Zhengzhou Wanyong Distillation Device JTZL-6 Sulfur Dioxide Analyzer Product Description: JTZL-6 Fully Automatic Integrated Distillation Analyzer is a new intelligent distillation treatment device that integrates constant temperature heating, automatic control of distillation endpoint, and cooling water circulation according to the laboratory distillation pretreatment operating procedures. Intelligent Distillation Instrument
Intelligent Distillation InstrumentThe heating device is mainly composed of four parts: heating device, distillation device, circulating cooling water device, and receiving device. The heating device is equipped with intelligent control function for heating rate, which can achieve precise control of heating temperature and heating efficiency; The intelligent integrated distillation instrument is designed with high distillation efficiency and good condensation effect; The circulating cooling device is equipped with cooling temperature display and control functions to ensure the cooling effect. The receiving device is equipped with distillation endpoint detection and automatic heating stop functions, achieving intelligent heating control.
Intelligent Distillation InstrumentOperation steps for use
1. Install and fix the supporting glass instrument on the given glass device bracket.
2. Connect all hoses properly. Familiarize yourself with the nitrogen blowing control pipeline again (only sulfur dioxide is selected); The inlet and outlet pipelines of circulating water; Anti backflow suction (four or six way anti vacuum valve) pipeline; Automatic cleaning of pipelines.
3. Familiar with the main power switch, touch screen (visible operation upon startup), heating control (four or six way single temperature control; right rotation temperature increases, otherwise it decreases), and mixing control (four or six way single control, right rotation speed increases, otherwise it decreases) on the panel
4. Familiar with the main power sockets (temperature control, mixing control power) and auxiliary power sockets (cooling, circulating water) on the right side; And connect the corresponding power supply (220V, 50HZ)
5. Familiarize yourself with the main power switch (temperature control, mixing control power), auxiliary power switch (cooling, circulating water), circulating water control switch, and water addition switch on the right side (before turning on the cooling and circulating water, connect the water plastic joint with a silicone hose and insert the other end into the prepared clean bucket of water; then turn on the water addition switch, and the water will automatically add; there is a water level gauge on the left side behind the instrument; the water level is displayed; the water level should be at the middle height of the water level gauge)
6. Turn on the main power switch (on the left side of the panel).
7. Click on "Intelligent Integrated Distillation Instrument"
8. Click on the "Sulfur Dioxide Residue Detector"
9. Select "Recipe Setting" at the lower right corner of the main interface above, the following recipe setting interface will be displayed (please note that the recipe setting will change to "ammonia nitrogen", "volatile phenol", etc. in turn with the click of "previous mode" - "next mode"; select and set the required test parameters as required). Set the temperature and time of each route according to the experimental requirements. (Because there is no weight requirement for sulfur dioxide, it can be uniformly set to 300 or 500 (technical personnel clarify this question)
10. Click to open "Heating Parameters" and set the heating parameters
11. Once the formula is set, please follow the instructions in the diagram, click "Save Parameters", and then click "Use Parameters"; Then click 'Return', and the main interface icon will appear again
Technical Specifications:
style |
Simple type |
economy |
standard form |
model |
JTZL-6C |
Jtzl- 6 y |
JTZL-6 |
Heating method |
Electric heating,Single hole power≤800W |
heating unit |
6 far-infrared ceramic heating furnaces, capable of single hole independent control |
蒸馏速度 |
12ml/min |
time control |
0-999min |
cooling method |
External condensate water |
Built in air-cooled circulating water device, no need for external circulating water. If high circulating water temperature is required, other circulating water devices can be connected externally |
Temperature Display |
digital display |
7-inch large LCD touch screen |
Anti backflow suction |
Optional |
Equipped with anti vacuum solenoid valve and anti reverse suction function |
warm-up time |
8-20min |
5-20min |
Temperature control range |
room temperature-400 degrees |
Endpoint control |
Manual control, optional weighing function |
Automatic detection of distillation endpoint function, dual control of weighing and time, automatic stop of heating, control accuracy±2ml |
Distillation flask specifications |
500mlX6 or 250ml * 6 (optional according to customer requirements) |
rated voltage/Frequency |
220V/50HZ |