- Phone
-
Address
Room 1206, No. 2, Lane 3620, Zhongshan North Road, Shanghai
Shanghai Boshitong Electric Appliance Co., Ltd
Room 1206, No. 2, Lane 3620, Zhongshan North Road, Shanghai
612The Type I online submersible water quality analysis and control device can perform long-term, unsupervised analysis and testing. Data can be transmitted remotely through telephone lines or wireless.
Conduct the following analysis using specially provided software
1. Set storage 2. View historical data
3. Trigger sampling 4. Verify stored data
5. Data Storage Options 6. Instant Data Storage
7. Multi parameter graphics of data 8. Extracting data
9. Calibrate the sensor
Main technical parameters
·Outer packaging: PVC plastic, formaldehyde plastic, and 316 stainless steel
·Storage capacity: 130K, 12 bits
·Storage format: 16300 sets, all sample numbers, dates, and times
·Battery: Alkaline battery
·Power consumption: 120mA during working hours, 50uA during sampling intervals
·Surface material: 316 stainless steel coating
·Communication output: RS—232·Software: can beWindows95—2000Use on top
Optional componentsTurbidity automatic cleaner.
Pressure outlet deviceA gas outlet pipe in the probe cable and an atmospheric pressure outlet gas on the water surface can accurately measure depth with an accuracy of ±0.01m.
Solar power sourcesuitable611,612And other series. This power supply is suitable for long-term unmanned operation,Especially suitable for telemetry options.
telemetryapplicationUHFWireless orGSMPerform long-distance data transmission, data storage, and downloading.
softwareAll models of instruments can be used for data download, data control, wire description, interactive calibration programs, and instrument settings.
|
serial number |
parameter |
sensor type |
measurement range |
accuracy |
resolution |
response time |
|
1 |
temperature |
Platinum electrode sensor, stainless steel sheath, one or two point calibration |
-5-50℃ |
±0.05℃. |
0.01℃ |
15second |
|
2 |
dissolved oxygen |
Galvanic battery type sensor, silver cathode, lithium anode,0.025mmMembrane filter,0%satand100%satCalibration, one or two point calibration |
0-200%sat(0-20mg/l) |
±0.5% . |
0.1%sat(0.1mg/l) |
4second |
|
3 |
depth |
Silicon stress electrode, stainless steel sheath |
0-100m |
±0.5% FRO . |
0.1m |
4second |
|
4 |
turbidity |
90oScattering, infrared light source, wavelength850nm |
0-600NTU |
±0.2NTU(0-200NTU) ±4NTU(200-600NTU) |
0.3NTU |
4second |
|
5 |
conductivity |
The automatic range selection function4Electrode catheter, capable of displaying three sets of ranges:0-8ms/cm;0-8000us/cm;0-60ppt(Salinity) |
0-80ms/cm 0-8000us/cm |
±0.02ms/cm ±3.0us/cm |
0.02ms/cm 2.0us/cm |
90%in10Within seconds |
|
6 |
salinity |
Calculated by conductivity and temperature |
0-60ppt |
±0.05ppt |
0.02ppt |
5second |
|
7 |
PH |
Glass composite electrode |
0-14PH |
±0.03PH |
0.01PH |
5second |
|
8 |
ORP |
Platinum electrode |
±900mv |
±2mv |
1mv |
5second |