 MHz Phased array (extended): 0.1 dB/10 ns
Conventional ultrasound: 40 dB/10 ns
display
刷新率 A-scan: 60 Hz; S-scan: 20 Hz to 30 Hz process
Envelope (echo dynamic mode) Yes: Volume corrected S-scan (30 Hz) The maximum number of scanned data points for A tallReaching 16384
A-scan height Up to 800% Real time average PA:2、4、8、16
UT:2、4、8、16、32、64
synchronize detection RF, full wave, positive half wave, negative half wave
According to the internal clock 1 Hz ~ 10 kHz filtering PA channel: 3 low-pass, 6 band-pass, and 4 high pass filters
External walking speed have UT channel: 8 low-pass, 6 band-pass, and 4 high pass filters (3 low-pass filters when configured as TOFD)
According to the encoder Dual axis: 1 step to 65536 steps Video filtering Smooth (adjusted according to the frequency range of the probe)
Data Technical Specifications
The maximum number of scanned data points for A tallReaching 16384
Real time average PA:2、4、8、16
UT:2、4、8、16、32、64
detection RF, full wave, positive half wave, negative half wave
filtering PA channel: 3 low-pass, 6 band-pass, and 4 high pass filters
UT channel: 8 low-pass, 6 band-pass, and 4 high pass filters (3 low-pass filters when configured as TOFD)
Video filtering Smooth (adjusted according to the frequency range of the probe)
Programmable TCG
number of points 32, each focusing rule has a TCG (Time Correction Gain) curve
scope Phased array (standard): 40 dB, step pitch of 0.1 dB
Phased array (extended): 65 dB, step size of 0.1 dB
Conventional ultrasound: 100 dB, step size of 0.1 dB
Maximum slope Phased array (standard): 40 dB/10 ns
Phased array (extended): 0.1 dB/10 ns
Conventional ultrasound: 40 dB/10 ns

Shanghai Jiuheng Instrument Co., Ltd. is a professional enterprise engaged in the research and development, production, sales, and service of instruments and meters. Ultrasonic flaw detectors: EPOCH 600 ultrasonic flaw detector, EPOCH 6LT ultrasonic flaw detector, EPOCH 650 ultrasonic flaw detector, EPOCH 1000 ultrasonic flaw detector, USM 36 ultrasonic flaw detector, USN 60 ultrasonic flaw detector, USM go+ultrasonic flaw detector Ominscan SXPhased array flaw detector、 OmniScan MX2 phased array TOFD flaw detector, new OmniScan MX3 fully focused phased array ultrasonic flaw detector); Ultrasonic thickness gauge: (27MG ultrasonic thickness gauge, 45MG ultrasonic thickness gauge, 38DL PLUS ultrasonic thickness gauge, DM5E ultrasonic thickness gauge, CL5 ultrasonic thickness gauge, MAGNA-MIKE 8600 Hall effect thickness gauge, ETG-100 electromagnetic high-temperature thickness gauge); Ultrasonic probe; Customized phased array probe; Eddy current flaw detector: (NORTEC 600 eddy current flaw detector); Hardness tester; DR flat panel detector: 1313 DX ,1515DXT-I ,2520DX ,2530HE,4343HE, Dexela 1207 ,Dexela 1313 ,Dexela 2121S ,Dexela CMOS Flat panel detector, XRD 0822 ,XRD 1611 ,XRD 1621 ,XRD 4343RF ,XRD 3025 ,XRpad 4336 ,XRpad 4343 F ,XRpad2 3025 ,XRpad2 4336。 Endoscopes: (IPLEX NX Olympus Endoscope, IPLEX GX/GT Olympus Endoscope, IPLEX G Lite Olympus Endoscope, IPLEX TX Olympus Endoscope, IPLEX YS Olympus Endoscope, XLGo+Industrial Video Endoscope GE, XL Detect+Industrial Video Endoscope GE New, UVin Professional UV Video Endoscope); Pulse generator: (DRP300 replaces Olympus' 5072PR, 5073PR); Alloy analyzer: (Vanta Olympus handheld alloy analyzer, DELTA Professional handheld alloy analyzer Olympus, DELTA Element handheld alloy analyzer Olympus) and other instruments and meters