JCB5085-02R Petroleum Product Thermal Stability Tester meets the requirements of HG/T5085-2016 (latest version) "Rubber Plasticizers - Determination of Thermal Stability of Cycloalkyl Mineral Oil Appendix B", and is used for the determination of thermal and optical stability of rubber plasticizers - Cycloalkyl mineral oil, rubber oil, white oil, and other light colored petroleum products with similar requirements.
Introduction
JCB5085-02R typePetroleum product thermal stability testerMeets the requirements of HG/T5085-2016 (latest version) "Rubber Plasticizers - Determination of Thermal Stability of Alkyl Mineral Oils - Appendix B", used for the determination of thermal and optical stability of rubber plasticizers - Alkyl mineral oils, rubber oils, white oils, and other light colored petroleum products with similar requirements.
Two main technical indicators
1. Input voltage: AC 220V ± 10% 50HZ
2. Temperature range: Room temperature -250 ℃
3. Rated power: 2000w
4. Temperature control accuracy: ± 1 ℃
5. Sample quantity: 8 pieces
Instructions for using the three instruments
1. After unpacking the instrument, please refer to the packing list to check if the accessories are complete.
2. Place the thermal stability test chamber on a horizontal table and keep the turntable in a horizontal position.
3. Take 120ml of sample separately from the measuring cylinder, pour it into a clean and dry test cup, place it in the position of the test cup on the turntable of the test chamber, install the thermometer on the thermometer holder, and insert the mercury head of the thermometer into the sample to measure the temperature of the sample. Close the door of the test chamber and turn on the power.
4. Turn on the power switch, the temperature control meter starts working, set the temperature on the temperature control meter, and the box begins to heat up. Turn on the start switch, and the turntable will start working.
5. When the temperature of the temperature control meter reaches the set temperature, adjust the timer time to 4 hours, turn on the timer switch, and the timer starts to work.
6. After the timing is completed, the timer will sound an alarm reminder. This should be done by sequentially turning off the timer, starting, and power switches, and opening the box door to allow the sample to cool naturally for 30 minutes. Then take out the sample and cool it to room temperature in a dryer, and measure the color of the sample according to GB/T3555.
7. During the operation of the test chamber, the temperature of the sample should be recorded every hour. When observing the temperature of the sample, turn on the lighting, turn off the motor, stop the turntable in a convenient position for observation, and observe the thermometer through the observation window. Do not open the test door, and the observation time should not exceed 1 minute each time. After observation, quickly start the turntable and turn off the lighting.
Four precautions
1. This instrument must be reliably grounded before use.
2. When the instrument malfunctions, it should be handled by professionals and should not be disassembled at will.
3. Keep the instrument clean, prevent oil samples and other dirt from splashing on the instrument, and avoid water and moisture.
Five common faults and troubleshooting methods
Fault occurrence |
Possible reasons |
treatment method |
The temperature control meter is not displaying properly |
Sensor malfunction |
Replace the sensor with the same model |
Turn on the power switch and trip |
Heating tube damaged |
Replace the heating tube |
Packing List
Instrument name:Petroleum product thermal stability tester Factory number: 1571210324 Instrument model: JCB5085-02R |
Accessory name |
quantity |
host |
1 unit |
Test cup |
16 of them |
Temperature gauge 155-170 ℃ |
1 unit |
Instruction Manual |
1 copy |
Certificate of Conformity |
1 sheet |
Packing List |
1 copy |
|
|
|
|
|
|
Inspector: Packer: