1、 Product Description
At present, sterilization workshops, laboratories, and operating rooms in industries such as pharmaceuticals, biological products, health products, beverages, and hospitals widely use the colony assay to determine the number of bacteria in the air.
Although this method is simple, it has many drawbacks:
(1) Cannot know the number of bacteria contained in a unit volume of air;
(2) Due to the slow natural sedimentation of bacteria, it is difficult to collect, with low efficiency and long collection time;
(3) The sedimentation of bacteria is influenced by airflow, resulting in significant errors.
JCQ-5Integrated planktonic bacteria samplerOvercoming these shortcomings, we designed and manufactured according to the requirements of China's GMP standards and referenced similar foreign products. Has the following advantages:
(1) The collection port is designed with 397 micropores, which can reduce the overlap of dust and bacteria during the sampling process and reduce counting errors (Anderson principle level 5);
(2) SuXin independently develops flow sensors with stable sampling flow;
(3) Infrared remote control operation to avoid the influence of human factors during the sampling process;
(4) Parameters such as sampling volume and sampling time are stored on a page by page basis and can be communicated with a computer through a USB port;
(5) Large capacity rechargeable battery, capable of continuous sampling for over 6 hours;
(6) When used in conjunction with the SuXin compressed gas detection device, it can detect the concentration of microorganisms in compressed gas.
IIJCQ-5Integrated planktonic bacteria samplertechnical indicators
sampling flow |
100L/min±5% |
Sampling head material |
The sampling head is made of aluminum alloy material |
sampling flow |
100L/min±5% |
Sampling period adjustment range |
1-6000L (adjustable) |
Impact velocity |
10.8m/s (Anderson Impact Level 5 Principle) |
Overall dimensions |
180*260*151mm(W*D*H) |
weight |
4Kg |
execution standard |
ISO 14698-1/2,GB/T 16293-2010 |
3、 Usage conditions
1. Normal atmospheric pressure environment
2. Temperature: 5-40 ℃
3. Relative humidity: 10~70% RH
4、 Product Features
1. The large sampling period can reach 6000L;
2. The impact velocity of the sampling head is around 10.8m/s, ensuring that particles larger than 1 μ m can be captured;
3. Porous sampling head, using standard ɸ 90mm * 15mm agar culture dish.