The WM-2000 fully automatic open flash point tester follows the technical standard GB/T3536-2008 and is compatible with GB267 ASTM D92 ISO2592 ASTMD1310. It can be widely used in departments such as petroleum, chemical, metallurgical, power, transportation inspection, and scientific research.
Instrument features:
1. WM-2000Fully automatic open flash point testerImplement the technical standard GB/T3536-2008 and be compatible with GB267 ASTM D92 ISO2592 ASTMD1310. It can be widely used in departments such as petroleum, chemical, metallurgical, power, transportation inspection, and scientific research.
2. The core host adopts TI's AM3354 processor, Cortex-A8 core, and a 1GHz clock speed; The operating system adopts Windows Embedded Compact 7 real-time industrial control system. Abandoning the use of non nuclear, non operational microcontrollers, the modernization of instrument control has been truly achieved, ushering instruments into a new era of intelligence.
3. WM-2000Fully automatic open flash point testerThe monitor adopts the original 7.0-inch 800 × 480 pixel true color TFT-LCD display screen from Innolux; The keyboard adopts a human body sensing touch screen. The fully Chinese operating interface displays delicate, intuitive, and elegant graphics, making it easy to control and touch freely.
4. Historical data storage adopts FLASH data storage, which can store 2010 historical data, and the stored data cannot be changed.
5. The printer adopts a micro embedded thermal printer, which prints quieter, faster, and clearer; Customized needle printers can also be used, and the printed results can be stored for a long time without fading.
6. Rear USB and 232 serial port, can be connected to a computer and later software upgrades, can export experimental data to a computer and USB drive.
7. Temperature measurement and control adopt imported PT100 platinum resistance temperature sensor, high-precision AD converter, excellent linearization mathematical model, and control algorithm, making temperature measurement and control faster, more accurate, and stable.
8. The ignition adopts electronic flame direct ignition, which has the same effect as gas flame ignition, safe, convenient, fast, reliable, and consumes no fuel.
9. Flash fire detection adopts high-frequency ion ring flame detection technology, which enables rapid and accurate capture of sample flash fire instantly, avoiding false detection and missed detection.
10. The atmospheric pressure measurement adopts a fully digital structure atmospheric pressure sensor imported from Germany, which measures the local atmospheric pressure in real time and automatically corrects the impact of atmospheric pressure changes on the measurement data.
11. Automatically complete heating, detection, calculation, printing and other operations.
Product Technical Parameters:
Measurement range: 79 ℃~400 ℃
Measurement accuracy: ± 0.1 ℃
Repeatability: ≤ 4 ℃
Ignition method: High voltage self-excited arc direct ignition
Flash fire detection: using pulse high-voltage ion ring flame detection technology.
Data storage: It can store up to 2010 historical data, cannot be deleted, and will be retained in case of power failure.
Cooling method: Strong air cooling
Environmental temperature: 10-45 ℃
Environmental humidity: ≤ 85%
Power supply voltage: 220 ± 10% V.AC
Power frequency: 50 ± 10% Hz
Power consumption: 700W
Dimensions: 500 * 440 * 340
Instrument weight: 20Kg