IM2200-EM spectroscopy is designed to provide users with applications that require high sensitivity, such as low light, ultraviolet response, and near-infrared response. Electronic circuits are carefully designed to have considerable flexibility and can be connected to various modules and external interfaces. The IM2000 series spectrometer can communicate with PCs, PLCs, and other embedded controllers through USB 2.0, UART, and SPI, where SPI can achieve bus parallel connection for multiple devices. USB connection can provide VCP interface. VCP communication development is simple, cross platform, and does not require installation and development of drivers. Users can operate the spectrometer by sending and receiving serial commands according to our protocol.
Main features:
High sensitivity from ultraviolet to near-infrared, eliminating high-order diffraction interference;
Circuit design with low noise, low power consumption, and resistance to EMI interference;
16 bit high-precision sampling;
Onboard pulse generator, which can be used to trigger other devices;
• Onboard GPIO, high-precision analog power output;
• Wavelength calibration coefficient, linearity correction coefficient;
• Support software and hardware customization, support secondary development, provide QT C++, LabVIEW, Python, C #, SDK Demo source code and communication protocols;
• System support: Windows, Linux, MAC, Android;
Application
• LED sorting; Multi parameter online water quality analysis; Fluorescence spectroscopy and Raman spectroscopy; • Biochemical analyzer;
• Reflectance detection; • LIBS; • Micro and rapid spectrophotometer; • Transmittance detection;
Pre established spectral range vs. resolution:
spectral range |
spectral resolution |
|||
25um |
50um |
100um |
200um |
|
185-450nm |
0.6nm |
0.8nm |
1.2nm |
2nm |
380-780nm |
0.6nm |
0.8nm |
1.2nm |
2nm |
790nm-1120nm |
0.7 nm |
0.9 nm |
- |
- |
300nm-1000nm |
1.5 nm |
- |
- |
- |
185nm-1030nm |
1.6 nm |
1.8 nm |
3.5 nm |
5.8 nm |
The spectrometer can customize spectral range, resolution and other parameters according to customer needs
Main technical parameters
name/model |
Miniature Fiber Optic Spectrometer / IM2200-EM |
Sensor chip type |
high sensitivityCMOS |
interface |
USB-Type C,TTL |
effective pixels |
2048 x 1 |
pixel size |
14 x 200 μm |
sensitivity |
1300 V/(lx·s) |
signal-to-noise ratio |
370:1 |
Spectral range of detector |
180~1100nm |
wavelength range |
200nm~400nm,380-780nm,790-1120nm,185-1025nm,300-1000nm, customizable |
Incident light interface |
SMA905 0.22NA, Free Space |
optical resolution |
0.1-4 nm(depending on the slit and spectral range) |
TTLSerial port baud rate |
115200/384000/921600switchable |
Width of the incident slit |
10、25、50、100、150The200 μmOptional, other sizes can be customized
|
TTLlevel |
3.3V |
optical platform |
f/4Cross asymmetricC-Toptical path |
TTLanalog output |
0-3.3V,0.8mV,12bithigh precisionDAC |
Maximum frame rate |
USB>120 fps,TTLserial port>20 fps(@921600Baud rate) |
exposure time |
20us ~65sAdjustable precision1us |
average frequency |
1-255 |
power consumption |
0.2A@5V,1.5W |
power supply |
3-5.5VThe maximum cannot exceed5.5V |
size |
76×61×32mm |
weight |
240g |
Storage temperature |
-20℃~60℃ |
Operating Temperature |
-20℃~60℃ |
Spectral response diagram |
|
TTLInterface (connector type)SH1.0-16P, voltage level3.3V):

pin |
name |
function |
1 Pin |
VSYS |
3-5.2Vpower input |
2 Pin |
VGND |
Negative power supply |
3 Pin |
USB DP |
USBdataD+ |
4 Pin |
USB DM |
USBdataD- |
5 Pin |
SPI_MOSI |
SPIdata cable |
6 Pin |
SPI_MISO |
SPIdata cable |
7 Pin |
SPI_SCK |
SPIClock signal line |
8 Pin |
SPI_NSS |
SPIChip selection signal line |
9 Pin |
UART RX |
Serial data interface port |
10Pin |
UART TX |
Serial data transmission port |
11Pin |
Trigger In1 |
Single external trigger input |
12Pin |
Trigger In2 |
Continuous external trigger input |
13Pin |
Sync |
Synchronous signal output |
14Pin |
Trigger Out2 |
Programmable output |
15Pin |
Analog Out |
0-3.3VProgrammable voltage output |
16Pin |
GND |
signal ground |
l Trigger In1Start capturing the current image at the rising edge of the pulse, and automatically output the image upon completion. The exposure time isMCUcontrol,Default high level, triggered by falling edge.
l Trigger In2External trigger pin,3.3VLevel,PWMPulse input mode, high level duration isCCDThe integration time, low-level time refers to the acquisition and transmission time, and the minimum low-level time needs to last4msThe minimum duration required for high level is10us.
l SyncOutput synchronous trigger signal of sensor, positive pulse output, rising edge image sensor electronic shutter opens, sensor begins integration, high-level duration is exposure time, can be directly connected to xenon lamptrigPin.
l Trigger Out2Programmable output provides a control interface, allowing users to control the output status of the port through software.
l Analog OutProgrammable analog voltage output interface,0-3.3VVoltage range and resolution0.8mV.
l VGNDandGNDdifference:GNDConnect the internal signal ground,VGNDThere is a model inductor connected to the signal ground.
Size chart

Application Examples
8.1, UseIMSpectralSuitSoftware measures the UV spectrum of mercury lamps and real-time detects the waveform of changes in single point wavelength values:

8.2 405nmLaser measured spectrum:

8.3 185-1030nm@25umSlit, mercury lamp measured spectrum:

185-1030nm@25umNarrow slit,LEDActual measured spectrum:

8.4 Use785Laser probe testing of acetonitrile Raman spectroscopy(500mW @500ms,50umSlit):