Laser Spectral Analyzer,FLR9
Wide harmonic laser spectroscopy gas analyzer,FLR9,It is a real-time environmental gas analyzer specifically designed for environmental pollution investigation.
FLR9Perform real-time measurement of multiple gases and analyze trace gas components. This analyzer can measure in real-timeOver 500 types of organic compoundsVOCs(Aldehydes, benzene derivatives, terpenes, alcohols, halogenated hydrocarbons, esters, amines, chlorides, thiols, etc.)French Plutonium Industry Science Company TechnologyBreakthrough will be achievedBTLThe wide tuning laser source is combined with a detection unit to analyze and detect the concentration of each gas in real time. Based on different measurement ranges for analysis, the minimum that can be analyzed isPPBLevel. It can also provide an emergency detection mode to adapt to faster and more effective emergency inspections, and can be used to detect on-site gas leaks in industrial processes.
Product Features:FLR9The series is used for environmental monitoring, and the equipment is easy to move and durable, suitable for various sites, and can be operated in a few minutes. It can be controlled locally or remotely through a tablet based interface. Its design considers minimizing maintenance requirements (no consumables, no need for frequent calibration), and the display of measurement results is on the local control device and/Both can be customized on remote control devices: this allows users to control data information when needed. The measurement data is securely stored on the analytical instrument.
Product Application:If you need to conduct environmental inspection and supervision,FLR9The analyzer will immediately address many specific pollutant issues for you. It will help you draw the entire process and environmental status: where, when, and what concentration of pollutants are emitted.
If you need to monitor for environmental violations,FLR9 analyzerIt can be installed continuously or temporarily on site, continuously measuring pollutants of interest and reporting alerts in real-time.
Measuring principle |
Laser absorption spectroscopy and photoacoustic detection technology |
Testing Scope* |
ppb-% |
Detection limit* |
1 ppb (direct measurement,No need for pre enrichment) |
Measure components |
>500 types |
reproducibility* |
±1%Half range |
linear |
R2 > 0.99 |
Measurement frequency** |
Up to 1 Hz |
Ambient temperature |
10 °C~+35 °C |
Maximum sampling temperature |
|
humidityRH |
maximum95% (non-condensing) |
Application Area |
Under normal working conditions, explosion-proof areas are optionalATEX / IECex in Ex |
protection level |
Cabinet style container: IP 20Mobile box: IP 54 |
Power supply |
24 V DC/ 3.5 A |
power** |
40W typical |
Installation Method |
cabinet-type(6 U)/trolley case/Explosion proof shell |
size |
cabinet-type524 x 448 x 260 (mm) trolley case621 x 520 x 422 (mm) |
weight**** |
cabinet-type17-19 kg |
trolley case29-31 kg | |
sampling method |
extractive(Process or environmental gases) |
* The actual measurement range and measurement limits depend on different applications and the gas being measured. Please contact us for further details.
**According to different applications
***Most components have strong absorption peaks that can be analyzed and measured
****According to the optional items