For routine meteorological and other data
Designed for recording applications
HygroVUE5 digital temperature and humidity sensorTemperature and relative humidity sensors are designed for routine meteorological and environmental monitoring. It adoptsSDI-12Communication protocol, can connect to any supported networkSDI-12The protocol's data collector simplifies installation and programming work. Sensor components can be replaced on-site, reducing downtime and calibration costs,HygroVUE™ 5belong toHygroVUE™Entry level temperature and humidity sensors in the product line.
Advantages and Characteristics
u Use a combination of pre calibrated humidity and temperature components
u Can replace components on site for quick recalibration
u SDI-12output,Allow long-distance communication
u Simple data acquisition programming
u low power consumption
u Wide working voltage range
u Durable and sturdy packaging design
u Small size, suitable for small radiation shields
Details
HygroVUE5 digital temperature and humidity sensorTemperature and relative humidity sensors are designed for routine meteorological and environmental monitoring. It adoptsUsers can replace the sensor chip components, which include a temperature sensor and a relative humidity sensor. Each component can be individually calibrated using pre stored calibration coefficients on the chip.
HygroVUE™ 5Control the measurement of sensor components through internal electronic devices, perform linearization correction on temperature readings, and useSDI-12Provide data to the data collector.
HygroVUE™ 5The stainless steel filter screen not only reduces the impact of dust and dirt on the sensor, but also ensures air circulation around the sensor element and reduces the possibility of condensate remaining inside the filter. In addition, the sensor surface is covered with a small polytetrafluoroethylene film filter to prevent dust or mold from affecting the measurement.
The sensor housing design can be exposed to any weather conditions at all times. Suitable for a range of radiation protection covers, including small protective covers such asRAD066-layer radiation shielding cover.
In order to improve reliability, sensors are installed with fixed length cables in the factory, and then the sensor body is encapsulated to seal electronic devices and cable connectors. If there are special needs, other lengths of cables can be customized.
Technical Specifications:
Inductive element |
SHT35 derivative (special coating ensures reliability) |
|
Calibration traceability |
NIST and NPL standards |
|
power supply voltage |
7 to 28 VDC |
|
warm-up time |
Sensors are usually powered on continuously. If the power is turned off, the preheating time is1.8 s |
|
Main shell material |
whitePET-P |
|
Protection level |
IP67 (electronic device casing) |
|
Sensor protection |
Equipped with nominal aperture<30 μ m polypropylene filter screen. The sensor element is equipped with a polytetrafluoroethylene protective film, which has a filtration efficiency of over 99.99% for particles with a size of 200 nm and larger. |
|
comply withEMC standard |
Tested and compliantIEC61326:2013 |
|
calibration |
Sensor components are calibrated separately during the manufacturing process |
|
On site chip replacement or recalibration is available |
On site replaceable chips |
|
diameter |
12.5 mm (0.49 in) sensor section, maximum 16 mm (0.63 in) cable end |
|
length |
115mm (4.52 in) sensor section (excluding cables) |
|
Maximum current consumption |
||
dormant state |
50 μA |
|
During the measurement period |
0.6 mA (0.5 s measurement time) |
|
air temperature |
||
measurement range |
-40 to+70 ℃ |
|
accuracy |
± 0.4 ℃ (-40 to+70 ℃) ± 0.3 ℃ (-20 to+60 ℃) |
|
Long term drift |
<± 0.03 ℃ per year |
|
resolution |
0.001 ℃ |
|
repetitiveness |
0.04 ℃ (3 standard deviations measured 25 times at constant temperature) |
|
response time |
130 seconds (63% of the reference response time at an air movement speed of 1 m/s) |
|
unit |
Celsius |
|
relative humidity |
||
measurement range |
0 to 100% RH |
|
accuracy |
± 3% (25 ℃, 80 to 100% RH) ± 1.8% (25 ℃, 0 to 80% RH) |
|
Additional errors at other temperatures |
<± 1% RH (-40 to+60 ℃) |
|
Short term delay |
<±1% RH |
|
long-term stability |
± 0.5% per year (maximum drift under clean air conditions) |
|
resolution |
0.001% RH |
|
repetitiveness |
0.05% (3 standard deviations measured 25 times at constant humidity) |
|
Response time of the filter |