E-mail
info@giantforce.cn
-
Phone
18911365393
-
Address
1311, Building B, Wanda Plaza, No. 58 Xinhua West Street, Tongzhou District, Beijing
Juli Optoelectronics (Beijing) Technology Co., Ltd
info@giantforce.cn
18911365393
1311, Building B, Wanda Plaza, No. 58 Xinhua West Street, Tongzhou District, Beijing
Large area LED solar simulator
Main applications:
Large area LED solar simulatorMeet the testing requirements for various types of solar cells, such as amorphous silicon (a-Si), microcrystalline silicon (nc Si, mc Si), semiconductor compounds [CdS, CdTe (cadmium telluride) CuInSe2]、 Testing of dye-sensitized solar cells (DSSCs), CIGS (copper indium selenide), organic/polymer solar cells, perovskite solar cells, stacked solar cells, etc;
Key Features:
Long life LED light source: an average of 20000 hours;
Effective spot area: 500mmX500mm;
Match standard AM1.5G spectrum;
• Effective spectral range: 300nm~1200nm;
• Initial light intensity of 1.3sun;
Match the international standard IEC60904-9 Ed.3;
Spectral mismatch:<± 12.5%A+level
• Uneven light intensity:<2%A-level
• Light intensity instability:<1%A+level(LTI)
• Light intensity software adjustment function, ranging from 10% to 100%;
Spectral software adjustment function;
Real time feedback control unit for light intensity;
• Working distance: 510mm for Class A non-uniformity;
• Cooling method: air cooling;
• Room size: W 900 × H 780 × D 1177 mm
• Power size: Size: W 652 × H 693 × D 786 mm
Actual measured spectrum:


IV Testing Unit:
KTITHLEY Digital Source Table

Standard battery:

• Traceability: Measurement institutions;
Standard battery size: 20mm * 20mm, made of monocrystalline silicon battery material;
Window glass options: KG2, KG5, and BK5
• Built in PT100 standard temperature sensor;
Certification body issues a certificate;
• Attached test data.
solar cellIV Testing Software:
1. IV characteristic curve
(1) Under illumination conditions, a single test (including forward voltage scan and reverse voltage scan)Record the IV characteristic curve of the photovoltaic cell and output various indicator parameters, including Voc (open circuit voltage), Isc (short-circuit current), Pmax, Vmaxp, Imaxp, Rs (series resistance), Rsh (parallel resistance), Jsc (short-circuit current density), FF (fill factor), and η (luminous efficiency). At the same time, the IV curve of the entire quadrant can be viewed;
(2) Under illumination conditions, multiple consecutive measurements (including forward voltage scan and reverse voltage scan)Record the current and multiple measurements of the photovoltaic cell's IV characteristic curve, and display real-time statistical data of each test result. The interval time between multiple measurements can be customized;
(3) Voltage scanning direction controlUsers can perform single or multiple forward or reverse scan voltage IV tests as needed, as well as single or multiple alternating forward and reverse voltage tests;
(4) In the dark state, measure the parallel resistance Rsh of the battery, including single measurement and multiple continuous measurements;
2. The calibration of the solar simulator uses standard photovoltaic cells to calibrate the intensity of the solar simulator;
3. Dark state testing measures the series resistance Rs to verify the characteristics of the battery sample;
4. Steady state (bias) testing at Vpmax ormustUnder bias voltage, measure the stability curves of efficiency η, current density Jsc, current I, and power point Pmax over time t;
5. MPPT tracking test scans forward or backward, always tracking and measuring the changes in Pmax power point over time; Simultaneously measure the time-varying curves of Vmpp, Impp, and η conversion efficiency corresponding to this condition;
6. Repeatability Test measures the attenuation curves of various parameters of solar cells over multiple measurements or over time;
7. Automatic or manual data saving function: CSV, TXT, RTF, and JPG screenshots.

Sample testing fixture(can be customized according to requirements)