Large stroke, vertical precision Z-direction locator piezoelectric ceramic nano stage NM-300 customized price, please contact
Please contact customer service for customized product prices
Product Description: How to Translate
The NM-300 compact linear nano positioning platform can provide motion of 100 μ m or greater along the X direction. Its compact design makes it an ideal tool for many photonics and fiber positioning applications. The Z-axis nano positioning linear displacement platform is a linear displacement platform driven by piezoelectric ceramics.
The nanodisplacement platform uses highly reliable multi-layer low-voltage piezoelectric stacked ceramics as driving elements, and an integrated flexible guiding system optimized by finite element method can ensure good parallel motion. Due to the use of a frictionless flexible mechanism, the piezoelectric displacement stage requires no maintenance, has no wear, and has fast response and stable performance, making it suitable for positioning control, such as step or continuous scanning. Piezoelectric ceramic actuators are resin insulated. This can prevent moisture and avoid faults caused by increased leakage current. Its lifespan has been proven to run 100 billion cycles without any issues.
The closed-loop version has a high-resolution strain gauge position sensor, which can achieve high-precision and repeatable motion, and can also compensate for the creep of piezoelectric ceramics. The sensor adopts a fully closed-loop bridge design. It can operate in open-loop or closed-loop control.
Product Parameters:
| model |
NM-300 |
NM-300S |
tolerance |
| motion axis |
Z |
Z |
|
| Integrated sensors |
|
SGS |
|
| Nominal travel range (@ 0~+120V) |
300μm |
300μm |
±10% |
| Maximum travel range (@ -20~+150V) |
350μm |
350μm |
±20% |
| resolution |
3.0nm |
6.0nm |
type. |
| Minimum moving step size |
10nm |
10 nm |
type. |
| Linear error cha |
|
0.2% |
type. |
| Repetitive positioning accuracy (full stroke) |
|
0.1% |
type. |
| Angular deviation (pitch/yaw/roll) |
50μrad |
50μrad |
type. |
| bearing capacity |
10N |
10N |
Max. |
| Resonant frequency (0g) |
360Hz |
360 Hz |
±20% |
| Movement direction stiffness |
0.25N/μm |
0.25N/μm |
±20% |
| Movement direction thrust/tension |
10N/8N |
10N /8N |
±20% |
| drive mechanism |
piezoelectric ceramic |
piezoelectric ceramic |
|
| Working temperature range |
-20~+80℃ |
-20~+80 °℃ |
type. |
| Cable length |
1.5m |
1.5 m |
±10mm |
| Hollow size |
|
|
|
| Dimensions (length x width x height) |
50×50×17.5mm |
50×50×17.5mm |
|
| weight |
250g |
250g |
±5% |
| PZT&Sensor connector |
LEMO |
LEMO |
|
| body material |
Steel and aluminum |
Steel and aluminum |
|
Note: All specification parameters are measured at room temperature (22 ° C ± 3 ° C). For more information, please contact us. The "-" in the table indicates not applicable.
1. The suffix "- S" indicates a closed loop with a built-in position sensor. Open loop suffix without "S", no linear error, and no repeatability data.
2. The maximum driving voltage is -20V~+150V. For high reliability long-term use, it is recommended to use a driving voltage of 0.+120V. For larger travel, please consult. The factory has a nominal stroke.
3. Resolution refers to positioning noise. Because the piezoelectric ceramic nano positioning system has no friction, the system resolution is only limited by amplifier noise and measurement techniques. Usually measured and specified using RMS (1 σ) value.
4. It refers to the minimum step size value that can be moved by the piezoelectric platform measured by the interferometer, which is mainly determined by the DAC resolution of the piezoelectric controller. In the best case, the displacement step size value corresponding to 16 Bit is listed in the parameter table.
5. Axial load, please consult for higher loads. Note that exceeding the nominal load may cause damage to the flexible mechanism.
6. Please consult for longer cable lengths.
7. The weight of the platform does not include the weight of cables and connectors.
8. Max represents the maximum value in the tolerance; Typ. represents the standard typical data value, and the actual measurement data value will be provided in the testing report.
Product dimensions: