- Phone
-
Address
Shuangluan District Power Plant Industrial Park, Chengde City
Chengde Donghai Testing Machine Manufacturing Co., Ltd
Shuangluan District Power Plant Industrial Park, Chengde City
Oxygen Index Tester Chengde Donghai Testing MachineXCY-100 Oxygen Index Tester
Oxygen Index Tester Chengde Donghai Testing MachineIt is a new type of product developed according to the technical requirements specified in the national standard GB/T2406. This instrument is used to determine the volume percentage of oxygen required during the combustion process of polymers. The polymer oxygen index value is the lowest oxygen concentration (also known as the oxygen index) required to maintain combustion for 50mm or 180s in a mixture of oxygen and nitrogen after the substance is ignited.
This instrument can not only serve as a means of identifying the flame retardancy of polymers, but also as a research tool, providing powerful testing methods for laboratory research on flame retardant formulations and the development of new flame retardant materials. It is suitable for the flammability test of solid materials, laminated materials, foamed plastics, fabrics, films and films.
The instrument structure is designed reasonably and easy to operate, use, and maintain. The testing system uses imported oxygen sensors and displays the results digitally. It has the advantages of accurate judgment and good reproducibility, making it an ideal testing equipment for scientific research and production quality control.
Main technical indicators
1. Using imported oxygen sensors, the digital display of oxygen concentration does not require calculation, with higher and more accurate precision and a range of 0100%
2. Digital resolution: ± 0.1%
3. Measurement accuracy: 0.4 level
4. Flow regulation range: 0-10L/min (60-600L/h)
5. Response time:<5 seconds
6. Combustion cylinder: inner diameter ≥ 75mm, height 400mm
7. Gas flow velocity inside the combustion cylinder: 40mm ± 10mm/s
8. Pressure gauge accuracy level 2.5, resolution: 0.01MPa
9. Flow meter: adjustable from 1-15L/min (60-900L/H)
10. Oxygen/nitrogen mixed gas inlet: including stabilizing valve, flow regulating valve, gas filter, and mixing chamber.
11The sample holder can be used for soft and hard plastics, textiles, fire doors, etc. 12. Propane (butane) ignition system, flame length 5mm-60mm can be freely adjusted
13. Gas: Industrial nitrogen, oxygen, purity>99%; (User provided).
14. Igniter: It is made of a metal tube and has a nozzle with an inner diameter of Φ 2 ± 1mm at the end, which can be inserted into the combustion cylinder to ignite the sample. The flame length is 16 ± 4mm and the size is adjustable.
15. Self supporting material sample clamp: can be fixed at the axis position of the combustion cylinder and can vertically clamp the sample
16.Non self-supporting material specimen clamp: capable of fixing both vertical edges of the specimen to the frame simultaneously
Working conditions of the instrument
1. Environmental temperature: -10 ℃ to 45 ℃
2. Relative humidity:<85%;
3. Voltage and power: 220V, 50HZ, 100W
4. Gas used: GB3863 Industrial Gaseous Oxygen; GB3864 Gaseous nitrogen for industrial use; Two bottles of gasAll require pressure regulating valves.
5. Input pressure: 0.2-0.3MPa
6. Working pressure: Nitrogen 0.1-0.15Mpa Oxygen 0.1-0.15Mpa Operating procedure