- Phone
-
Address
Room 2201-03, Building A1, Greenland Central Plaza, No. 112 Science Avenue, Huangpu District, Guangzhou City
Southeast Science Instrument
Room 2201-03, Building A1, Greenland Central Plaza, No. 112 Science Avenue, Huangpu District, Guangzhou City
Technical parameters:
model |
External dimensions (mm) |
External dimensions (mm) |
Internal dimensions (mm) |
Airflow rate (inflow) |
Airflow rate (sinking) |
power supply voltage |
AC2-4D1 |
1340 x 753 x 1400 |
1340 x 810 x 1400 |
1220 x 580 x 660 |
0.45 m/s (90 fpm) |
0.30m/s (60 fpm) |
220 V 50 Hz |
AC2-6D1 |
1950 x 753 x 1400 |
1950 x 810 x 1400 |
1830 x 580 x 660 |
0.45 m/s (90 fpm) |
0.30m/s (60 fpm) |
220 V 50 Hz |
Performance characteristics:
The EscoSentinel Gold ultra large screen microcomputer programmable system displays safety information on one screen.
The airflow sensor will remind users if the airflow can provide higher safety.
The RS 232 serial interface sends operational information to the Building Management System (BMS).
Esco ISOCIDE antibacterial coating prevents the growth and contamination of bacteria and microorganisms on the surface of the cabinet.
Single piece stainless steel workbench with arched design to prevent liquid leakage.
The operating chamber is formed from a single steel plate, without joints or screws, and is easy to clean.
Optimize the isolation efficiency of the side interception zone and negative pressure side wall.
Ergonomic design; Raising the armrest helps prevent the grille from intercepting and creates a more comfortable working posture.
We have the following size models: 1.2m (4 ') and 8m (6').
ULPA high-efficiency air filters have a retention efficiency of 99.999% for 3-micron particle systems and 99.995% for 0.1 micron particle systems.
The Esco AirstreamA2 type secondary safety cabinet adopts a DCECM maintenance free fan system with a constant wind speed, and comes with voltage fluctuation compensation function to maintain a constant wind speed within a wide voltage fluctuation range of 190-250V; Equipped with resistance sensing compensation function, it can provide safe wind speed even when the filter blockage pressure increases by 30%. Adopting ULPA ji ultra high efficiency air filters with high safety performance, it can provide vertical laminar airflow to the workbench to protect the samples, while meeting the Class 1 air cleanliness standard (compliant with Class 3 of 1S0 1464.1 standard). The cabinet surface is coated with ESCO ISOCIDE antibacterial coating, which can effectively inhibit the growth of microorganisms on the cabinet surface. Compliant with EN 12469 certification, NSF/ANSI49 standard in the United States, and YY0569 standard in China.
Product Features:
Comes with wind speed compensation function
High Efficiency and Energy Saving
High cleanliness level in the workspace
Antibacterial coating on the surface of the cabinet reduces microbial contamination
Low noise and ergonomic design