- Phone
-
Address
Chengnan Industrial Zone, Baoying County, Jiangsu Province
Yangzhou Kotuo Electric Technology Co., Ltd
Chengnan Industrial Zone, Baoying County, Jiangsu Province
Fully automatic insulation oil dielectric strength testerIt is a high-precision, fully digital industrial instrument developed by all scientific research and technical personnel of our company, based on the relevant provisions of national standard GB507-1986 and industry standard DL/T846.7-2004, leveraging their own advantages, through multiple on-site tests and long-term unremitting efforts. To meet the needs of different users, this series of instruments can be divided into single cup three cup and multi cup models. The instrument is easy to operate and has a beautiful and elegant appearance. Due to the use of fully automatic digital microcomputer control, it has high measurement accuracy, strong anti-interference ability, and is safe and reliable.
1. The instrument is controlled by a large capacity microcontroller, which ensures stable and reliable operation;
2. The instrument is equipped with a wide range watchdog circuit to prevent system crashes;
3. Multiple operation options, instrument program setGB507-1986TheGB507-2002Two national standard methods、industry standardDL/429International standardsIEC156andCustom operations that can adapt to various choices of different users;
4. The instrument oil cup is made of special glass and cast in one go, which eliminates the occurrence of interference phenomena such as oil leakage;
5. The high voltage sampling design of the instrument allows the test values to enter directlyA/DConverter avoids errors caused in analog circuits, making measurement results more accurate;
6. The instrument has functions such as overcurrent, overvoltage, and short circuit protection inside, and has anti-interference ability and good electromagnetic compatibility;
7. Portable structure, easy to move, convenient for both indoor and outdoor use.
1. Capacity of booster1.5 kVA
2. Boosting speed0.5kV/s-5.0kV/s(Every0.5increasing)TenFile selection
mistake Poor to 0.2kV/s
3. Output voltage0~80 kV(optional)
4. Voltage accuracy±(2%reading+2Character)
5. Power distortion rate<1%
6. Electrode gap standard2.5 mm
7. Test frequency6 Next time(1-10Secondary optional)
8. Rest time900S(0-5900 Soptional)
9.Interval static time300S(0-5900Soptional)
10.Mixing time15S(0-99SOptional)
11.Display method: Large screen LCD Chinese character display
12.Communication interface standard configuration232Communication interface.
13.Exterior dimensions long430 mm; Wide350 mm; tall370 mm
14.Instrument weight36 kg
1. ambient temperature 0~40℃
2. relative humidity ≤85%
3. working power supply AC 220V ±20
4. electricitySource frequency 50 Hz ± 5Hz
5. powerconsume <200 W