Welcome Customer !

Membership

Help

Beijing Guancheng Measurement Technology Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

PMC531、 534 type pressure transmitter

NegotiableUpdate on 05/16
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

PMC531、 534 type pressure transmitter

Product Details


Product Features

Imported dry ceramic capacitive sensors are used. Good temperature characteristics and long-term stability. Obtain high reliability and measurement accuracy.

On the basis of PMC133 standard pressure transmitter, an oil filled welded sealed isolation diaphragm structure is added. Avoid direct contact pressure of the medium

The measuring diaphragm and sealing material of the force transmitter provide a reliable way for pressure measurement under special conditions.

Oil filled welding sealing isolation device, with multiple different process connection structure forms and can be made of various materials, especially suitable for

High temperature, low temperature, strong corrosion, high viscosity, and crystalline media; Suitable for clean and explosive hazardous environments. Having a negative table

Pressure measurement.

working principle

As shown in the figure, the pressure of the measured medium is applied to the oil filled sealing isolation diaphragm, and the process pressure is applied to the sensor through the intermediate fluid.

1Intermediate fluid inside the diaphragm membrane

2Transmit pressure to the ceramic measuring diaphragm

3The weak offset of the capacitor electrode results in a change in capacitance

4. The connected circuit converts the change in capacitance into an electrical signal output




Technical Specifications

·PMC531measurement range

Relative pressure: Maximum measurement range:0~7MPa;Minimum measurement range: 0~20KPa

Negative relative pressure: Maximum measurement range: -0.1MPa~7MPaMinimum measurement range:10Kpa~10Kpa

Absolute pressure: Maximum measurement range:0~7MPa;Minimum measurement range: 0~20KPa

·PMC534measurement range

Relative pressure: Maximum measurement range:0~20MPa;Minimum measurement range: 0~20KPa

Negative relative pressure: Maximum measurement range: -0.1MPa~7MPaMinimum measurement range:-10KPa~10KPa

Absolute pressure: Maximum measurement range: 0~20MPaMinimum measurement range: 0~20KPa

·Measurement accuracy

Comprehensive error (including linearity, repeatability, hysteresis): ± 0.2% FS or ± 0.5% FS

·Long term stability

Better than 0.1% F.s per year

·Allow ambient temperature

(-40~80)℃

·Allow storage temperature

(-40~120)℃

·Environmental impact

(Excluding temperature coefficient of oil filled isolation diaphragm) -20 ℃~+80 ℃ ± 0.15%/10K

-20℃~-40℃ ±0.25%/10K; -80℃~+120℃ ±0.25%/10K

·Structure and Materials

(Excluding oil filled isolation membrane structure)

Shell: plastic or die cast aluminum

Size: Approximately 70 × 70 × 130mm

Protection level: IP65 [DIN40050]

Weight: 0.6Kg to 0.8Kg

·Sealing components

(Excluding oil filled isolation membrane structure) Fluororubber, fluorosilicone EPDM、PTFE、KALRE




Exterior structure (PMC531 type)



Exterior structure (PMC534 type)


electrical connection

Terminal block: M3 screw, cable hole: internal thread M20 * 1.5, cable sealing sleeve: external thread M20 * 1.5
The function of a chain diode is to ensure the output current of the measuring instrument without cutting off the signal line during operation.
The wiring terminal with on-site display transmitter is placed at the back of the on-site display, and the display is connected through a metal socket and
When connecting the flexible cable to the transmitter, please gently pull up the monitor and plastic bracket together with the user's wiring
[Caution: Do not apply too much force to avoid disconnecting the flexible cable], the wiring terminal can be exposed.
[See the diagram of zero and range adjustment for details]

load characteristic

The analog output type feed voltage is 13V to 30V [excluding the cascading diode, allowing for 12.5 to 30V voltage].
The maximum load resistance depends on the instrument feed voltage.

Zero point and range adjustment




Order Selection Table