- Phone
-
Address
L19-005, Building L19, South China International Leather and Leather Raw Material Logistics Zone (Phase I), No.1 South China Avenue, Hehua Community, Pinghu Street, Longgang District, Shenzhen City
Shenzhen Yimai Optoelectronics Co., Ltd
L19-005, Building L19, South China International Leather and Leather Raw Material Logistics Zone (Phase I), No.1 South China Avenue, Hehua Community, Pinghu Street, Longgang District, Shenzhen City
Product application
◇ UV, visible, and infrared spectroscopic analysis
◇ Fluorescence analysis
Medical: Enzyme immunoassay analyzer, urine imaging device, small POCT biochemical analyzer
◇ Industrial automation strobe light source ◇ Air NO, SO2 pollution gas analysis
◇ Laboratory and online water pollution analysis
Product Features
High radiation intensity and instantaneous flash power are beneficial for improving the signal-to-noise ratio of the instrument
◇ Wide spectral range, covering ultraviolet, visible, and infrared, with a range of 185-2000nm
◇ Long service life, flash frequency ≥ 109 times
Good stability, CV value ≤ 2% after 1000 flashes
◇ Small size, low heat generation, conducive to optical and mechanical stability
◇ No preheating required, start immediately, saving testing time
DB-9 connector, plug and play
Basic Product Parameters Table
output parameters
project |
parameter |
|
XFM050-*** |
XFM050-***-a |
|
Discharge arc length |
1.5 ±0.2 mm |
3.0 ±0.3 mm |
Transparent panel material |
throughUVglass |
|
Wavelength range of optical radiation |
185 - 2000 nm |
|
Light output stability |
Max 2.0% CV ; Max6.5% PV -1000flash*① |
Max 1.5% CV ; Max4.2% PV -1000flash*① |
Type 0.75% CV; Typ3.5% PV -1000flash*① |
Type0.55% CV; Max3.0% PV -1000flash*① |
|
Lifespan (number of flashes) |
≥109time |
|
Maximum average power |
continuous5W *② |
|
Discharge capacitor |
0.047、0.11、0.22The0.28 μF *③ |
|
Main discharge voltage |
400 - 600 V *④ |
|
Input electrical parameters
project |
parameter |
input voltage |
DC -Min 11.0 - Max 28.0 V |
input current |
11 V:Max 1.4 A |
trigger voltage |
5VSquare wave pulse; High level pulse width≥10 μS |
Maximum triggering frequency |
C=0.047μF,V=600V : Max 530 Hz *⑤ |
C=0.11μF,V=600V : Max 252 Hz | |
C=0.22μF,V=600V : Max 126Hz | |
C=0.28μF,V=600V: Max 100 Hz |
operating conditions
project |
parameter |
Environmental temperature for use |
0 - +40°C |
Storage temperature |
-40 - +90°C |
Use and store humidity |
Max 95% Rh |
cooling method |
natural cooling *⑥ |
quality |
①Light output stability
CV=Standard deviation of light output energy value(Stdev)/Average value)×100%, 取1000Secondary flash data.
PV=Peak difference value of light output energy(Max-Min)/Average value)×100%, 取1000Secondary flash data.
②Maximum average power
average powerP(W)=Single flash energy(J)×Flicker frequency(Hz)The light bulb is outputting5WThe following can work normally.
Single flash energy(J)=1/2CU2Among themCFor the discharge capacitance value,UThe main discharge voltage value at both ends of the capacitor.
③Discharge capacitance value
Under the same main discharge voltage, the energy of a single flash depends on the capacitance value, and the larger the capacitance value, the greater the energy of a single flash.
If the output of the flashing xenon lamp module is kept at the same power condition, the larger the capacitor, the lower the corresponding maximum operating frequency.
④Main discharge voltage
The voltage applied across the discharge capacitor,XFM050The series is400-600VAdjustable.
adjustment method:INTMode, rotate the potentiometer on the back;EXTpattern,DB-9interface3&8Needle, inputDC3.2-4.8VVoltage.
⑤Maximum triggering frequency
According to the power calculation formulaP=1/2CU2×fWhen the main discharge voltageUWhen constant, byPmax=5W, can be calculatedfValue.
But in order for the xenon lamp to charge the main discharge capacitor within the flash interval, the triggering frequency needs to be adjusted≤530Hz.
⑥cooling method