Welcome Customer !

Membership

Help

Shanghai Boshitong Electric Appliance Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products
Product Categories

Shanghai Boshitong Electric Appliance Co., Ltd

  • E-mail

  • Phone

  • Address

    Room 1206, No. 2, Lane 3620, Zhongshan North Road, Shanghai

Contact Now

BOD Wade portable biological oxygen demand detector

NegotiableUpdate on 01/30
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
BOD Wade portable biological oxygen demand detector
Product Details

The BOD Wade and BOD Wade Pro portable biological oxygen demand detectors can obtain real-time levels of biological oxygen demand (BOD) in water bodies. The instrument measures by detecting tryptophan ultraviolet fluorescence, which is correlated with 5-day biological oxygen demand (BOD5). BOD Wade includes a UviLux tryptophan fluorescence meter from CTG company and a display storage unit Hawk. BOD Reader Pro includes two UviLux tryptophan fluorescence meters and UviLux CDOM fluorescence meters, used to distinguish tryptophan fluorescence in colored dissolved organic matter (CDOM).
UviLux fluorescence meter has industry-leading sensitivity and is optimized for turbidity environments to minimize the impact of turbidity on measurement. The handheld device Hawk is equipped with a color display touch screen, which can view and store data in real time. Hawk can be programmed according to requirements to apply user set standards to measurement data. Hawk is equipped with a GPS receiver that can locate and time stamp data.
Instrument features
Real time display of BOD5 concentration (mg/l) with high sensitivity
• Touchscreen display
• Record the time and location of data collection
Easy to operate, large storage capacity (2GB)
Application
BOD monitoring and effluent compliance assessment of water treatment units in water treatment plants
Industrial wastewater discharged into surface water for BOD monitoring
• River pollution and investigation BOD monitoring
• Point source pollution investigation
• Groundwater quality monitoring
• Combined sewer overflow (CSO) monitoring
• Fecal indicator (FI) monitoring
Technical parameters
(1) UviLux fluorescence meter
• Size: 70 x 150 mm including connector
• Weight: 800g
• Depth level: 60bar
(2) UviLux parameters
• BOD range: 0.001-50 mg/l
BOD sensitivity: 0.001mg/l
CDOM range: 0.02-900ug/l
CDOM sensitivity: 0.01ug/l
(3) Hawk handheld display and storage devices
• Display screen: 320 x 240 pixel qVGA backlit LCD
• Display screen size: 70 x 50 mm
• Size: 210 x 110 x 45mm
• Weight: 500g
• Memory card: 2G
• Battery: 4 hours of continuous use
• Protection level: IP67
• 操作温度: -2 ºC ~40 ºC
• Storage temperature: -40 º C~70 º C