The main technical indicators of the non dispersive infrared carbon monoxide analyzer meet the technical requirements of the national secondary instrument and can obtain the secondary calibration certificate from the China Institute of Metrology. This instrument has a built-in zero adjustment filter and a six way valve for switching between zero adjustment and measurement, making it easy and flexible to operate.
Product Details

OneNon dispersive infrared carbon monoxide analyzerOverview

This instrument complies with the national standards of GB/T18204.2-2014 "Methods for hygiene inspection in public places" and GB/T9801-88 "Determination of carbon monoxide in air quality - Non dispersive infrared method"; Comply with the national metrological verification regulations of JJG635-2011 "Infrared Gas Analyzer for Carbon Monoxide and Carbon Dioxide".

The main technical indicators of this instrument meet the technical requirements of national second-class instruments and can obtain a second-class calibration certificate from the Chinese Academy of Metrology.

This instrument has a built-in zero adjustment filter and a six way valve for switching between zero adjustment and measurement, making it easy and flexible to operate.

This instrument can customize the measurement range.

It has hourly mean, 8-hour mean, daily average display, and storage functions.


Equipped with wireless Bluetooth printer connection(Optional).

II

Non dispersive infrared carbon monoxide analyzerMain technical indicators


Measurement principle: Non dispersive infrared analysis method/Non dispersive infrared method (NDIR)

Sampling method: built-in pump suction

Measurement range: 0~50.0 × 10-6 or 0~200.0 × 10-6

Resolution: 0.1 × 10-6

Repeatability: ≤ 1% F · S

Zero drift: ≤± 2% F · S/h

Span drift: ≤± 2% F · S/3h

Linear deviation: ≤± 2% F · S

Response time: t0~t90 ≤ 45S

Preheating time: ≤ 30 minutes

Flow range: (0.5-2.0) L/min

Standard configuration: host, sampler, strap, rechargeable battery, charger, small screwdriver, technical documents, portable case

Wireless connection to Bluetooth printer (optional)


Power supply: AC/DC dual-use, 220AVC (± 10%) or built-in rechargeable battery

Dimensions: 245 × 190 × 85 (mm)

Weight: ≤ 3kg